Sodium inversion recovery MRI of the knee joint in vivo at 7T

被引:56
作者
Madelin, Guillaume [1 ,2 ]
Lee, Jae-Seung [1 ,2 ]
Inati, Souheil [3 ]
Jerschow, Alexej [1 ]
Regatte, Ravinder R. [2 ]
机构
[1] NYU, Dept Chem, New York, NY 10012 USA
[2] NYU, Med Ctr, Ctr Biomed Imaging, New York, NY 10012 USA
[3] NIH, Bethesda, MD 20892 USA
关键词
Osteoarthritis; Cartilage; Sodium; Magnetic resonance imaging; Inversion recovery; Adiabatic inversion; BOVINE ARTICULAR-CARTILAGE; FIXED CHARGE-DENSITY; MAGNETIC-RESONANCE; ADIABATIC PULSES; HUMAN BRAIN; NA-23; MRI; NMR;
D O I
10.1016/j.jmr.2010.08.003
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The loss of proteoglycans (PG) in the articular cartilage is an early signature of osteoarthritis (OA). The ensuing changes in the fixed charge density in the cartilage can be directly linked to sodium concentration via charge balance. Sodium ions in the knee joint appear in two pools: in the synovial fluids or joint effusion where the ions are in free motion and bound within the cartilage tissue where the Na+ ions have a restricted motion. The ions in these two compartments have therefore different T-1 and T-2 relaxation times. The purpose of this study is to demonstrate the feasibility of a fluid-suppressed 3D ultrashort TE radial sodium sequence by implementing an inversion recovery (IR) preparation of the magnetization at 7T. This method could allow a more accurate and more sensitive quantification of loss of PG in patients with OA. It is shown that adiabatic pulses offer significantly improved performance in terms of robustness to B-1 and B-0 inhomogeneities when compared to the hard pulse sequence. Power deposition considerations further pose a limit to the RF inversion power, and we demonstrate in simulations and experiments how a practical compromise can be struck between clean suppression of fluid signals and power deposition levels. Two IR sequences with different types of inversion pulses (a rectangular pulse and an adiabatic pulse) were tested on a liquid phantom, ex vivo on a human knee cadaver and then in vivo on five healthy volunteers, with a (Nyquist) resolution of similar to 3.6 mm and a signal-to-noise ratio of similar to 30 in cartilage without IR and similar to 20 with IR. Due to specific absorption rate limitations, the total acquisition time was similar to 17 min for the 3D radial sequence without inversion or with the rectangular IR, and 24:30 min for the adiabatic IR sequence. It is shown that the adiabatic IR sequence generates a more uniform fluid suppression over the whole sample than the rectangular IR sequence. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:42 / 52
页数:11
相关论文
共 30 条
[1]  
[Anonymous], ADV MRI KNEE OSTEOAR
[2]   Gd-DTPA(2-) as a measure of cartilage degradation [J].
Bashir, A ;
Gray, ML ;
Burstein, D .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :665-673
[3]   Quantifying sodium in the human wrist in vivo by using MR imaging [J].
Borthakur, A ;
Shapiro, EM ;
Akella, SVS ;
Gougoutas, A ;
Kneeland, JB ;
Reddy, R .
RADIOLOGY, 2002, 224 (02) :598-602
[4]   Sodium and T1ρ MRI for molecular and diagnostic imaging of articular cartilage [J].
Borthakur, Arijitt ;
Mellon, Eric ;
Niyogi, Sampreet ;
Witschey, Walter ;
Kneeland, J. Bruce ;
Reddy, Ravinder .
NMR IN BIOMEDICINE, 2006, 19 (07) :781-821
[5]   The return of the frequency sweep: Designing adiabatic pulses for contemporary NMR [J].
Garwood, M ;
DelaBarre, L .
JOURNAL OF MAGNETIC RESONANCE, 2001, 153 (02) :155-177
[6]   Accelerating the nonuniform fast Fourier transform [J].
Greengard, L ;
Lee, JY .
SIAM REVIEW, 2004, 46 (03) :443-454
[7]  
GREENGARD L, 2006, APPL MATH COMPUT SCI, V1, P121
[8]  
Hargreaves B., BLOCH EQUATION SIMUL
[9]  
INATI S, 2006, P 14 ANN M ISMRM SEA
[10]  
INATI S, 2005, P 13 ANN M ISMRM MIA