Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courier

被引:17
作者
Peter, John Victor [2 ]
Patole, Shalom [3 ]
Fleming, Jude Joseph [4 ]
Selvakumar, Ratnasamy [4 ]
Graham, Petra L. [1 ]
机构
[1] Macquarie Univ, Dept Stat, Sydney, NSW 2109, Australia
[2] Christian Med Coll & Hosp, Med ICU, Vellore 632004, Tamil Nadu, India
[3] Christian Med Coll & Hosp, Dept Med, Vellore 632004, Tamil Nadu, India
[4] Christian Med Coll & Hosp, Dept Clin Biochem, Vellore 632004, Tamil Nadu, India
关键词
agreement; arterial blood gas; Bland-Altman; concordance; pneumatic tube system; point of care testing; TUBE SYSTEM; AIR BUBBLES; OXYGEN-TENSION; TIME; TEMPERATURE; SPECIMENS;
D O I
10.1515/CCLM.2011.611
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Rapid accurate assessment of metabolic derangements is crucial in the critically ill. We evaluated if arterial blood gas (ABG) samples transported through a pneumatic tube system (PTS) agreed with values transported by a human courier. Methods: In this prospective study of 50-paired ABG samples, the couriered reference ABG was compared with those transported by PTS. Agreement was summarised by the mean difference with 95% limits of agreement (LOA) and Lin's concordance correlation (pc). Results: The mean (+/- SD) time from sampling to analysis was 35.7 +/- 23.2 (courier) and 38.6 +/- 22.1 (PTS) minutes. Agreement was good between courier and PTS for pH, PaCO(2), bicarbonate, oxygen saturation and PaO(2) values (pc > 0.97). Although the mean difference in PaO(2) values between PTS and courier was small (-0.9 mm Hg) and the agreement was good, individual differences were clinically significant (95% LOA -40.8 to 39.0). For PaO(2) < 160 mm Hg, analysis of PTS samples yielded erroneously high PaO(2) values and vice versa for PaO(2) > 160 mm Hg compared to manual courier. This suggested exaggerated oxygen movement between the blood sample and air in the PTS. Conclusions: In this study, analysis of samples transported through the PTS resulted in clinically unacceptable PaO(2) values. Delay in transport and analysis of ABG samples should be avoided and samples transported manually if they cannot be assessed on-site.
引用
收藏
页码:1303 / 1309
页数:7
相关论文
共 15 条
  • [1] Astles JR, 1996, ARCH PATHOL LAB MED, V120, P642
  • [2] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [3] Evaluation and improvement strategy of analytical turnaround time in the Stat Laboratory
    Chien, Tzu-I
    Lu, Jin-Ying
    Kao, Jau-Tsuen
    Cheng, Ya-Chih
    Lee, Ya-Fen
    [J]. JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2007, 106 (07) : 558 - 564
  • [4] Changes in blood gas samples produced by a pneumatic tube system
    Collinson, PO
    John, CM
    Gaze, DC
    Ferrigan, LF
    Cramp, DG
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 2002, 55 (02) : 105 - 107
  • [5] Use of glass capillaries avoids the time changes in high blood Po2 observed with plastic syringes
    d'Ortho, MP
    Delclaux, C
    Zerah, F
    Herigault, R
    Adnot, S
    Harf, A
    [J]. CHEST, 2001, 120 (05) : 1651 - 1654
  • [6] The effects of time delay and temperature on capillary blood gas measurements
    Dent, RG
    Boniface, DR
    Fyffe, J
    Yousef, Z
    [J]. RESPIRATORY MEDICINE, 1999, 93 (11) : 794 - 797
  • [7] Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality
    Fernandes, CMB
    Worster, A
    Eva, K
    Hill, S
    McCallum, C
    [J]. JOURNAL OF EMERGENCY NURSING, 2006, 32 (02) : 139 - 143
  • [8] Knowles TP, 2006, RESPIR CARE, V51, P732
  • [9] A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY
    LIN, LI
    [J]. BIOMETRICS, 1989, 45 (01) : 255 - 268
  • [10] Lu JY, 2003, J FORMOS MED ASSOC, V102, P246