Rotational stability of a single-piece toric acrylic intraocular lens

被引:82
作者
Koshy, John J. [1 ]
Nishi, Yutaro [1 ]
Hirnschall, Nino [1 ,2 ]
Crnej, Alja [1 ]
Gangwani, Vinod [1 ]
Maurino, Vincenzo [1 ]
Findl, Oliver [1 ,2 ]
机构
[1] Moorfields Eye Hosp NHS Fdn Trust, London, England
[2] Hanusch Hosp, VIROS, Karl Landsteiner Inst, Dept Ophthalmol, Vienna, Austria
关键词
CATARACT-SURGERY; ASTIGMATISM;
D O I
10.1016/j.jcrs.2010.05.018
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To assess the rotational stability of a single-piece toric hydrophobic acrylic intraocular lens (IOL) during the first 6 postoperative months. SETTING: Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom. DESIGN: Prospective case series. METHOD: Eyes with age-related cataract and corneal astigmatism (1.00 to 3.00 D diopter [D]) were measured by partial coherence interferometry (IOLMaster). Preoperatively, the horizontal axis was marked. Surgical technique included a standardized temporal clear corneal incision, phacoemulsification, and in-the-bag implantation of a toric IOL (AcrySof SN60TT). Rotational stability of the IOL was assessed using retroillumination photographs postoperatively at 1 hour and at 1 and 6 months. Uncorrected (UDVA) and corrected distance visual acuities, residual refractive error, and keratometric and refractive cylinders were measured. RESULTS: The study enrolled 30 eyes (30 patients). The mean absolute IOL rotation was 2.44 degrees +/- 1.84 (SD) at 1 month and 2.66 +/- 1.99 degrees at 6 months. The photographic technique had high reproducibility of axis measurement, with consecutive measurements varying by less than 2.0 degrees. The mean UDVA was 0.16 logMAR (range 0.42 to -0.18 logMAR) at 1 month and 0.20 logMAR (range 0.60 to -0.20 logMAR) at 6 months. The residual refractive cylinder by autorefraction was -0.84 D (range -2.00 to 0.00 D) and -0.80 D (range -1.75 to 0.00 D), respectively. CONCLUSIONS: The acrylic toric IOL was rotationally stable within the first 6 months postoperatively. The photographic and axis analysis method to evaluate stability had high reproducibility and detected small changes in rotation.
引用
收藏
页码:1665 / 1670
页数:6
相关论文
共 17 条
[1]  
B. Alcon Inc., ACRYSOF TOR IOL WEB
[2]   Astigmatism management in cataract surgery with the AcrySof toric intraocular lens [J].
Bauer, Noel J. C. ;
de Vries, Niels E. ;
Webers, Carroll A. B. ;
Hendrikse, Fred ;
Nuijts, Rudy M. M. A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (09) :1483-1488
[3]   Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses [J].
Chang, David F. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (11) :1842-1847
[4]   Early rotational stability of the longer Staar toric intraocular lens - Fifty consecutive cases [J].
Chang, DF .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (05) :935-940
[5]   Early experience with the AcrySof toric IOL for the correction of astigmatism in cataract surgery [J].
Dardzhikova, Albena ;
Shah, Chirag R. ;
Gimbel, Howard V. .
CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2009, 44 (03) :269-273
[6]   Evaluation of a toric intraocular lens with a Z-haptic [J].
De Silva, Don Julian ;
Ramkissoon, Yashin D. ;
Bloom, Philip A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (09) :1492-1498
[7]  
Dong E Y, 2001, Korean J Ophthalmol, V15, P22
[8]   Comparison of the biomechanical behavior of foldable intraocular lenses [J].
Lane, SS ;
Burgi, P ;
Millos, GS ;
Orchowski, MW ;
Vaughan, M ;
Schwarte, E .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (11) :2397-2402
[9]  
Lane SS, 2006, CATARACT REFRACT MAY, P66
[10]   Digital overlay technique for documenting toric intraocular lens axis orientation [J].
Nguyen, TM ;
Miller, KM .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (10) :1496-1504