A Monte Carlo simulation of an encapsulated Ir-192 source for high dose rate (HDR) prostate brachytherapy is presented. Basic dosimetric quantities related to this source are derived from the simulation data following the AAPM Task Group 43 formalism, including dose rate constant, source kerma strength, anisotropy function and radial dose function. Dose rate distribution data in a water phantom were obtained using cylindrical geometry covering +/-15 cm in the longitudinal and transverse directions from the center of the source.
引用
收藏
页码:128 / 133
页数:6
相关论文
共 8 条
[1]   Spectra and air-kerma strength for encapsulated 192Ir sources [J].
Borg, J ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (11) :2441-2444
[2]   Determination of the tissue attenuation factor along two major axes of a high dose rate (HDR) 192Ir source [J].
Cho, SH ;
Muller-Runkel, R ;
Hanson, WF .
MEDICAL PHYSICS, 1999, 26 (08) :1492-1497
[3]   A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source [J].
Karaiskos, P ;
Angelopoulos, A ;
Baras, P ;
Sakelliou, L ;
Sandilos, P ;
Dardoufas, K ;
Vlachos, L .
MEDICAL PHYSICS, 1999, 26 (08) :1498-1502
[4]   DOSIMETRY OF INTERSTITIAL BRACHYTHERAPY SOURCES - RECOMMENDATIONS OF THE AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO 43 [J].
NATH, R ;
ANDERSON, LL ;
LUXTON, G ;
WEAVER, KA ;
WILLIAMSON, JF ;
MEIGOONI, AS .
MEDICAL PHYSICS, 1995, 22 (02) :209-234
[5]  
*SECR SAL, 1997, COMP REG HIST NEOPL
[6]   An algorithm for Monte Carlo simulation of coupled electron-photon transport [J].
Sempau, J ;
Acosta, E ;
Baro, J ;
FernandezVarea, JM ;
Salvat, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1997, 132 (03) :377-390
[7]   Monte Carlo dosimetry of the VariSource high dose rate 192Ir source [J].
Wang, RQ ;
Sloboda, RS .
MEDICAL PHYSICS, 1998, 25 (04) :415-423
[8]   MONTE-CARLO AIDED DOSIMETRY OF THE MICROSELECTRON PULSED AND HIGH-DOSE-RATE IR-192 SOURCES [J].
WILLIAMSON, JF ;
LI, ZF .
MEDICAL PHYSICS, 1995, 22 (06) :809-819