Fracture strength of all-ceramic posterior inlay-retained fixed partial dentures

被引:49
作者
Wolfart, Stefan
Ludwig, Klaus
Uphaus, Anja
Kern, Matthias
机构
[1] Sch Dent, Dept Dent Mat, D-24105 Kiel, Germany
[2] Privat Praxis, D-46119 Oberhausen, Germany
[3] Univ Kiel, Sch Dent, Dept Prosthodont Propaedeut & Dent Mat, D-24105 Kiel, Germany
关键词
all-ceramic restoration; FPD; fracture load; zirconia ceramic; YPSZ; glass-ceramic; inlay-retained;
D O I
10.1016/j.dental.2006.12.006
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives. The purpose of this in vitro study was to compare the quasi-static (QSFS) and fatigue fracture strength of all-ceramic resin-bonded three unit inlay-retained fixed partial dentures (IRFPDs) made from a heat-pressed lithium-disilicate based glass-ceramic (LDGC) and a CAD/CAM-manufactured yttrium-oxide partially stabilized zirconia framework (YPSZ). Methods. Identical IRFPD-models of the maxilla were used. They consisted of a second premolar, a missing first molar and a second molar (Co-Cr-Mo alloy) integrated in a low melting alloy base. Roots were covered with a soft silicone-layer to simulate an artificial parodontium. Premolars had an occlusal-distal inlay-preparation and molars a mesial-occlusal inlay-preparation. Thirty-two IRFPDs were made from each ceramic using two different connector dimensions: 3 mm x 3 mm (LDGC_9, YPSZ_9); 4 mm x 4 mm (LDGC_16, YPSZ_16). All IRFPDs were cemented adhesively to the IRFPD-models, using composite resin cement. QSFS was tested in a universal-testing-machine (UTM) for six specimens. The other specimens were fatigued either with cyclic loading at 250 N (n = 4) in a chewing simulator or at 600 N (n = 6) using the UTM, and group YPSZ additionally at 1500 N. Results. The medians of QSFS (N) were 960 (LDGC_9), 1316 (LDGC_16), 3180 (YPSZ_9) and 3120 (YPSZ_16). For both QSFS and cyclic loading significant differences were found between LDGC_9 and LDGC_16 (p <= 0.03) but not between YPSZ_9 and YPSZ_16 (p > 0.05). Differences between LDGC and YPSZ were significant for both connector sizes (p <= 0.001). Significance. Considering the maximum chewing forces in the molar region it seems clinically possible, to use YPSZ as a core material for IRFPDs with a connector size between 9 and 16 mm(2). (c) 2007 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1513 / 1520
页数:8
相关论文
共 45 条
[1]   A NEW ALL-CERAMIC CROWN - A DENSE-SINTERED, HIGH-PURITY ALUMINA COPING WITH PORCELAIN [J].
ANDERSSON, M ;
ODEN, A .
ACTA ODONTOLOGICA SCANDINAVICA, 1993, 51 (01) :59-64
[2]  
Beham G., 1990, 6 IV VIV, P3
[3]   ENDODONTIC COMPLICATIONS FOLLOWING PERIODONTAL AND PROSTHETIC TREATMENT OF PATIENTS WITH ADVANCED PERIODONTAL-DISEASE [J].
BERGENHOLTZ, G ;
NYMAN, S .
JOURNAL OF PERIODONTOLOGY, 1984, 55 (02) :63-68
[4]  
Castellani D, 1994, Int J Prosthodont, V7, P149
[5]   MECHANICAL-PROPERTIES AND SHORT-TERM INVIVO EVALUATION OF YTTRIUM-OXIDE-PARTIALLY-STABILIZED ZIRCONIA [J].
CHRISTEL, P ;
MEUNIER, A ;
HELLER, M ;
TORRE, JP ;
PEILLE, CN .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1989, 23 (01) :45-61
[6]  
Edelhoff D, 2002, INT J PERIODONT REST, V22, P241
[7]  
Edelhoff D, 2001, QUINTESSENCE INT, V32, P269
[8]   DYNAMIC FATIGUE OF FELDSPATHIC PORCELAIN [J].
FAIRHURST, CW ;
LOCKWOOD, PE ;
RINGLE, RD ;
TWIGGS, SW .
DENTAL MATERIALS, 1993, 9 (04) :269-273
[9]  
Geis-Gerstdorfer J, 1999, DTSCH ZAHNARZTL Z, V54, P692
[10]   LIMITS OF HUMAN BITE STRENGTH [J].
GIBBS, CH ;
MAHAN, PE ;
MAUDERLI, A ;
LUNDEEN, HC ;
WALSH, EK .
JOURNAL OF PROSTHETIC DENTISTRY, 1986, 56 (02) :226-229