The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements

被引:111
作者
Raylman, Raymond R. [1 ]
Majewski, Stan [2 ]
Smith, Mark F. [2 ,3 ]
Proffitt, James [2 ]
Hammond, William [2 ]
Srinivasan, Amarnath [1 ]
McKisson, John [2 ]
Popov, Vladimir [2 ]
Weisenberger, Andrew [2 ]
Judy, Clifford O. [4 ]
Kross, Brian [2 ]
Ramasubramanian, Srikanth [1 ]
Banta, Larry E. [4 ]
Kinahan, Paul E. [5 ]
Champley, Kyle [5 ]
机构
[1] W Virginia Univ, Dept Radiol, Ctr Adv Imaging, Morgantown, WV 26506 USA
[2] Thomas Jefferson Natl Accelerator Facility, Newport News, VA USA
[3] Univ Maryland, Sch Med, Dept Radiol, Baltimore, MD 21201 USA
[4] W Virginia Univ, Dept Mech & Aerosp Engn, Morgantown, WV 26506 USA
[5] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
关键词
D O I
10.1088/0031-9155/53/3/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Tomographic breast imaging techniques can potentially improve detection and diagnosis of cancer in women with radiodense and/or fibrocystic breasts. We have developed a high-resolution positron emission mammography/tomography imaging and biopsy device (called PEM/PET) to detect and guide the biopsy of suspicious breast lesions. PET images are acquired to detect suspicious focal uptake of the radiotracer and guide biopsy of the area. Limited-angle PEM images could then be used to verify the biopsy needle position prior to tissue sampling. The PEM/PET scanner consists of two sets of rotating planar detector heads. Each detector consists of a 4 x 3 array of Hamamatsu H8500 flat panel position sensitive photomultipliers (PSPMTs) coupled to a 96 x 72 array of 2 x 2 x 15 mm(3) LYSO detector elements (pitch = 2.1 mm). Image reconstruction is performed with a three-dimensional, ordered set expectation maximization (OSEM) algorithm parallelized to run on a multiprocessor computer system. The reconstructed field of view (FOV) is 15 x 15 x 15 cm(3). Initial phantom-based testing of the device is focusing upon its PET imaging capabilities. Specifically, spatial resolution and detection sensitivity were assessed. The results from these measurements yielded a spatial resolution at the center of the FOV of 2.01 +/- 0.09 mm (radial), 2.04 +/- 0.08 mm (tangential) and 1.84 +/- 0.07 mm (axial). At a radius of 7 cm from the center of the scanner, the results were 2.11 +/- 0.08 mm (radial), 2.16 +/- 0.07 mm (tangential) and 1.87 +/- 0.08 mm (axial). Maximum system detection sensitivity of the scanner is 488.9 kcps mu Ci(-1) ml(-1) (6.88%). These promising findings indicate that PEM/PET may be an effective system for the detection and diagnosis of breast cancer.
引用
收藏
页码:637 / 653
页数:17
相关论文
共 67 条
[1]   Design and evaluation of the Clear-PEM scanner for positron emission mammography [J].
Abreu, MC ;
Aguiar, JD ;
Almeida, FG ;
Almeida, P ;
Bento, P ;
Carrico, B ;
Ferreira, M ;
Ferreira, NC ;
Gonçalves, F ;
Leong, C ;
Lopes, F ;
Lousa, P ;
Martins, MV ;
Matela, N ;
Mendes, PR ;
Moura, R ;
Nobre, J ;
Oliveira, N ;
Ortigao, C ;
Peralta, L ;
Pereira, R ;
Rego, J ;
Ribeiro, R ;
Rodrigues, P ;
Sampaio, J ;
Santos, AI ;
Silva, L ;
Silva, JC ;
Sousa, P ;
Teixeira, IC ;
Teixeira, JP ;
Trindade, A ;
Varela, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) :71-77
[2]   Clear-PEM: A dedicated pet camera for improved breast cancer detection [J].
Abreu, MC ;
Almeida, P ;
Balau, F ;
Ferreira, NC ;
Fetal, S ;
Fraga, F ;
Martins, M ;
Matela, N ;
Moura, R ;
Ortigao, C ;
Peralta, L ;
Rato, P ;
Ribeiro, R ;
Rodrigues, P ;
Santos, AI ;
Trindade, A ;
Varela, J .
RADIATION PROTECTION DOSIMETRY, 2005, 116 (1-4) :208-210
[3]   EVALUATION OF BREAST MASSES AND AXILLARY LYMPH-NODES WITH [F-18] 2-DEOXY-2-FLUORO-D-GLUCOSE PET [J].
ADLER, LP ;
CROWE, JP ;
ALKAISI, NK ;
SUNSHINE, JL .
RADIOLOGY, 1993, 187 (03) :743-750
[4]  
Aloj L, 1999, CANCER RES, V59, P4709
[5]  
[Anonymous], P 9 INT M FULL 3 DIM
[6]  
Avril N, 1997, J NUCL MED, V38, P1186
[7]   Metabolic characterization of breast tumors with positron emission tomography using F-18 fluorodeoxyglucose [J].
Avril, N ;
Dose, J ;
Janicke, F ;
Bense, S ;
Ziegler, S ;
Laubenbacher, C ;
Romer, W ;
Pache, H ;
Herz, M ;
Allgayer, B ;
Nathrath, W ;
Graeff, H ;
Schwaiger, M .
JOURNAL OF CLINICAL ONCOLOGY, 1996, 14 (06) :1848-1857
[8]  
Avril N, 2001, J NUCL MED, V42, P9
[9]   Scintimammography with 99mTc-MIBI and magnetic resonance imaging in the evaluation of breast cancer [J].
Bagni, B ;
Franceschetto, A ;
Casolo, A ;
De Santis, M ;
Bagni, I ;
Pansini, F ;
Di Leo, C .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (10) :1383-1388
[10]   High-resolution fluorodeoxyglucose positron emission tomography with compression ("positron emission mammography") is highly accurate in depicting primary breast cancer [J].
Berg, Wendie A. ;
Weinberg, Irving N. ;
Narayanan, Deepa ;
Lobrano, Mary E. ;
Ross, Eric ;
Amodei, Laura ;
Tafra, Lorraine ;
Adler, Lee P. ;
Uddo, Joseph ;
Stein, William, III ;
Levine, Edward A. .
BREAST JOURNAL, 2006, 12 (04) :309-323