Comparison of pleural fluid pH values obtained using blood gas machine, pH meter, and pH indicator strip

被引:52
作者
Cheng, DS
Rodriguez, RM
Rogers, J
Wagster, M
Starnes, DL
Light, RW
机构
[1] St Thomas Hosp, Pulm Dis Program, Dept Med, Nashville, TN 37202 USA
[2] St Thomas Hosp, Dept Radiol, Nashville, TN 37202 USA
[3] St Thomas Hosp, Dept Pathol, Nashville, TN 37202 USA
[4] Vanderbilt Univ, Nashville, TN USA
关键词
blood gas machine; empyema; pH; pH meter; pleural fluid;
D O I
10.1378/chest.114.5.1368
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study purpose: The purpose of this study was to compare the pleural fluid pH values obtained with a blood gas machine (pHbg), with a pH meter (pHmet), and with a pH indicator strip (pHstrip), to determine if the pleural fluid pH measured by a pH meter or a pH indicator strip was sufficiently accurate for clinical decisions, Methods: The pleural fluid pH was determined, within 20 min after being collected anaerobically, by a blood gas machine (CIBA-Corning model 288), pH meter (Corning pH meter 610A), and pH indicator strip (Baxter Diagnostic) following routine laboratory procedures in 50 pleural fluids. Pleural fluid pH was determined in seven additional samples with the blood gas machine and a pH meter at 25 and 37 degrees C respectively, initially, and after 30 min, Results: The mean pHbg (7.42 +/- 0.01) was significantly less than the mean pHmet (7.58 +/- 0.02) or the mean pHstrip (8.23 +/- 0.06), There were significant differences between the pHbg and the pHmet (p < 0.001), and between the pHbg and the pHstrip (p < 0.001), Analysis of the additional seven samples demonstrated that when the blood gas machine was set at 25 degrees C, the pHbg (pHbg = 7.54 +/- 0.02) and the pHmet (7.53 +/- 0.01) were almost identical. Conclusion: When the pleural fluid pH is going to be used for decision making, only the pH values provided by the blood gas machine are sufficiently accurate.
引用
收藏
页码:1368 / 1372
页数:5
相关论文
共 10 条
  • [1] PLEURAL FLUID CHEMICAL-ANALYSIS IN PARAPNEUMONIC EFFUSIONS - A METAANALYSIS
    HEFFNER, JE
    BROWN, LK
    BARBIERI, C
    DELEO, JM
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1995, 151 (06) : 1700 - 1708
  • [2] Is pH paper an acceptable, low-cost alternative to the blood gas analyzer for determining pleural fluid pH?
    Lesho, EP
    Roth, BJ
    [J]. CHEST, 1997, 112 (05) : 1291 - 1292
  • [3] DIAGNOSTIC SIGNIFICANCE OF PLEURAL FLUID PH AND PCO2
    LIGHT, RW
    MACGREGOR, MI
    BALL, WC
    LUCHSINGER, PC
    [J]. CHEST, 1973, 64 (05) : 591 - 596
  • [4] PLEURAL EFFUSIONS - DIAGNOSTIC SEPARATION OF TRANSUDATES AND EXUDATES
    LIGHT, RW
    LUCHSINGER, PC
    MACGREGOR, MI
    BALL, WC
    [J]. ANNALS OF INTERNAL MEDICINE, 1972, 77 (04) : 507 - +
  • [5] PARAPNEUMONIC EFFUSIONS
    LIGHT, RW
    GIRARD, WM
    JENKINSON, SG
    GEORGE, RB
    [J]. AMERICAN JOURNAL OF MEDICINE, 1980, 69 (04) : 507 - 512
  • [6] Light RW, 1995, Pleural Diseases
  • [7] PLEURAL FLUID PH IN PARAPNEUMONIC EFFUSIONS
    POTTS, DE
    LEVIN, DC
    SAHN, SA
    [J]. CHEST, 1976, 70 (03) : 328 - 331
  • [8] SURVIVAL-TIME OF PATIENTS WITH PLEURAL METASTATIC CARCINOMA PREDICTED BY GLUCOSE AND PH STUDIES
    RODRIGUEZPANADERO, F
    LOPEZMEJIAS, J
    [J]. CHEST, 1989, 95 (02) : 320 - 324
  • [9] LOW GLUCOSE AND PH LEVELS IN MALIGNANT PLEURAL EFFUSIONS - DIAGNOSTIC-SIGNIFICANCE AND PROGNOSTIC VALUE IN RESPECT TO PLEURODESIS
    RODRIGUEZPANADERO, F
    MEJIAS, JL
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (03): : 663 - 667
  • [10] THE PLEURA
    SAHN, SA
    [J]. AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (01): : 184 - 234