Studies on x-ray and UV emissions in electron cyclotron resonance x-ray source

被引:1
作者
Baskaran, R. [1 ]
Selvakumaran, T. S. [1 ]
机构
[1] Indira Gandhi Ctr Atom Res, Radiol Safety Div, Kalpakkam 603102, Tamil Nadu, India
关键词
D O I
10.1063/1.2801651
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel electron cyclotron resonance x-ray source is constructed based on the ECR technique. In this paper, the possibility of using the ECR x-ray source for producing UV rays by optimizing the plasma parameters is explored. X-ray and UV emissions from the ECR x-ray source are carried out for argon, nitrogen, and CO2 plasma. The x-ray spectral and dose measurements are carried with Nal(Tl) based spectrometer and dosimeter, respectively. For UV measurement, a quartz window arrangement is made at the exit port and the UV intensity is measured at 5 cm from the quartz plate using UV meter. The x-ray and UV emissions are carried out for different microwave power levels and gas pressures. The x-ray emission is observed in the pressure range <= 10(-5) Torr, whereas the UV emission is found to be negligible for the gas pressures < 10(-5) Torr and it starts increasing in the pressure range between 10(-5) and 10(-3) Torr. At high-pressure range, collision frequency of electron-atom is large which leads to the higher UV flux. At low pressure, the electron-atom collision frequency is low and hence the electrons reach high energy and by hitting the cavity wall produces higher x-ray flux. By choosing proper experimental conditions and plasma gas species, the same source can be used as either an x-ray source or an UV source. 0 2008 American Institute Of Physics.
引用
收藏
页数:3
相关论文
共 13 条
  • [1] A compact radiological X-ray source
    Bacal, M
    Gaudin, C
    Bourdier, A
    Bruneteau, J
    Buzzi, JM
    Golovanivsky, KS
    Hay, L
    Rouille, C
    Schwartz, L
    [J]. NATURE, 1996, 384 (6608) : 421 - 421
  • [2] Experimental studies on a compact electron cyclotron resonance plasma x-ray source
    Baskaran, R
    Selvakumaran, TS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (06) : 2637 - 2645
  • [3] Theoretical analysis of electron cyclotron resonance x-ray sources
    Baskaran, R
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (10) : 3510 - 3514
  • [4] Studies on enhancement of x-ray flux in the compact electron cyclotron resonance plasma x-ray source
    Baskaran, R
    Selvakumaran, TS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2000, 71 (02) : 1203 - 1205
  • [5] Studies on the effective energy of x-rays generated by ECR and their use for the calibration of thermoluminescent dosimeter badges in low energy region
    Baskaran, R
    Selvakumaran, TS
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (04)
  • [6] BASKARAN R, 2006, REV SCI INSTRUM, V77, P1
  • [7] BASKARAN R, IN PRESS INDIAN J ND
  • [8] BASKARAN R, 2006, REV SCI INSTRUM, V3, P1
  • [9] Electron cyclotron resonance light source from TE011 mode microwave plasma
    Chauhan, SS
    Rajyaguru, CC
    Ito, H
    Yugami, N
    Nishida, Y
    Yoshida, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (12) : 4344 - 4347
  • [10] FEED OPTIMIZATION FOR THE SLOTTED LINE ANTENNA FOR HIGH-DENSITY PLASMA PRODUCTION