High Failure at a Minimum 5-Year Follow-Up in Primary Total Hip Arthroplasty Using a Modular Femoral Trunnion

被引:4
作者
Nahhas, Cindy R. [1 ]
Yi, Paul H. [2 ]
Moric, Mario [1 ]
Puri, Rajeev [3 ]
Jacobs, Joshua J. [1 ]
Sporer, Scott M. [1 ]
机构
[1] Rush Univ, Med Ctr, Dept Orthoped Surg, Chicago, IL 60612 USA
[2] Johns Hopkins Univ, Sch Med, Dept Radiol, Baltimore, MD 21205 USA
[3] Riverside Hosp, Dept Orthoped Surg, Kankakee, IL USA
关键词
primary total hip arthroplasty; modular neck; modularity; cementless; survivorship; STEM; COMPONENT; HEAD; DISSOCIATION; INTERFACES; CORROSION; OUTCOMES;
D O I
10.1016/j.arth.2019.03.033
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Modular femoral trunnions enable the surgeon to independently adjust offset, leg length, and anteversion in total hip arthroplasty (THA). However, modularity may result in an increased risk of fretting and corrosion along with a higher risk of implant dissociation or fracture. The purpose of this study is to evaluate mid-term survivorship of THAs using a cementless modular system. Methods: A consecutive series of 221 patients who underwent a primary THA using the ALFA II modular stem by a single surgeon between 2002 and 2004 were reviewed. Survivorship of the ALFA II modular hip system was evaluated at a minimum of 5 years postoperatively. Results: Of the 221 patients, 28 (12.7%) died from causes unrelated to the surgery before adequate follow-up, and 64 (29.0%) patients were lost to follow-up. The remaining 129 patients had a mean 6.5-year (range: 5-8 years) follow-up. All-cause survivorship of the modular stem system was 81% (95% confidence interval = 69-90) at a mean 6.5-year follow-up. Of the 25 (19.4%) cases requiring revision surgery, 52.0% was for dissociation of the modular components, 32.0% was for fracture of the prosthesis, 12.0% was for instability/multiple dislocations, and 4.0% was for chronic septic THA. Body mass index (odds ratio = 1.080) and offset (odds ratio = 1.254) were independent risk factors for mechanical failures of the modular stem system. Conclusion: The modular stem hip system of interest in this study demonstrates a high failure rate at mid-term follow-up, and we caution against the use of similar designs in primary THAs. Published by Elsevier Inc.
引用
收藏
页码:1395 / 1399
页数:5
相关论文
共 24 条
[11]   What Is the Difference Between Modular and Nonmodular Tapered Fluted Titanium Stems in Revision Total Hip Arthroplasty [J].
Huang, Yong ;
Zhou, Yixin ;
Shao, Hongyi ;
Gu, Jianming ;
Tang, Hao ;
Tang, Qiheng .
JOURNAL OF ARTHROPLASTY, 2017, 32 (10) :3108-3113
[12]   Total hip arthroplasty using an omniflex modular system: 5 to 12 years followup [J].
Ito, H ;
Matsuno, T ;
Aok, Y ;
Minami, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (419) :98-106
[13]   Late detachment modular femoral component after primary total hip replacement [J].
Karaismailoglu, TN ;
Tomak, Y ;
Gulman, B .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2001, 121 (08) :481-482
[14]   White blood cell counts and differential in synovial fluid of aseptically failed total knee arthroplasty [J].
Kersey, R ;
Benjamin, J ;
Marson, B .
JOURNAL OF ARTHROPLASTY, 2000, 15 (03) :301-304
[15]   Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases [J].
Kop, Alan M. ;
Swarts, Eric .
JOURNAL OF ARTHROPLASTY, 2009, 24 (07) :1019-1023
[16]  
McKellop H A, 1992, J Arthroplasty, V7, P291, DOI 10.1016/0883-5403(92)90051-Q
[17]  
Namba RS, 2000, ORTHOPEDICS, V23, P489
[18]   Midterm Results of a Femoral Stem With a Modular Neck Design: Clinical Outcomes and Metal Ion Analysis [J].
Silverton, Craig D. ;
Jacobs, Joshua J. ;
Devitt, Jeffrey W. ;
Cooper, H. John .
JOURNAL OF ARTHROPLASTY, 2014, 29 (09) :1768-1773
[19]   Revision Hip Arthroplasty Using a Modular, Cementless Femoral Stem: Intermediate-Term Follow-Up [J].
Sivananthan, Sureshan ;
Lim, Chin-Tat ;
Narkbunnam, Rapeepat ;
Sox-Harris, Alex ;
Huddleston, James I., III ;
Goodman, Stuart B. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (04) :1245-1249
[20]   Midterm Outcomes of Revision Total Hip Arthroplasty Using a Modular Revision Hip System [J].
Smith, Marie Anne ;
Deakin, Angela H. ;
Allen, David ;
Baines, Joe .
JOURNAL OF ARTHROPLASTY, 2016, 31 (02) :446-450