Evaluation of the Airtraq® and Macintosh laryngoscopes in patients at increased risk for difficult tracheal intubation

被引:129
作者
Maharaj, C. H.
Costello, J. F.
Harte, B. H. [1 ]
Laffey, J. G. [1 ]
机构
[1] Univ Coll Hosp Galway, Dept Anaesthesia & Intens Care Med, Galway, Ireland
关键词
D O I
10.1111/j.1365-2044.2007.05316.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Airtraq (R), a novel single use indirect laryngoscope, has demonstrated promise in the normal and simulated difficult airway. We compared the ease of intubation using the Airtraq with the Macintosh laryngoscope, in patients at increased risk for difficult tracheal intubation, in a randomised, controlled clinical trial. Forty consenting patients presenting for surgery requiring tracheal intubation, who were deemed to possess at least three characteristics indicating an increased risk for difficulty in tracheal intubation, were randomly assigned to undergo tracheal intubation using a Macintosh (n = 20) or Airtraq (n = 20) laryngoscope. All patients were intubated by one of three anaesthetists experienced in the use of both laryngoscopes. Four patients were not successfully intubated with the Macintosh laryngoscope, but were intubated successfully with the Airtraq. The Airtraq reduced the duration of intubation attempts (mean (SD); 13.4 (6.3) vs 47.7 (8.5) s), the need for additional manoeuvres, and the intubation difficulty score (0.4 (0.8) vs 7.7 (3.0)). Tracheal intubation with the Airtraq also reduced the degree of haemodynamic stimulation and minor trauma compared to the Macintosh laryngoscope.
引用
收藏
页码:182 / 188
页数:7
相关论文
共 22 条
[11]   A comparison of tracheal intubation using the Airtraq® or the Macintosh laryngoscope in routine airway management:: a randomised, controlled clinical trial [J].
Maharaj, C. H. ;
O'Croinin, D. ;
Curley, G. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2006, 61 (11) :1093-1099
[12]   Learning and performance of tracheal intubation by novice personnel:: a comparison of the Airtraq® and Macintosh laryngoscope [J].
Maharaj, C. H. ;
Costello, J. F. ;
Higgins, B. D. ;
Harte, B. H. ;
Laffey, J. G. .
ANAESTHESIA, 2006, 61 (07) :671-677
[13]   Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy -: a manikin study [J].
Maharaj, CH ;
Higgins, BD ;
Harte, BH ;
Laffey, JG .
ANAESTHESIA, 2006, 61 (05) :469-477
[14]   Endotracheal intubation in patients with cervical spine immobilization - A comparison of Macintosh and Airtraq laryngoscopes [J].
Maharaj, Chrisen H. ;
Buckley, Elma ;
Harte, Brian H. ;
Laffey, John G. .
ANESTHESIOLOGY, 2007, 107 (01) :53-59
[15]   Tracheal intubation by inexperienced medical residents using the Airtraq and Macintosh laryngoscopes - a manikin study [J].
Maharaj, Chrisen H. ;
Chonghaile, Martina Ni ;
Higgins, Brendan D. ;
Harte, Brian H. ;
Laffey, John G. .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2006, 24 (07) :769-774
[16]   A CLINICAL SIGN TO PREDICT DIFFICULT TRACHEAL INTUBATION - A PROSPECTIVE-STUDY [J].
MALLAMPATI, SR ;
GATT, SP ;
GUGINO, LD ;
DESAI, SP ;
WARAKSA, B ;
FREIBERGER, D ;
LIU, PL .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1985, 32 (04) :429-434
[17]  
Montes FR, 2003, CAN J ANAESTH, V50, P824, DOI 10.1007/BF03019380
[18]  
Nathan N, 2004, CAN J ANAESTH, V51, P382, DOI 10.1007/BF03018244
[19]   CARDIOVASCULAR AND CATECHOLAMINE RESPONSES TO LARYNGOSCOPY WITH AND WITHOUT TRACHEAL INTUBATION [J].
SHRIBMAN, AJ ;
SMITH, G ;
ACHOLA, KJ .
BRITISH JOURNAL OF ANAESTHESIA, 1987, 59 (03) :295-299
[20]   Combined use of esmolol and nicardipine to blunt the haemodynamic changes following laryngoscopy and tracheal intubation [J].
Tan, PH ;
Yang, LC ;
Shih, HC ;
Lin, CR ;
Lan, KC ;
Chen, CS .
ANAESTHESIA, 2002, 57 (12) :1207-1212