Effects of nitinol Strecker stent placement on vascular response in normal and stenotic porcine iliac arteries

被引:8
作者
Froelich, JJ
Alfke, H
Wilke, A
Ramaswamy, A
Barth, KH
Hoppe, M
Wagner, HJ
Klose, KJ
机构
[1] Philipps Univ Hosp, Dept Radiol, D-35043 Marburg, Germany
[2] Philipps Univ Hosp, Dept Orthoped Surg, D-35043 Marburg, Germany
[3] Philipps Univ Hosp, Dept Pathol, D-35043 Marburg, Germany
[4] Georgetown Univ Hosp, Div Vasc & Intervent Radiol, Washington, DC 20007 USA
关键词
arteries; iliac; stenosis or obstruction; stents and prostheses;
D O I
10.1016/S1051-0443(99)70039-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: This experimental study was conducted to evaluate neointimal thickness, lumen diameters, and histologic changes in normal and stenotic porcine iliac arteries following placement of self-expanding nitinol Strecker stents, MATERIALS AND METHODS: Neointimal trauma causing slight vascular stenosis was induced unilaterally within external iliac arteries of 12 swines by means of endothelial abrasion and high cholesterol diet. Nitinol Strecker stents were placed within the stenotic and the normal contralateral vascular segments. For histopathologic evaluation, the pigs were killed 12 or 24 weeks after stent placement and luminal diamters were evaluated angiographically, RESULTS: Excluding one occlusion, 15% narrowing of the lumen diameter was induced unilaterally (P = .002), Initial luminal gain after stent placement was greater for stenotic than for normal arteries. The amount of neointima thickness was not different between stenotic and normal vessels (P > .05), Comparing vascular diameters before stent placement and at follow-up, luminal loss due to neointima proliferation was 22% within normal arteries (P = .0002), while a luminal gain by 15% was found within the stenotic arteries (P = .008), Maturation of neointima and endothelial coverage were complete after 24 weeks. CONCLUSIONS: Even though nitinol Strecker stents induce excessive neointimal proliferation, stenotic arteries seem to profit from great early luminal gain resulting in 15% of vascular expansion at follow-up while slight stenosis is induced within normal iliac arteries.
引用
收藏
页码:329 / 338
页数:10
相关论文
共 30 条
[1]   Paired comparison of vascular wall reactions to Palmaz stents, Strecker tantalum stents, and Wallstents in canine iliac and femoral arteries [J].
Barth, KH ;
Virmani, R ;
Froelich, J ;
Takeda, T ;
Lossef, SV ;
Newsome, J ;
Jones, R ;
Lindisch, D .
CIRCULATION, 1996, 93 (12) :2161-2169
[2]   FLEXIBLE TANTALUM STENTS IMPLANTED IN AORTAS AND ILIAC ARTERIES - EFFECTS IN NORMAL CANINES [J].
BARTH, KH ;
VIRMANI, R ;
STRECKER, EP ;
SAVIN, MA ;
LINDISCH, D ;
MATSUMOTO, AH ;
TEITELBAUM, GP .
RADIOLOGY, 1990, 175 (01) :91-96
[3]   TREATMENT OF EXPERIMENTALLY INDUCED ATHEROSCLEROSIS IN SWINE ILIAC ARTERIES - A COMPARISON OF SELF-EXPANDING AND BALLOON-EXPANDED STENTS [J].
CHUAPETCHARASOPON, C ;
WRIGHT, KC ;
WALLACE, S ;
DOBBEN, RL ;
GIANTURCO, C .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1992, 15 (03) :143-150
[4]   PERCUTANEOUS ARTERIAL GRAFTING [J].
CRAGG, AH ;
LUND, G ;
RYSAVY, JA ;
SALOMONOWITZ, E ;
CASTANEDAZUNIGA, WR ;
AMPLATZ, K .
RADIOLOGY, 1984, 150 (01) :45-49
[5]   PLASMA TREATMENT EFFECTS ON THE TANTALUM STRECKER STENT IMPLANTED IN FEMORAL ARTERIES OF SHEEP [J].
CROCHET, D ;
GROSSETETE, R ;
BACHLIJOUR, B ;
SAGAN, C ;
LECOMTE, E ;
LEURENT, B ;
BRUNEL, P ;
LENIHOUANNEN, JC .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 17 (05) :285-291
[6]   MATCHING BASED ON QUANTITATIVE CORONARY ANGIOGRAPHY AS A SURROGATE FOR RANDOMIZED STUDIES - COMPARISON BETWEEN STENT IMPLANTATION AND BALLOON ANGIOPLASTY OF NATIVE CORONARY-ARTERY LESIONS [J].
DEJAEGERE, PP ;
HERMANS, WR ;
RENSING, BJ ;
STRAUSS, BH ;
DEFEYTER, PJ ;
SERRUYS, PW .
AMERICAN HEART JOURNAL, 1993, 125 (02) :310-319
[7]   TRANS-LUMINAL EXPANDABLE NITINOL COIL STENT GRAFTING - PRELIMINARY-REPORT [J].
DOTTER, CT ;
BUSCHMANN, RW ;
MCKINNEY, MK ;
ROSCH, J .
RADIOLOGY, 1983, 147 (01) :259-260
[8]   TRANSLUMINALLY-PLACED COILSPRING ENDARTERIAL TUBE GRAFTS - LONG-TERM PATENCY IN CANINE POPLITEAL ARTERY [J].
DOTTER, CT .
INVESTIGATIVE RADIOLOGY, 1969, 4 (05) :329-+
[9]   FLEXIBLE BALLOON-EXPANDED STENT FOR SMALL VESSELS - WORK IN PROGRESS [J].
DUPRAT, G ;
WRIGHT, KC ;
CHARNSANGAVEJ, C ;
WALLACE, S ;
GIANTURCO, C .
RADIOLOGY, 1987, 162 (01) :276-278
[10]   SELF-EXPANDING METALLIC STENTS FOR SMALL VESSELS - AN EXPERIMENTAL EVALUATION [J].
DUPRAT, G ;
WRIGHT, KC ;
CHARNSANGAVEJ, C ;
WALLACE, S ;
GIANTURCO, C .
RADIOLOGY, 1987, 162 (02) :469-472