Performance of the XMM EPIC MOS CCD detectors

被引:39
作者
Short, AD [1 ]
Keay, A [1 ]
Turner, MJL [1 ]
机构
[1] Univ Leicester, Dept Phys & Astron, Xray Astron Grp, Leicester LE1 7RH, Leics, England
来源
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY IX | 1998年 / 3445卷
关键词
XMM; EPIC; CCD; open electrode; resolution; quantum efficiency; simulation;
D O I
10.1117/12.330276
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
Measurements made during the selection and evaluation of flight CCD detectors for the XMM EPIC MOS cameras have demonstrated near Fano limited resolution at X-ray energies above approximately 3keV. At lower energies some devices exhibit a fractional charge loss which is believed to be due to recombination at the epitaxy/oxide interface. This has been modelled through a Monte-Carlo simulation by assuming that the pinning implant in the etched electrode structure can cause electrons to flow to the front surface, rather than to the buried channel. In spite of this charge loss, spectral response may be characterised using a double Gaussian with residuals of <5%. Quantum efficiency has been measured using a lithium drifted silicon (Si(Li)) reference detector and these measurements combined with analytical and Monte Carlo simulation, event size ratios and cosmic ray detection, all give a value for the effective depletion depth of 30 to 35 mu m.
引用
收藏
页码:13 / 27
页数:15