Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures

被引:31
|
作者
Lever, James P. [2 ]
Aksenov, Sergei A. [2 ]
Zdero, Rad [1 ]
Ahn, Henry [2 ]
Mckee, Michael D. [2 ]
Schemitsch, Emil H. [1 ,2 ]
机构
[1] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[2] Univ Toronto, Dept Surg, Toronto, ON, Canada
关键词
biomechanics; proximal humerus; fracture; plate fixation;
D O I
10.1097/BOT.0b013e31815c89ce
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To biomechanically assess five proximal, humeral, fracture-plate-fixation systems. Methods: Surgical neck fractures, with and without cortical contact, were created in 25 fresh-frozen cadaveric humeri. Five methods of plate fixation were used for repair: construct A [an eight-hole, low contact dynamic compression (LCDC) plate contoured into a blade shape, supported by one, 70-mm-long, 4.5-mm-diameter cortical screw acting as a truss], construct B (a 10-hole LCDC plate arrangement identical to construct A, but using one, 70-mm-long, 3.5-mm-diameter cortical screw as a truss), construct C [a five-hole dynamic compression (DC) blade plate with one, 6.5-mm-diameter cancellous screw], construct D (a five-hole T-plate supported by three, 6.5-mm-diameter cancellous screws), and construct E (a five-hole cloverleaf plate supported by five, 4-mm-diameter cancellous screws). Plates were posterior to the bicipital groove, 10 mm distal to the greater tuberosity tip, on the lateral aspect of the humeral shaft. Screw fixation was done using standard AO compression plating techniques. Stiffness of constructs was measured in bending and axial compression. Locked plates were not assessed. Results: For cortical contact [abduction of 20 degrees (P = 0.02), flexion of 20 degrees (P = 0.02), flexion of 90 degrees (P = 0.005)] and no cortical contact [flexion of 90 degrees (P = 0.0001)], construct A was significantly stiffer than other constructs. For no cortical contact in abduction of 90 degrees (P = 0.05), construct A was significantly stiffer than other constructs. Conclusions: Construct A was significantly stiffer than other constructs.
引用
收藏
页码:23 / 29
页数:7
相关论文
共 50 条
  • [31] Biomechanical comparison of screw-based zones of a spatial subchondral support plate for proximal humerus fractures
    Jabran, Ali
    Peach, Chris
    Zou, Zhenmin
    Ren, Lei
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2019, 233 (03) : 372 - 382
  • [32] Biomechanical in vitro assessment of screw augmentation in locked plating of proximal humerus fractures
    Roederer, Goetz
    Scola, Alexander
    Schmoelz, Werner
    Gebhard, Florian
    Windolf, Markus
    Hofmann-Fliri, Ladina
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2013, 44 (10): : 1327 - 1332
  • [33] Double plate osteosynthesis provides better biomechanical stabilization than double tension band technique in distal humerus fractures
    Dogramaci, Yunus
    Esen, Erdinc
    Kurklu, Mustafa
    Kirici, Yalcin
    Atahan, Ali O.
    Komurcu, Mahmut
    EKLEM HASTALIKLARI VE CERRAHISI-JOINT DISEASES AND RELATED SURGERY, 2010, 21 (01): : 44 - 49
  • [34] Plate osteosynthesis versus hemiarthroplasty in proximal humerus fractures - Does routine screening of systemic inflammatory biomarkers makes sense?
    Horst, Klemens
    Hildebrand, Frank
    Pfeifer, Roman
    Koeppen, Karin
    Lichte, Philipp
    Pape, Hans-Christoph
    Dienstknecht, Thomas
    EUROPEAN JOURNAL OF MEDICAL RESEARCH, 2015, 20
  • [35] Mid-term follow-up evaluation of plate osteosynthesis and hemiarthroplasty results in multipart fractures of the proximal humerus
    Celik, Cem
    Gumustas, Seyit Ali
    Cecen, Gultekin Sitki
    Bulut, Guven
    Bekler, Halil Ibrahim
    ULUSAL TRAVMA VE ACIL CERRAHI DERGISI-TURKISH JOURNAL OF TRAUMA & EMERGENCY SURGERY, 2016, 22 (04): : 379 - 385
  • [36] Hemiarthroplasty for proximal humerus fractures
    Chambers L.
    Dines J.S.
    Lorich D.G.
    Dines D.M.
    Current Reviews in Musculoskeletal Medicine, 2013, 6 (1) : 57 - 62
  • [37] Plate osteosynthesis versus hemiarthroplasty in proximal humerus fractures – Does routine screening of systemic inflammatory biomarkers makes sense?
    Klemens Horst
    Frank Hildebrand
    Roman Pfeifer
    Karin Köppen
    Philipp Lichte
    Hans-Christoph Pape
    Thomas Dienstknecht
    European Journal of Medical Research, 20
  • [38] Can the osteosynthesis with locking plates be a better treatment for unstable fractures of the proximal humerus?
    Verdano M.A.
    Lunini E.
    Pellegrini A.
    Corsini T.
    Marenghi P.
    Ceccarelli F.
    MUSCULOSKELETAL SURGERY, 2014, 98 (1) : 27 - 33
  • [39] Trends in the Characterization of the Proximal Humerus in Biomechanical Studies: A Review
    Castro-Franco, Angel D.
    Mendoza-Munoz, Ismael
    Gonzalez-Angeles, Alvaro
    Cruz-Sotelo, Samantha E.
    Castaneda, Ana Maria
    Siqueiros-Hernandez, Miriam
    APPLIED SCIENCES-BASEL, 2020, 10 (18):
  • [40] Open reduction and plate/screw osteosynthesis of proximal phalanx fractures
    Unglaub, F.
    Langer, M. F.
    Loew, S.
    Hohendorff, B.
    Spies, C. K.
    OPERATIVE ORTHOPADIE UND TRAUMATOLOGIE, 2019, 31 (05): : 408 - 421