Design and test of a 64-channel charge measurement ASIC developed in CMOS 0.35 μm technology

被引:21
作者
La Rosa, A. [1 ,2 ]
Mazza, G. [2 ]
Donetti, M. [2 ,3 ]
Marchetto, F. [2 ]
Luetto, L. [1 ]
Attili, A. [2 ]
Bourhaleb, F. [1 ]
Cirio, R. [2 ]
Garella, M. A. [2 ]
Giordariengo, S. [2 ]
Givehchi, N. [1 ,2 ]
Iliescu, S. [2 ]
Pardo, J. [2 ]
Pecka, A. [1 ,2 ]
Peroni, C. [1 ,2 ]
Pitta, G. [4 ]
机构
[1] Univ Turin, Dipartimento Fis Sperimentale, I-10125 Turin, Italy
[2] Ist Nazl Fis Nucl, I-10125 Turin, Italy
[3] Fdn CNAO, I-20123 Milan, Italy
[4] Fdn TERA, I-28100 Novara, Italy
关键词
VLSI electronics; CMOS technology; current-to-frequency converter; Dosimetry; beam position monitor; radiotherapy; hadrontherapy;
D O I
10.1016/j.nima.2007.09.024
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 64-channel charge measurement (Application-Specific Integrated Circuit) ASIC has been designed and tested: it is intended to serve as a front-end electronic read-out for detectors to monitor and measure radiotherapeutical beams. The ASIC has been designed in a CMOS 0.35 mu m technology with particular attention to the linearity over a wide input range and can accept currents of both polarities. The linearity is better than 1.5% for a dynamic range of the input current between 500 pA and 3 mu A. For a charge resolution of 350 fC, the spread (r.m.s.) of the gain is less than 1%. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:461 / 468
页数:8
相关论文
共 6 条
[1]   Performances of a VLSI wide dynamic range current-to-frequency converter for strip ionization chambers [J].
Bonazzola, GC ;
Cirio, R ;
Donetti, M ;
Marchetto, F ;
Mazza, G ;
Peroni, C ;
Zampieri, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1998, 405 (01) :111-120
[2]   A pixel chamber to monitor the beam performances in hadron therapy [J].
Bonin, R ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Garelli, E ;
Giordanengo, S ;
Marchetto, F ;
Peroni, C ;
Freire, CJS ;
Simonetti, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 519 (03) :674-686
[3]  
GHIVECHI N, 2007, NUCL INSTRUM METH A, V573, P1094
[4]   CHARGE-BALANCING CURRENT INTEGRATOR WITH LARGE DYNAMIC-RANGE [J].
GOTTSCHALK, B .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 207 (03) :417-421
[5]   A pixel ionization chamber used as beam monitor at the Institut Curie-Centre de Protontherapie de Orsay (CPO) [J].
La Rosa, A. ;
Garella, M. A. ;
Bourhaleb, F. ;
Cirio, R. ;
Donetti, M. ;
Giordanengo, S. ;
Givehchi, N. ;
Marchetto, F. ;
Martin, F. ;
Meyroneinc, S. ;
Peroni, C. ;
Pitta, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 565 (02) :833-840
[6]   A 64-channel wide dynamic range charge measurement ASIC for strip and pixel ionization detectors [J].
Mazza, G ;
Cirio, R ;
Donetti, M ;
La Rosa, A ;
Luparia, A ;
Marchetto, F ;
Peroni, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (04) :847-853