Accurate measurement of extended half-life and unmodified factor VIII low levels with one-stage FVIII assays is dependent on the matrix of calibration curves

被引:5
作者
Van den Bossche, Dorien [1 ]
Toelen, Jelle [1 ]
Schoeters, Joke [1 ]
Van Horenbeeck, Isa [1 ]
Vanlinthout, Ingrid [1 ]
Debasse, Mirjam [1 ]
Peerlinck, Kathelijne [2 ,3 ]
Jacquemin, Marc [1 ,2 ]
机构
[1] Univ Hosp Leuven, Lab Med, Clin Dept, Leuven, Belgium
[2] Univ Leuven, Dept Cardiovasc Sci, Ctr Mol & Vasc Biol, Leuven, Belgium
[3] Univ Hosp Leuven, Vasc Med & Hemostasis, Leuven, Belgium
关键词
extended half-life FVIII; factor VIII; FVIII one-stage FVIII assay; haemophilia A; parallelism; PARALLELISM; PLASMA;
D O I
10.1111/hae.13656
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction The monitoring of factor VIII (FVIII) replacement therapy relies on the accurate measurement of FVIII activity over a large concentration range. However, unexplained overestimation of low FVIII levels has recently been reported with extended half-life recombinant FVIIIs. Aim The objective of this study was to confirm previous publications indicating that the reagents used to generate the calibration curves determine the accuracy of the measurement of low FVIII levels. Methods We generated FVIII calibration curves with FVIII-deficient plasmas or a commercial diluent buffer. We then measured FVIII levels in FVIII-deficient plasma spiked with plasma FVIII, a full-length recombinant FVIII (Advate((R)), Shire, Brussels, Belgium) and two extended half-life recombinant FVIIIs (Elocta((R)) , Swedish Orphan Biovitrum, Woluwe Saint-Lambert, Belgium and Afstyla((R)), CSL Behring, Mechelen, Belgium). FVIII levels were also analyzed in spiked samples prediluted two- and fourfold, either in diluent buffer or in FVIII-deficient plasma to evaluate parallelism. Results Coagulation times of calibration curves generated with diluent buffer were longer than those with FVIII-deficient plasmas. This resulted in an overestimation of FVIII levels lower than 25 IU/dL in spiked samples and in detection of FVIII activity (>= 1 IU/dL) in FVIII-deficient plasma. Predilution of samples with diluent buffer rather than with FVIII-deficient plasma also led to discordant results. Conclusion Our data confirm that the generation of calibration curves by dilution in FVIII-deficient plasma is crucial for the accurate measurement of low FVIII levels. When serial dilutions of samples are analyzed, predilution in FVIII-deficient plasma is required to respect the parallelism criteria. These methods should be more generally implemented for coagulation instruments.
引用
收藏
页码:E19 / E26
页数:8
相关论文
共 10 条
  • [1] The coagulation laboratory monitoring of Afstyla single-chain FVIII concentrate
    Bowyer, A.
    Key, N.
    Dalton, D.
    Kitchen, S.
    Makris, M.
    [J]. HAEMOPHILIA, 2017, 23 (05) : e469 - e470
  • [2] Impact of a product-specific reference standard for the measurement of a PEGylated rFVIII activity: the Swiss Multicentre Field Study
    Bulla, O.
    Poncet, A.
    Alberio, L.
    Asmis, L. M.
    Gahler, A.
    Graf, L.
    Nagler, M.
    Studt, J. -D.
    Tsakiris, D. A.
    Fontana, P.
    [J]. HAEMOPHILIA, 2017, 23 (04) : E335 - E339
  • [3] Accuracy of FVIII:C assay by one-stage method can be improved using hemophilic plasma as diluent
    Cinotti, S
    Paladino, E
    Morfini, M
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2006, 4 (04) : 828 - 833
  • [4] AN EFFECT OF PREDILUTION ON VIII-C DETERMINATION BY THE ONE STAGE ASSAY
    FRANDSEN, H
    STENDER, S
    SCHEIBEL, E
    [J]. THROMBOSIS RESEARCH, 1987, 48 (02) : 195 - 203
  • [5] Evaluation of the activated partial thromboplastin time assay for clinical monitoring of PEGylated recombinant factor VIII (BAY 94-9027) for haemophilia A
    Gu, J-M.
    Ramsey, P.
    Evans, V.
    Tang, L.
    Apeler, H.
    Leong, L.
    Murphy, J. E.
    Laux, V.
    Myles, T.
    [J]. HAEMOPHILIA, 2014, 20 (04) : 593 - 600
  • [6] Koerts JR, 2010, THROMB HAEMOSTASIS, V104, P1080, DOI [10.1160/TH09.12.0812, 10.1160/TH09-12-0812]
  • [7] Measurement of factor VIII activity using one-stage clotting assay: a calibration curve has not to be systematically included in each run
    Lattes, S.
    Appert-Flory, A.
    Fischer, F.
    Jambou, D.
    Toulon, P.
    [J]. HAEMOPHILIA, 2011, 17 (01) : 139 - 142
  • [8] DETERMINATION OF PARALLELISM AND NONPARALLELISM IN BIOASSAY DILUTION CURVES
    PLIKAYTIS, BD
    HOLDER, PF
    PAIS, LB
    MASLANKA, SE
    GHEESLING, LL
    CARLONE, GM
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 1994, 32 (10) : 2441 - 2447
  • [9] Comparative field study evaluating the activity of recombinant factor VIII Fc fusion protein in plasma samples at clinical haemostasis laboratories
    Sommer, J. M.
    Moore, N.
    McGuffie-Valentine, B.
    Bardan, S.
    Buyue, Y.
    Kamphaus, G. D.
    Konkle, B. A.
    Pierce, G. F.
    [J]. HAEMOPHILIA, 2014, 20 (02) : 294 - 300
  • [10] A world-wide survey and field study in clinical haemostasis laboratories to evaluate FVIII:C activity assay variability of ADYNOVATE and OBIZUR in comparison with ADVATE
    Turecek, P. L.
    Romeder-Finger, S.
    Apostol, C.
    Bauer, A.
    Crocker-Buque, A.
    Burger, D. A.
    Schall, R.
    Gritsch, H.
    [J]. HAEMOPHILIA, 2016, 22 (06) : 957 - 965