Explantation of the Novel Ahmed Glaucoma Valve M4 Implant
被引:4
|作者:
Hu, Wanda D.
[1
]
Pro, Michael J.
[1
]
Fudemberg, Scott J.
[1
]
Moster, Marlene R.
[1
]
机构:
[1] Wills Eye Hosp & Res Inst, Philadelphia, PA 19107 USA
关键词:
glaucoma;
drainage device;
tube shunt;
polyethylene;
Ahmed Glaucoma Valve;
explantation;
CLINICAL-EXPERIENCE;
DRAINAGE DEVICES;
COMPLICATIONS;
OUTCOMES;
D O I:
10.1097/IJG.0000000000000192
中图分类号:
R77 [眼科学];
学科分类号:
100212 ;
摘要:
Purpose: To report a series of cases involving Ahmed Glaucoma Valve M4 (AGV) explantation and to discuss the surgical technique to remove the drainage device. Methods: Four cases were identified that presented with AGV M4 postoperative complications necessitating tube shunt removal. Three patients presented with conjunctival erosion and 1 patient with persistent diplopia. AGV M4 implants were removed successfully between 1.5 and 9 months after implantation. Results: Successful explantation of the AGV M4 novel implant was achieved in all cases without intraoperative or postoperative complications. Conclusions: If necessary, AGV M4 explantation can be successfully performed in the early postoperative period.