Application of the MTSVD unfolding method for reconstruction of primary X-ray spectra using semiconductor detectors

被引:7
作者
Gallardo, S. [1 ]
Rodenas, J. [1 ]
Querol, A. [1 ]
Verdu, G. [1 ]
机构
[1] Univ Politecn Valencia, Dept Ingn Quim & Nucl, Valencia 46022, Spain
关键词
X-ray; Unfolding; Semiconductor detectors; Quality control; MONTE-CARLO-SIMULATION; COMPTON SPECTROSCOPY; ENERGY-RANGE; SPECTROMETER; RESOLUTION;
D O I
10.1016/j.pnucene.2011.06.013
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A thorough knowledge of the primary spectrum is very important to perform a quality control (QC) of X-ray tubes. In previous works, a methodology to assess primary spectrum using a Compton spectrometer and applying the Monte Carlo (MC) method has been analyzed. By means of a Monte Carlo model, a response matrix can be built, relating the Pulse Height Distribution (PHD) recorded in the detector to the primary X-ray spectrum. Subsequently, an unfolding method based on the application of a Modified Truncated Singular Value Decomposition (MTSVD) is applied to the response matrix to assess the primary spectrum. Germanium (Ge), Silicon (Si) and Cadmium-Telluride (CdTe) detectors are considered in this work. The main goal of the paper is to determine whether the MTSVD unfolding method is adequate to provide an acceptable reproduction of spectra for these detectors. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1136 / 1139
页数:4
相关论文
共 12 条
  • [1] Cranley K., 1997, CATALOGUE DIAGNOSTIC
  • [2] FORSYTHE E, 1976, COMPUTER METHODS MAT
  • [3] Analysis of shielding materials in a Compton spectrometer applied to X-ray tube quality control using Monte Carlo simulation
    Gallardo, S
    Ródenas, J
    Verdú, G
    Villaescusa, JI
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) : 375 - 379
  • [4] X-ray spectrum unfolding using a regularized truncated SVD method
    Gallardo, S
    Ginestar, D
    Verdú, G
    Ródenas, J
    Puchades, V
    Villaescusa, JI
    [J]. X-RAY SPECTROMETRY, 2006, 35 (01) : 63 - 70
  • [5] Monte Carlo simulation of the Compton scattering technique applied to characterize diagnostic x-ray spectra
    Gallardo, S
    Ródenas, J
    Verdú, G
    [J]. MEDICAL PHYSICS, 2004, 31 (07) : 2082 - 2090
  • [6] Golub G. H., 1996, MATRIX COMPUTATIONS
  • [7] Knoll GF., 1989, Radiation detection and measurement
  • [8] Compton-scattering measurement of diagnostic x-ray spectrum using high-resolution Schottky CdTe detector
    Maeda, K
    Matsumoto, M
    Taniguchi, A
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1542 - 1547
  • [9] COMPTON SPECTROSCOPY IN THE DIAGNOSTIC-X-RAY ENERGY-RANGE .1. SPECTROMETER DESIGN
    MATSCHEKO, G
    CARLSSON, GA
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (02) : 185 - 197
  • [10] COMPTON SPECTROSCOPY IN THE DIAGNOSTIC-X-RAY ENERGY-RANGE .2. EFFECTS OF SCATTERING MATERIAL AND ENERGY RESOLUTION
    MATSCHEKO, G
    CARLSSON, GA
    RIBBERFORS, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (02) : 199 - 208