Proarrhythmic effect of pacemaker stimulation in patients with implanted cardioverter-defibrillators

被引:54
作者
Himmrich, E [1 ]
Przibille, O [1 ]
Zellerhoff, C [1 ]
Liebrich, A [1 ]
Rosocha, S [1 ]
Andreas, K [1 ]
Nebeling, D [1 ]
Omogbehin, B [1 ]
Meyer, J [1 ]
机构
[1] Univ Hosp, Med Clin 2, D-55131 Mainz, Germany
关键词
electrical stimulation; tachyarrhythmias; pacing; cardioverter-defibrillator;
D O I
10.1161/01.CIR.0000080291.65638.CC
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-We sought to determine the potential of right ventricular VVI backup pacing to induce ventricular tachyarrhythmias in patients with implanted cardioverter-defibrillators. Methods and Results-All consecutive patients presenting exclusively with pacemaker-induced tachycardias (PITs) were included in a prospective study using a crossover protocol. Patients were randomized to either group 1 (augmentation of the baseline frequency of the pacemaker to 60 bpm) or group 2 (pacemaker turned off) and were followed up for 1 year and then crossed over to the other programming, looking for reoccurrence of PIT. Of 150 consecutive patients, 39 (26%) had PIT, 13 of them exclusively (8.6%). Forty of 1063 analyzed tachyarrhythmias of all the patients were PIT (3%). Before inclusion in the study, the patients had 2.7 +/- 0.9 PITs in 11 +/- 6.5 months with their pacemakers programmed empirically at 42.3 bpm. During the study phase, no PIT occurred while the pacemaker was turned off, whereas programming to 60 bpm led to the recurrence of PIT in 5 of 6 patients (1.4 +/- 0.6 per patient). At the end of the study, 9 patients underwent a prolonged follow-up with their pacemakers turned off, resulting in spontaneous episodes of ventricular tachycardia/fibrillation in 5 patients, but PITs were no longer observed. Conclusions-This crossover protocol proves the potential proarrhythmic effect of pacemaker stimulation in implantable cardioverter-defibrillator patients. Resulting PITs led to clinical symptoms and antitachycardia therapy by the implantable cardioverter-defibrillator. Thus, in patients presenting with PIT but without a pacemaker indication, the pacemaker feature should be turned off, or, alternatively, the longest possible escape interval should be programmed.
引用
收藏
页码:192 / 197
页数:6
相关论文
共 16 条
[1]   CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALKINS, H ;
BRINKER, J ;
VELTRI, EP ;
GUARNIERI, T ;
LEVINE, JH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :666-673
[2]   PACED BEATS FOLLOWING SINGLE NONSENSED COMPLEXES IN A CODEPENDENT CARDIOVERTER DEFIBRILLATOR AND BRADYCARDIA PACING SYSTEM - POTENTIAL FOR VENTRICULAR-TACHYCARDIA INDUCTION [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (08) :1281-1287
[3]   TORSADE-DE-POINTES AND OTHER PAUSE-INDUCED VENTRICULAR TACHYCARDIAS - THE SHORT-LONG-SHORT SEQUENCE AND EARLY AFTERDEPOLARIZATIONS [J].
CRANEFIELD, PF ;
ARONSON, RS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (06) :670-678
[4]   FACILITATION OF VENTRICULAR-TACHYCARDIA INDUCTION WITH ABRUPT CHANGES IN VENTRICULAR CYCLE LENGTH [J].
DENKER, S ;
LEHMANN, M ;
MAHMUD, R ;
GILBERT, C ;
AKHTAR, M .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (04) :508-515
[5]   Pacemaker-mediated polymorphic ventricular tachycardia [J].
Goldman, DS ;
Levine, PA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10) :1993-1995
[6]  
GOMES JA, 1989, J AM COLL CARDIOL, V14, P1618
[7]   PACEMAKER-INDUCED VENTRICULAR-TACHYCARDIA IN NORMALLY FUNCTIONING VENTRICULAR DEMAND PACEMAKERS [J].
KARBENN, U ;
BORGGREFE, M ;
BREITHARDT, G .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (01) :120-122
[8]  
KAY GN, 1983, J AM COLL CARDIOL, V2, P806
[9]   SUDDEN BRADYARRHYTHMIC DEATH IN PATIENTS WITH THE IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - REPORT OF 2 CASES [J].
KHASTGIR, T ;
AARONS, D ;
VELTRI, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (03) :395-398
[10]  
LEFROY DC, 1994, BRIT HEART J, V71, P481