Toward a pre-clinical irradiator using clinical infrastructure

被引:4
作者
Parsons, David [1 ]
Church, Cody [2 ]
Syme, Alasdair [2 ,3 ]
机构
[1] Univ Texas Southwestern Med Ctr Dallas, Dept Radiat Oncol, 2280 Inwood Rd, Dallas, TX 75390 USA
[2] Dalhousie Univ, Dept Phys & Atmospher Sci, 5820 Univ Ave, Halifax, NS B3H 1V7, Canada
[3] Dalhousie Univ, Dept Radiat Oncol, 5820 Univ Ave, Halifax, NS B3H 1V7, Canada
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 58卷
关键词
Pre-clinical irradiator; Small animal irradiator; Dynamic couch trajectories; Virtual isocentre; RADIATION RESEARCH PLATFORM; X-RAY; CONFORMAL RADIOTHERAPY; SYSTEM; PERFORMANCE; DOSIMETRY; THERAPY; MICRORT; QUALITY; LESIONS;
D O I
10.1016/j.ejmp.2019.01.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Pre-clinical irradiation systems use kilovoltage x-ray systems to deliver small fields of radiation in static beam arrangements or arcs. The systems are costly and the radiobiological effectiveness of kilovoltage beams is known to differ from the megavoltage photon beams used clinically. This work used Developer mode on the Varian TrueBeam STx linear accelerator to create a pre-clinical irradiator capable of treating millimeter-sized targets. Materials and methods: A treatment field defined by a single opposed leaf pair was used to deliver arc-based treatments. Dynamic couch trajectories were used to create a shortened virtual isocentre. Initially, a pre-treatment imaging procedure was used to quantify target misalignment at control points along the arcs and determine appropriate couch positional corrections. This was followed by the treatment arcs in which the positional corrections were implemented. Monte Carlo simulations and radiochromic film were used to calculate and measure dose distributions. Results: A 1 mm leaf separation produced the optimal dose distributions. Couch position corrections up to 2.1 mm were required to maintain a target at virtual isocentre. Application of couch corrections reduced noncoplanar arc treatments dose profile by 1.2 mm at 30% of the maximum dose. Treatment of a 1 mm diameter target would result in falloff distances to the 80%, 50% and 25% of the 90% prescription line of 0.3 mm, 0.5 mm and 1.3 mm from the target edge respectively. Conclusions: This work has demonstrated that it is possible to deliver highly compact dose distributions using megavoltage photon beams from existing clinical infrastructure.
引用
收藏
页码:21 / 31
页数:11
相关论文
共 34 条
  • [1] RADIOBIOLOGICAL EFFECTIVENESS (RBE) OF MEGAVOLTAGE X-RAY AND ELECTRON-BEAMS IN RADIOTHERAPY
    AMOLS, HI
    LAGUEUX, B
    CAGNA, D
    [J]. RADIATION RESEARCH, 1986, 105 (01) : 58 - 67
  • [2] Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: phase III results of the RTOG 9508 randomised trial
    Andrews, DW
    Scott, CB
    Sperduto, PW
    Flanders, AE
    Gaspar, LE
    Schell, MC
    Werner-Wasik, M
    Demas, W
    Ryu, J
    Bahary, JP
    Souhami, L
    Rotman, M
    Mehta, MP
    Curran, WJ
    [J]. LANCET, 2004, 363 (9422) : 1665 - 1672
  • [3] The safety and efficacy of robotic image-guided radiosurgery system treatment for intra- and extracranial lesions: A systematic review of the literature
    Calcerrada Diaz-Santos, Nieves
    Blasco Amaro, Juan Antonio
    Ariza Cardiel, Gloria
    Andradas Aragones, Elena
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 89 (03) : 245 - 253
  • [4] Stereotactic Radiotherapy for Unresectable Adenocarcinoma of the Pancreas
    Chang, Daniel T.
    Schellenberg, Devin
    Shen, John
    Kim, Jeff
    Goodman, Karyn A.
    Fisher, George A.
    Ford, James M.
    Desser, Terry
    Quon, Andrew
    Koong, Albert C.
    [J]. CANCER, 2009, 115 (03) : 665 - 672
  • [5] Characterization of image quality and image-guidance performance of a preclinical microirradiator
    Clarkson, R.
    Lindsay, P. E.
    Ansell, S.
    Wilson, G.
    Jelveh, S.
    Hill, R. P.
    Jaffray, D. A.
    [J]. MEDICAL PHYSICS, 2011, 38 (02) : 845 - 856
  • [6] Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA-compliant phase space calculations
    Constantin, Magdalena
    Perl, Joseph
    LoSasso, Tom
    Salop, Arthur
    Whittum, David
    Narula, Anisha
    Svatos, Michelle
    Keall, Paul J.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4018 - 4024
  • [7] The small-animal radiation research platform (SARRP): dosimetry of a focused lens system
    Deng, Hua
    Kennedy, Christopher W.
    Armour, Elwood
    Tryggestad, Erik
    Ford, Eric
    McNutt, Todd
    Jiang, Licai
    Wong, John
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (10) : 2729 - 2740
  • [8] Design and evaluation of a variable aperture collimator for conformal radiotherapy of small animals using a microCT scanner
    Graves, Edward E.
    Zhou, Hu
    Chatterjee, Raja
    Keall, Paul J.
    Gambhir, Sanjiv Sam
    Contag, Christopher H.
    Boyer, Arthur L.
    [J]. MEDICAL PHYSICS, 2007, 34 (11) : 4359 - 4367
  • [9] The variation in biological effectiveness of X-rays and gamma rays with energy
    Hill, MA
    [J]. RADIATION PROTECTION DOSIMETRY, 2004, 112 (04) : 471 - 481
  • [10] Ibbott G.S., 1985, Med. Phys, V12, P813, DOI 10.1118/1.595780