Reduced Image Noise at Low-Dose Multidetector CT of the Abdomen with Prior Image Constrained Compressed Sensing Algorithm

被引:68
作者
Lubner, Meghan G. [1 ]
Pickhardt, Perry J. [1 ]
Tang, Jie [2 ]
Chen, Guang-Hong [1 ,2 ]
机构
[1] Univ Wisconsin, Sch Med & Publ Hlth, Dept Radiol, Madison, WI 53792 USA
[2] Univ Wisconsin, Sch Med & Publ Hlth, Dept Med Phys, Madison, WI 53792 USA
关键词
STATISTICAL ITERATIVE RECONSTRUCTION; TUBE CURRENT MODULATION; AUTOMATIC EXPOSURE CONTROL; COMPUTED-TOMOGRAPHY; COLORECTAL-CANCER; ROW CT; REDUCTION; COLONOGRAPHY; QUALITY; OPTIMIZATION;
D O I
10.1148/radiol.11101380
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the effect of prior image constrained compressed sensing (PICCS) on noise reduction and image quality at low-dose computed tomography (CT). Materials and Methods: This HIPAA-compliant institutional review board-approved retrospective study was performed by using DICOM CT colonography data sets obtained in 20 adult patients. Informed consent was waived. Low-dose CT colonography was performed with 64-detector CT by using the standard protocol with mean effective dose per series of 3.06 mSv (range, 1.4-7.7 mSv). PICCS was applied to standard filtered back-projection (FBP) series. For FBP and PICCS series, mean and standard deviation (SD) of attenuation were obtained with 100-mm(2) circular region of interest (ROI) at six sites (240 soft-tissue, colonic gas, and subcutaneous fat measurements). Two abdominal radiologists reviewed two-and three-dimensional CT colonography displays and graded image quality with a five-point scale. Phantom studies were performed to compare spatial resolution and image quality between FBP and PICCS. Mean image noise and image quality scores were calculated and compared for clinical and phantom data sets. Bland-Altman, generalized estimating equation regression model, and Student t tests were used to obtain limits of agreement and to compare noise ratios and subjective image quality. Results: Mean SD of attenuation (image noise) for ROIs was 38.0 for FBP and 12.2 for PICCS, corresponding to a noise-reduction factor of 3.1 (P<.001). Average noise reduction was 3.3 for soft tissue, 2.8 for air, and 3.0 for fat attenuation. Attenuation did not substantially change between FBP and PICCS images. Average two-dimensional image quality was 2.45 for FBP and 3.4 for PICCS (P<.001). Average three-dimensional image quality at three sites in the colon was 3.5 for FBP and 3.7 for PICCS (P = .34). Phantom data sets revealed no loss of spatial resolution in a line phantom and reduced noise in a liver tumor phantom when PICCS was compared with FBP. Conclusion: Application of PICCS to standard FBP low-dose multidetector CT abdominal images results in substantial noise reduction and improved image quality. (C) RSNA, 2011
引用
收藏
页码:248 / 256
页数:9
相关论文
共 30 条
  • [1] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [2] Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets
    Chen, Guang-Hong
    Tang, Jie
    Leng, Shuai
    [J]. MEDICAL PHYSICS, 2008, 35 (02) : 660 - 663
  • [3] Prior Image Constrained Compressed Sensing (PICCS) and Applications in X-ray Computed Tomography
    Chen, Guang-Hong
    Tang, Jie
    Nett, Brian
    Qi, Zhihua
    Leng, Shuai
    Szczykutowicz, Timothy
    [J]. CURRENT MEDICAL IMAGING, 2010, 6 (02) : 119 - 134
  • [4] Evaluation of image quality and lesion perception by human readers on 3D CT colonography: comparison of standard and low radiation dose
    Fisichella, Valeria A.
    Bath, Magnus
    Johnsson, Ase Allansdotter
    Jaderling, Fredrik
    Bergsten, Tommy
    Persson, Ulf
    Mellingen, Kristin
    Hellstrom, Mikael
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (03) : 630 - 639
  • [5] Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study
    Flicek, Kristina T.
    Hara, Amy K.
    Silva, Alvin C.
    Wu, Qing
    Peter, Mary B.
    Johnson, C. Daniel
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) : 126 - 131
  • [6] Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies
    Gies, M
    Kalender, WA
    Wolf, H
    Suess, C
    Madsen, MT
    [J]. MEDICAL PHYSICS, 1999, 26 (11) : 2235 - 2247
  • [7] Dose reduction and image quality in MDCT colonography using tube current modulation
    Graser, A.
    Wintersperger, B. J.
    Suess, C.
    Reiser, M. F.
    Becker, C. R.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (03) : 695 - 701
  • [8] Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions
    Greess, H
    Wolf, H
    Baum, U
    Lell, M
    Pirkl, M
    Kalender, W
    Bautz, WA
    [J]. EUROPEAN RADIOLOGY, 2000, 10 (02) : 391 - 394
  • [9] THE EFFECT OF MAS VARIATION UPON COMPUTED-TOMOGRAPHY IMAGE QUALITY AS EVALUATED BY INVIVO AND INVITRO STUDIES
    HAAGA, JR
    MIRALDI, F
    MACINTYRE, W
    LIPUMA, JP
    BRYAN, PJ
    WIESEN, E
    [J]. RADIOLOGY, 1981, 138 (02) : 449 - 454
  • [10] Effective dose determination using an anthropomorphic phantom and metal oxide semiconductor field effect transistor technology for clinical adult body multidetector array computed tomography protocols
    Hurwitz, Lynne M.
    Yoshizumi, Terry T.
    Goodman, Philip C.
    Frush, Donald P.
    Nguyen, Giao
    Toncheva, Greta
    Lowry, Carolyn
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) : 544 - 549