Computer simulations of an irrigated radiofrequency cardiac ablation catheter and experimental validation by infrared imaging

被引:5
作者
Rossmann, Christian [1 ,2 ]
Motamarry, Anjan [1 ,3 ,4 ]
Panescu, Dorin [5 ]
Haemmerich, Dieter [1 ,6 ]
机构
[1] Med Univ South Carolina, Dept Pediat, Charleston, SC 29425 USA
[2] AdditiveLab, Leuven, Belgium
[3] Med Univ South Carolina, Dept Drug Discovery & Biomed Sci, Charleston, SC 29425 USA
[4] Harvard Univ, Wellman Ctr Photomed, Massachusetts Gen Hosp, Boston, MA 02115 USA
[5] Biotronik, Santa Clara, CA USA
[6] Clemson Univ, Dept Bioengn, Clemson, SC 29634 USA
基金
美国国家卫生研究院;
关键词
Radiofrequency; microwave; thermal ablation; cardiac catheter ablation; modeling (i; e; heat transfer; ultrasound; EM; integrated; treatment planning); PULSATILE BLOOD-FLOW; DIELECTRIC-PROPERTIES; THERMAL-CONDUCTIVITY; BIOLOGICAL TISSUES; LESION SIZE; BIOPHYSICS; ESTIMATOR; LOCATIONS; ISOTHERM; MODEL;
D O I
10.1080/02656736.2021.1961027
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose To develop and validate a three-dimensional (3-D) computer model based on accurate geometry of an irrigated cardiac radiofrequency (RF) ablation catheter with microwave radiometry capability, and to test catheter performance. Methods A computer model was developed based on CAD geometry of a RF cardiac ablation catheter prototype to simulate electromagnetic heating, heat transfer, and computational fluid dynamics (blood flow, open irrigation, and natural convection). Parametric studies were performed; blood flow velocity (0-25 cm/s) and irrigation flow (0-40 ml/min) varied, both with perpendicular (PE) and parallel (PA) catheter orientations relative to tissue. Tissue Agar phantom studies were performed under similar conditions, and temperature maps were recorded via infrared camera. Computer model simulations were performed with constant voltage and with voltage adjusted to achieve maximum tissue temperatures of 95-105 degrees C. Results Model predicted thermal lesion width at 5 W power was 5.8-6.4 mm (PE)/6.5-6.6 mm (PA), and lesion depth was 4.0-4.3 mm (PE)/4.0-4.1 mm (PA). Compared to phantom studies, the mean errors of the computer model were as follows: 6.2 degrees C(PE)/4.3 degrees C (PA) for maximum gel temperature, 0.7 mm (10.9%) (PE)/0.1 mm (0.8%) (PA) for lesion width, and 0.3 mm (7.7%)(PE)/0.7 mm (19.1%) (PA) for lesion depth. For temperature-controlled ablation, model predicted thermal lesion width was 7-9.2 mm (PE)/8.6-9.2 mm (PA), and lesion depth was 4.3-5.5 mm (PE)/3.4-5.4 mm (PA). Conclusions Computer models were able to reproduce device performance and to enable device evaluation under varying conditions. Temperature controlled ablation of irrigated catheters enables optimal tissue temperatures independent of patient-specific conditions such as blood flow.
引用
收藏
页码:1149 / 1163
页数:15
相关论文
共 50 条
[31]   Effects of Open-Irrigated Radiofrequency Ablation Catheter Design on Lesion Formation and Complications: In Vitro Comparison of 6 Different Devices [J].
Guerra, Jose M. ;
Jorge, Esther ;
Raga, Silvia ;
Galvez-Monton, Carolina ;
Alonso-Martin, Concepcion ;
Rodriguez-Font, Enrique ;
Cinca, Juan ;
Vinolas, Xavier .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (10) :1157-1162
[32]   Near-Field Ultrasound Imaging During Radiofrequency Catheter Ablation Tissue Thickness and Epicardial Wall Visualization and Assessment of Radiofrequency Ablation Lesion Formation and Depth [J].
Haines, David E. ;
Wright, Matthew ;
Harks, Erik ;
Deladi, Szabolcs ;
Fokkenrood, Steven ;
Brink, Rob ;
Belt, Harm ;
Kolen, Alexander F. ;
Mihajlovic, Nenad ;
Zuo, Fei ;
Rankin, Darrell ;
Stoffregen, William ;
Cockayne, Debra ;
Cefalu, Joseph .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (12)
[33]   Should fluid dynamics be included in computer models of RF cardiac ablation by irrigated-tip electrodes? [J].
Gonzalez-Suarez, Ana ;
Perez, Juan J. ;
Berjano, Enrique .
BIOMEDICAL ENGINEERING ONLINE, 2018, 17
[34]   Near-Field Ultrasound Imaging During Radiofrequency Catheter Ablation Lessons for Today and Hope for Tomorrow [J].
Richardson, Travis ;
Michaud, Gregory F. .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (08) :1073-1074
[35]   Long-Term Success of Irrigated Radiofrequency Catheter Ablation of Sustained Ventricular Tachycardia Post-Approval THERMOCOOL VT Trial [J].
Marchlinski, Francis E. ;
Haffajee, Charles I. ;
Beshai, John F. ;
Dickfeld, Timm-Michael L. ;
Gonzalez, Mario D. ;
Hsia, Henry H. ;
Schuger, Claudio D. ;
Beckman, Karen J. ;
Bogun, Frank M. ;
Pollak, Scott J. ;
Bhandari, Anil K. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (06) :674-683
[36]   Evaluation of biochemical markers of myocardial lesion after radiofrequency catheter ablation. Usefulness of cardiac troponin I [J].
delRey, JM ;
Madrid, AH ;
Novo, L ;
Sanchez, A ;
Martin, J ;
Rubi, J ;
Pallares, E ;
Kais, J ;
Villa, BG ;
Manzano, JG ;
Silvestre, I ;
Jimenez, A ;
Ripoll, E ;
Moro, C .
REVISTA ESPANOLA DE CARDIOLOGIA, 1997, 50 (08) :552-560
[37]   Radiofrequency ablation using a novel insulated-tip ablation catheter can create uniform lesions comparable in size to conventional irrigated ablation catheters while using a fraction of the energy and irrigation [J].
Aryana, Arash ;
Irastorza, Ramiro M. ;
Berjano, Enrique ;
Cohen, Richard J. ;
Kraus, Jeffrey ;
Haghighi-Mood, Ali ;
Reddy, Vivek Y. ;
d'Avila, Andre .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (06) :1146-1156
[38]   Characteristics of Radiofrequency Catheter Ablation Lesion Formation in Real Time In Vivo Using Near Field Ultrasound Imaging [J].
Wright, Matthew ;
Harks, Erik ;
Deladi, Szabolcs ;
Fokkenrood, Steven ;
Brink, Rob ;
Belt, Harm ;
Kolen, Alexander F. ;
Rankin, Darrell ;
Stoffregen, William ;
Cockayne, Debra A. ;
Cefalu, Joseph ;
Haines, David E. .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (08) :1062-1072
[39]   The Disappearance of 'Double Peak Sign' With Tissue Doppler Imaging After Radiofrequency Catheter Ablation in Patients With Atrial Fibrillation [J].
Fujii, Akira ;
Inoue, Katsuji ;
Nagai, Takayuki ;
Nishimura, Kazuhisa ;
Iio, Chiharuko ;
Fujimoto, Kaori ;
Kohno, Tamami ;
Uetani, Teruyoshi ;
Suzuki, Jun ;
Funada, Jun-ichi ;
Okura, Takafumi ;
Higaki, Jitsuo ;
Ogimoto, Akiyoshi .
CIRCULATION, 2013, 128 (22)
[40]   Computational Modeling of Open-Irrigated Electrodes for Radiofrequency Cardiac Ablation Including Blood Motion-Saline Flow Interaction [J].
Gonzalez-Suarez, Ana ;
Berjano, Enrique ;
Guerra, Jose M. ;
Gerardo-Giorda, Luca .
PLOS ONE, 2016, 11 (03)