Amperometric biosensors with a laminated distribution of enzymes in their coating. Steady-state kinetics

被引:2
作者
Sorochinskii, VV [1 ]
Kurganov, BI [1 ]
机构
[1] RUSSIAN ACAD SCI,AN BACH INST BIOCHEM,MOSCOW 117071,RUSSIA
关键词
amperometric biosensors; steady-state kinetics;
D O I
10.1016/0956-5663(96)83712-2
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Calculations are given of the stationary concentration of substrate and products in the coating of amperometric biosensors involving consecutive multi-enzyme transformations of a substrate. The coating of a biosensor consists of an outer and inner membrane making a multi-enzyme sandwich comprising layers of separately immobilized enzymes. The relationships controlling the effect of the concentration of enzymes and their kinetic parameters, and the permeability and thickness of the coating components, as well as the parameters of the indicator reaction of biosensor sensitivity and the value of the concentration range suitable for substrate analysis were evaluated.
引用
收藏
页码:45 / 51
页数:7
相关论文
共 8 条
[1]  
KUBO T, 1983, P INT M CHEM SENSORS, P660
[2]   AMPEROMETRIC ACETYLCHOLINE AND CHOLINE SENSORS WITH IMMOBILIZED ENZYMES [J].
MASCINI, M ;
MOSCONE, D .
ANALYTICA CHIMICA ACTA, 1986, 179 :439-444
[3]   HYBRID UREA SENSOR USING NITRIFYING BACTERIA [J].
OKADA, T ;
KARUBE, I ;
SUZUKI, S .
EUROPEAN JOURNAL OF APPLIED MICROBIOLOGY AND BIOTECHNOLOGY, 1982, 14 (03) :149-154
[4]   PLANT-TISSUE HYDRID ELECTRODE FOR DETERMINATION OF PHOSPHATE AND FLUORIDE [J].
SCHUBERT, F ;
RENNEBERG, R ;
SCHELLER, FW ;
KIRSTEIN, L .
ANALYTICAL CHEMISTRY, 1984, 56 (09) :1677-1682
[5]   MATHEMATICAL-MODELING OF AMPEROMETRIC ENZYME ELECTRODES WITH PERFORATED MEMBRANES [J].
SCHULMEISTER, T ;
PFEIFFER, D .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (02) :75-79
[6]  
SOROCHINSKII VV, 1986, ZH ANAL KHIM, V41, P2055
[7]  
SOROCHINSKII VV, 1988, ITOP NAUKI TEKHNIK B, V13
[8]  
SOROCHINSKII VV, 1991, KINET KATAL, V32, P312