Evaluation of the YSI 2300 glucose analyzer: Algorithm-corrected results are accurate and specific

被引:26
作者
Astles, JR [1 ]
Sedor, FA [1 ]
Toffaletti, JG [1 ]
机构
[1] DUKE UNIV,MED CTR,DEPT PATHOL,DIV CLIN LABS,DURHAM,NC 27710
关键词
whole blood; hematocrit; glucose electrode; acetaminophen;
D O I
10.1016/0009-9120(95)02010-1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To insure commutability (equivalence) between whole blood and plasma glucose values and to assess potential interferences in glucose oxidase methods. Design and Methods: We compared plasma glucose results by the Ektachem 700 analyzer to glucose results on whole blood by the Nova Stat Profile 6, and by the YSI 2300, both without hematocrit correction (YSIunc) and with two different hematocrit (Hct) corrections. The two correction methods were: (a) whole blood results multiplied by 1.11 (YSI1.11), and (b) whole blood results corrected for Hct with the programmed YSI algorithm (YSIcor). Several compounds were tested for interference. Results: 88% of YSIcor results agreed with plasma results within +/-5%, and 78% of Nova and 74% of YSI1.11 results were within this limit. YSIcor results were unaffected by Hct and were valid even when erythrocyte size was abnormal. At low Hct, Nova results were falsely decreased, while YSI1.11 results were falsely elevated. The Ektachem was most affected by ascorbic acid, and the Nova was especially affected by acetaminophen. Conclusions: The YSI algorithm-corrected whole blood glucose results were commutable with plasma results and the YSI was unaffected by the compounds tested, including acetaminophen and acetylsalicylic acid.
引用
收藏
页码:27 / 31
页数:5
相关论文
共 16 条
[1]  
COPLAND AM, 1992, BRIT J GEN PRACT, V42, P259
[2]   PARACETAMOL INTERFERENCE WITH BLOOD-GLUCOSE ANALYSIS - A POTENTIALLY FATAL PHENOMENON [J].
FARAH, DA ;
BOAG, D ;
MORAN, F ;
MCINTOSH, S .
BRITISH MEDICAL JOURNAL, 1982, 285 (6336) :172-172
[3]  
FLEETWOOD JA, 1981, CLIN CHEM, V27, P1945
[4]   DIRECT READING GLUCOSE ELECTRODES DETECT THE MOLALITY OF GLUCOSE IN PLASMA AND WHOLE-BLOOD [J].
FOGHANDERSEN, N ;
WIMBERLEY, PD ;
THODE, J ;
SIGGAARDANDERSEN, O .
CLINICA CHIMICA ACTA, 1990, 189 (01) :33-38
[5]   THE NECESSITY OF ACHIEVING GOOD LABORATORY PERFORMANCE [J].
FRASER, CG .
DIABETIC MEDICINE, 1990, 7 (06) :490-493
[6]  
HENDERSON AR, 1994, TIETZ TXB CLIN CHEM, P1576
[7]   PLASMA OR WHOLE-BLOOD GLUCOSE [J].
MORRISON, B ;
FLECK, A .
CLINICA CHIMICA ACTA, 1973, 45 (03) :293-297
[8]  
PAINTER PC, 1994, TIETZ TXB CLIN CHEM, P2161
[9]  
REX CS, 1981, DIABETES CARE, V4, P392
[10]   INVITRO DRUG INTERFERENCE WITH HOME BLOOD-GLUCOSE-MEASUREMENT SYSTEMS [J].
RICE, GK ;
GALT, KA .
AMERICAN JOURNAL OF HOSPITAL PHARMACY, 1985, 42 (10) :2202-2207