A comparative analysis between fixed bearing total knee arthroplasty (PFC Sigma) and rotating platform total knee arthroplasty (PFC-RP) with minimum 3-year follow-up

被引:13
作者
Jawed, Akram [1 ]
Kumar, Vijay [1 ]
Malhotra, R. [1 ]
Yadav, C. S. [1 ]
Bhan, S. [1 ]
机构
[1] All India Inst Med Sci, Dept Orthopaed, New Delhi 110029, India
关键词
Mobile bearing; Fixed bearing; Rotating platform; Knee score; RANDOMIZED CONTROLLED-TRIAL; MOBILE-BEARING; PROSTHESIS; REPLACEMENT; OUTCOMES; PATIENT; DESIGN; SYSTEM; METAANALYSIS; RATIONALE;
D O I
10.1007/s00402-012-1482-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Since the introduction of mobile bearing total knee designs nearly 30 years back, many studies have been done to evaluate its long-term result. Comparison with fixed bearing designs has been done in the past, but the studies were confounded by variables such as disease, surgeon, bone quality, pain tolerance, etc. We attempt to eliminate these variables in this study. A total of 50 patients who had bilateral arthritis of the knee with similar deformity and pre-operative range of motion on both sides agreed to have one knee replaced with mobile bearing total knee design (PFC-RP) and the other with a fixed bearing design (PFC Sigma) were prospectively evaluated. Comparative analysis of both the designs was done at a mean follow-up of 40 months, minimizing patient, surgeon and observer related bias. Clinical and radiographic outcome, survival and complication rates were compared. At a mean follow-up of 40 months (range 36-47 months), no benefit of mobile bearing (PFC-RP) over fixed bearing design (PFC Sigma) could be demonstrated with respect to Knee Society scores, pain scores, range of flexion, subject preference or patello-femoral complication rates. Radiographs showed no difference in prosthetic alignment. No patient required a revision surgery till last follow-up. Our study demonstrated no advantage of the mobile-bearing arthroplasty over fixed bearing arthroplasty with regard to clinical results at short-term follow-up. However, longer follow-up is necessary to confirm whether these results are sustained.
引用
收藏
页码:875 / 881
页数:7
相关论文
共 36 条
[1]   Results of rotating-platform, low-contact-stress knee prosthesis [J].
Bhan, S ;
Malhotra, R .
JOURNAL OF ARTHROPLASTY, 2003, 18 (08) :1016-1022
[2]   A comparison of fixed-bearing and mobile-bearing total knee arthroplasty at a minimum follow-up of 4.5 years [J].
Bhan, S ;
Malhotra, R ;
Kiran, EK ;
Shukla, S ;
Bijjawara, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (10) :2290-2296
[3]   NEW METHOD OF MEASURING PATELLAR HEIGHT [J].
BLACKBURNE, JS ;
PEEL, TE .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1977, 59 (02) :241-242
[4]   Less anterior knee pain with a mobile-bearing prosthesis compared with a fixed-bearing prosthesis [J].
Breugem, Stefan J. M. ;
Sierevelt, Inger N. ;
Schafroth, Matthias U. ;
Blankevoort, Leendert ;
Schaap, Gerard R. ;
van Dijk, C. Niek .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (08) :1959-1965
[5]  
BRYAN RS, 1982, CLIN ORTHOP RELAT R, P116
[6]   THE NEW-JERSEY LOW-CONTACT-STRESS KNEE REPLACEMENT SYSTEM - BIOMECHANICAL RATIONALE AND REVIEW OF THE 1ST 123 CEMENTED CASES [J].
BUECHEL, FF ;
PAPPAS, MJ .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1986, 105 (04) :197-204
[7]   Cemented rotating-platform total knee replacement - A nine to twelve-year follow-up study [J].
Callaghan, JJ ;
Squire, MW ;
Goetz, DD ;
Sullivan, PM ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (05) :705-711
[8]   PATIENT OUTCOMES FOLLOWING TRICOMPARTMENTAL TOTAL KNEE REPLACEMENT - A METAANALYSIS [J].
CALLAHAN, CM ;
DRAKE, BG ;
HECK, DA ;
DITTUS, RS .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 271 (17) :1349-1357
[9]   Total knee replacement in young, active patients - Long-term follow-up and functional outcome [J].
Diduch, DR ;
Insall, JN ;
Scott, WN ;
Scuderi, GR ;
FontRodriguez, D .