Toward a pre-clinical irradiator using clinical infrastructure

被引:4
作者
Parsons, David [1 ]
Church, Cody [2 ]
Syme, Alasdair [2 ,3 ]
机构
[1] Univ Texas Southwestern Med Ctr Dallas, Dept Radiat Oncol, 2280 Inwood Rd, Dallas, TX 75390 USA
[2] Dalhousie Univ, Dept Phys & Atmospher Sci, 5820 Univ Ave, Halifax, NS B3H 1V7, Canada
[3] Dalhousie Univ, Dept Radiat Oncol, 5820 Univ Ave, Halifax, NS B3H 1V7, Canada
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 58卷
关键词
Pre-clinical irradiator; Small animal irradiator; Dynamic couch trajectories; Virtual isocentre; RADIATION RESEARCH PLATFORM; X-RAY; CONFORMAL RADIOTHERAPY; SYSTEM; PERFORMANCE; DOSIMETRY; THERAPY; MICRORT; QUALITY; LESIONS;
D O I
10.1016/j.ejmp.2019.01.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Pre-clinical irradiation systems use kilovoltage x-ray systems to deliver small fields of radiation in static beam arrangements or arcs. The systems are costly and the radiobiological effectiveness of kilovoltage beams is known to differ from the megavoltage photon beams used clinically. This work used Developer mode on the Varian TrueBeam STx linear accelerator to create a pre-clinical irradiator capable of treating millimeter-sized targets. Materials and methods: A treatment field defined by a single opposed leaf pair was used to deliver arc-based treatments. Dynamic couch trajectories were used to create a shortened virtual isocentre. Initially, a pre-treatment imaging procedure was used to quantify target misalignment at control points along the arcs and determine appropriate couch positional corrections. This was followed by the treatment arcs in which the positional corrections were implemented. Monte Carlo simulations and radiochromic film were used to calculate and measure dose distributions. Results: A 1 mm leaf separation produced the optimal dose distributions. Couch position corrections up to 2.1 mm were required to maintain a target at virtual isocentre. Application of couch corrections reduced noncoplanar arc treatments dose profile by 1.2 mm at 30% of the maximum dose. Treatment of a 1 mm diameter target would result in falloff distances to the 80%, 50% and 25% of the 90% prescription line of 0.3 mm, 0.5 mm and 1.3 mm from the target edge respectively. Conclusions: This work has demonstrated that it is possible to deliver highly compact dose distributions using megavoltage photon beams from existing clinical infrastructure.
引用
收藏
页码:21 / 31
页数:11
相关论文
共 34 条
  • [21] Commissioning of a novel microCT/RT system for small animal conformal radiotherapy
    Rodriguez, Manuel
    Zhou, Hu
    Keall, Paul
    Graves, Edward
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (12) : 3727 - 3740
  • [22] Rogers, 2018, DOSXYZNRC USERS MANU
  • [23] BEAM - A MONTE-CARLO CODE TO SIMULATE RADIOTHERAPY TREATMENT UNITS
    ROGERS, DWO
    FADDEGON, BA
    DING, GX
    MA, CM
    WE, J
    MACKIE, TR
    [J]. MEDICAL PHYSICS, 1995, 22 (05) : 503 - 524
  • [24] A Phase I trial of stereotactic body radiation therapy (SBRT) for liver metastases
    Schefter, TE
    Kavanagh, BD
    Timmerman, RD
    Cardenes, HR
    Baron, A
    Gaspar, LE
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (05): : 1371 - 1378
  • [25] An x-ray image guidance system for small animal stereotactic irradiation
    Song, K. H.
    Pidikiti, R.
    Stojadinovic, S.
    Speiser, M.
    Seliounine, S.
    Saha, D.
    Solberg, T. D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) : 7345 - 7362
  • [26] Patient specific methods for room-mounted x-ray imagers for monoscopic/stereoscopic prostate motion monitoring
    Stevens, M. Tynan R.
    Parsons, Dave D.
    Robar, James L.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (04): : 40 - 50
  • [27] Continuous monitoring of prostate position using stereoscopic and monoscopic kV image guidance
    Stevens, M. Tynan R.
    Parsons, Dave D.
    Robar, James L.
    [J]. MEDICAL PHYSICS, 2016, 43 (05) : 2558 - 2568
  • [28] Online virtual isocenter based radiation field targeting for high performance small animal microirradiation
    Stewart, James M. P.
    Ansell, Steve
    Lindsay, Patricia E.
    Jaffray, David A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (23) : 9031 - 9046
  • [29] Two-dimensional inverse planning and delivery with a preclinical image guided microirradiator
    Stewart, James M. P.
    Lindsay, Patricia E.
    Jaffray, David A.
    [J]. MEDICAL PHYSICS, 2013, 40 (10)
  • [30] MicroRT - Small animal conformal irradiator
    Stojadinovic, S.
    Low, D. A.
    Hope, A. J.
    Vicic, M.
    Deasy, J. O.
    Cui, J.
    Khullar, D.
    Parikh, P. J.
    Malinowski, K. T.
    Izaguirre, E. W.
    Mutic, S.
    Grigsby, P. W.
    [J]. MEDICAL PHYSICS, 2007, 34 (12) : 4706 - 4716