Near-Infrared Spectra in Buccal Tissue as a Marker for Detection of Hypoxia

被引:3
作者
Amini, Maryam
Hisdal, Jonny
Gjovaag, Terje
Kapetanovic, Nihad
Strand, Trond-Eirik
Owe, Jan Ove
Horthe, John Ragnar
Mirtaheri, Peyman
机构
[1] Oslo & Akershus Univ Coll Appl Sci, Oslo, Norway
[2] Norwegian Armed Forces Med Serv, Inst Aviat Med, Oslo, Norway
[3] Oslo Univ Hosp, Oslo, Norway
关键词
hypobaric chamber; near infrared spectroscopy; synchronization; multivariate calibration; BLOOD-FLOW; OXYGEN-SATURATION; SPECTROSCOPY; CONSUMPTION; PRESSURE; OXIMETRY;
D O I
10.3357/AMHP.4510.2016
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
INTRODUCTION: Hypoxia caused by high altitude exposure can impair cerebral and mental functions. Blood flow and oxygenation of the buccal tissue can be reliable markers to detect hypoxia. In this study, near infrared spectroscopy was used in combination with a novel optical probe to evaluate the applicability of the novel probe in measuring hypoxia markers in buccal tissue under a hypoxic condition. METHODS: Six healthy participants were tested at altitudes from 2000 to 16,000 ft inside a hypobaric chamber. The buccal reference measurements of blood flow and oxygen saturation were synchronized with the spectral measurements of the novel near infrared probe and the relationship between the reference measurements and spectral data were evaluated by multivariate partial least square method. In addition, finger oxygen saturation was measured during the experiment and the recordings were compared with buccal oxygen saturation. RESULTS: The spectral analysis illustrated that the spectral data from the near infrared probe correlated strongly with the absorption features of both buccal flow and oxygenation measured by the reflectance sensors (average R-2 = 0.89).The results showed probably overestimated values for buccal oxygen saturation recorded by the reference pulse oximeter in comparison with finger oxygen saturation, with the mean difference increasing from 1.8% at 2000 ft to 11.4% at 16,000 ft. CONCLUSION: The novel near infrared probe showed promising results for simultaneous measurement of blood flow and oxygen saturation in the buccal tissue. The suggested method can be used as a new technique for early indication of hypoxia in future clinical applications.
引用
收藏
页码:498 / 504
页数:7
相关论文
共 22 条
[1]   Novel Design of an Optical Probe for Detecting Perfusion Changes in Buccal Tissue [J].
Amini, Maryam ;
Gjovaag, Terje F. ;
Hisdal, Jonny ;
Mirtaheri, Peyman .
IEEE SENSORS JOURNAL, 2012, 12 (06) :1861-1867
[2]   Blood flow and oxygen consumption with near-infrared spectroscopy and venous occlusion: spatial maps and the effect of time and pressure of inflation [J].
Casavola, C ;
Paunescu, LA ;
Fantini, S ;
Gratton, E .
JOURNAL OF BIOMEDICAL OPTICS, 2000, 5 (03) :269-276
[3]   Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation A Comparison of Three Cerebral Oximetty Technologies [J].
Davie, Sophie N. ;
Grocott, Hilary P. .
ANESTHESIOLOGY, 2012, 116 (04) :834-840
[4]   Accuracy and precision of buccal pulse oximetry [J].
De Jong, Marla J. ;
Schmelz, Joseph ;
Evers, Karen ;
Bradshaw, Patricia ;
McKnight, Kathleen ;
Bridges, Elizabeth .
HEART & LUNG, 2011, 40 (01) :31-40
[5]   NONINVASIVE MEASUREMENT OF FOREARM BLOOD-FLOW AND OXYGEN-CONSUMPTION BY NEAR-INFRARED SPECTROSCOPY [J].
DEBLASI, RA ;
FERRARI, M ;
NATALI, A ;
CONTI, G ;
MEGA, A ;
GASPARETTO, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1994, 76 (03) :1388-1393
[6]   COMPARISON OF PULSE OXIMETERS - ACCURACY AT LOW ARTERIAL-PRESSURE IN VOLUNTEERS [J].
FALCONER, RJ ;
ROBINSON, BJ .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (04) :552-557
[7]  
GUNTER JB, 1989, ANESTH ANALG, V69, P417
[8]  
Kamat V., 2002, INDIAN J ANAESTH, V46, P261
[9]  
Kobayashi A, 2000, AVIAT SPACE ENVIR MD, V71, P177
[10]   Cerebral Near-Infrared Spectroscopy to Evaluate Anti-G Straining Maneuvers in Centrifuge Training [J].
Kobayashi, Asao ;
Kikukawa, Azusa ;
Kimura, Mikihiko ;
Inui, Takuo ;
Miyamoto, Yoshinori .
AVIATION SPACE AND ENVIRONMENTAL MEDICINE, 2012, 83 (08) :790-794