Use of monoatomic and polyatomic projectiles for the characterisation of polylactic acid by static secondary ion mass spectrometry

被引:8
|
作者
Boschmans, B [1 ]
Van Royen, P [1 ]
Van Vaeck, L [1 ]
机构
[1] Univ Antwerp CDE, Dept Chem, MiTAC, B-2610 Antwerp, Belgium
关键词
D O I
10.1002/rcm.2089
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The application of polyatomic primary ions is a strongly developing branch of static secondary ion mass spectrometry (S-SIMS), since these projectiles allow a significant increase in the secondary ion yields to be achieved. However, the different limitations and possibilities of certain polyatomic primary ions for use on specific functional classes of samples are still not completely known. This paper compares the use of monoatomic and polyatomic primary ions in S-SIMS for thin layers of polylactic acid (PLA), obtained by spin-coating solutions on silicon wafers. Bombardment with Ga+, Xe+ and SF5+ primary ions allowed the contribution of the projectile mass and number of atoms in the gain in ion yield and molecular specificity (relative importance of high m/z and low m/z signals) to be assessed. Samples obtained by spin-coating solutions with increasing concentration showed that optimal layer thickness depended on the primary ion used. In comparison with the use of Ga+ projectiles, the yield of structural ions increased by a factor of about 1.5 to 2 and by about 7 to 12 when Xe+ and SF5+ primary ion bombardment were applied, respectively. A detailed fragmentation pattern was elaborated to interpret ion signal intensity changes for different projectiles in terms of energy deposition and collective processes in the subsurface, and the internal energy of radical and even-electron precursor ions. Copyright (C) 2005 John Wiley & Sons, Ltd.
引用
收藏
页码:2517 / 2527
页数:11
相关论文
共 50 条
  • [41] Secondary ion mass spectrometry
    Griffiths, Jennifer
    ANALYTICAL CHEMISTRY, 2008, 80 (19) : 7194 - 7197
  • [42] Suppression and enhancement of secondary ion formation due to the chemical environment in static-secondary ion mass spectrometry
    Jones, Emrys A.
    Lockyer, Nicholas P.
    Kordys, Jeanette
    Vickerman, John C.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (08) : 1559 - 1567
  • [43] Secondary ion mass spectrometry
    不详
    NATURE REVIEWS METHODS PRIMERS, 2024, 4 (01):
  • [44] Secondary ion mass spectrometry
    Handley, Judith
    Analytical Chemistry, 2002, 74 (11)
  • [45] Secondary ion mass spectrometry
    1600, AOAC International (87):
  • [46] Secondary ion mass spectrometry
    Sherma, J
    JOURNAL OF AOAC INTERNATIONAL, 2004, 87 (05) : 142A - 148A
  • [47] Secondary ion mass spectrometry
    Sangely, Laure
    Boyer, Bernard
    De Chambost, Emmanuel
    Valle, Nathalie
    Audinot, Jean-Nicolas
    Ireland, Trevor
    Wiedenbeck, Michael
    Aléon, JéRôme
    Jungnickel, Harald
    Barnes, Jean-Paul
    Bienvenu, Philippe
    Breuer, Uwe
    New Developments in Mass Spectrometry, 2015, 2015-January (03): : 439 - 499
  • [48] Suppression and Enhancement of Secondary Ion Formation Due to the Chemical Environment in Static-Secondary Ion Mass Spectrometry
    Jones, Emrys A.
    Lockyer, Nicholas P.
    Kordys, Jeanette
    Vickerman, John C.
    Journal of the American Society for Mass Spectrometry, 2007, 18 (08): : 1559 - 1567
  • [49] USE OF SECONDARY ION MASS-SPECTROMETRY IN SURFACE ANALYSIS
    WERNER, HW
    SURFACE SCIENCE, 1975, 47 (01) : 301 - 323
  • [50] THE ANALYSIS OF VARIOUS POLYCARBONATES BY STATIC SECONDARY-ION MASS-SPECTROMETRY
    LUB, J
    BUNING, GHW
    POLYMER, 1990, 31 (06) : 1009 - 1017