Experimental analysis of insertion torques and forces of threaded and press-fit acetabular cups by means of ex vivo and in vivo measurements

被引:7
作者
Vogel, Danny [1 ]
Rathay, Andreas [1 ]
Teufel, Stephanie [1 ]
Ellenrieder, Martin [1 ]
Zietz, Carmen [1 ]
Sander, Manuela
Bader, Rainer [1 ]
机构
[1] Univ Rostock, Biomech & Implant Technol Res Lab, Dept Orthopaed, Doberaner Str 142, D-18057 Rostock, Germany
关键词
insertion force; insertion torque; acetabular cup; press-fit; threaded; synthetic bone; TOTAL HIP-ARTHROPLASTY; PRIMARY STABILITY; FIXATION; BONE; DEFORMATION; COMPONENTS; SURFACE;
D O I
10.5277/ABB-00723-2016-02
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Purpose: In THA a sufficient primary implant stability is the precondition for successful secondary stability. Industrial foams of different densities have been used for primary stability investigations. The aim of this study was to analyse and compare the insertion behaviour of threaded and press-fit cups in vivo and ex vivo using bone substitutes with various densities. Methods: Two threaded (Bicon Plus (R), Trident (R) TC) and one press-fit cup (Trident PSL (R)) were inserted by orthopaedic surgeons (S1, S2) into 10, 20 and 31 pcf blocks, using modified surgical instruments allowing measurements of the insertion forces and torques. Furthermore, the insertion behaviour of two cups were analysed intraoperatively. Results: Torques for the threaded cups increased while bone substitute density increased. Maximum insertion torques were observed for S2 with 102 Nm for the Bicon Plus (R) in 20 pcf blocks and 77 Nm for the Trident (R) TC in 31 pcf blocks, which compares to the in vivo measurement (85 Nm). The average insertion forces for the press-fit cup varied from 5.2 to 6.8 kN (S1) and 7.2-11.5 kN (S2) ex vivo. Intraoperatively an average insertion force of 8.0 kN was determined. Conclusions: Implantation behaviour was influenced by acetabular cup design, bone substitute and experience of the surgeon. No specific density of bone substitute could be favoured for ex vivo investigations on the implantation behaviour of acetabular cups. The use synthetic bone blocks of high density (31 pcf) led to problems regarding cup orientation and seating. Therefore, bone substitutes used should be critically scrutinized in terms of the comparability to the in vivo situation.
引用
收藏
页码:155 / 163
页数:9
相关论文
共 25 条
[1]  
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[2]   Primary cup stability in THA with augmentation of acetabular defect. A comparison of healthy and osteoporotic bone [J].
Amirouche, F. ;
Solitro, G. ;
Broviak, S. ;
Goldstein, W. ;
Gonzalez, M. ;
Barmada, R. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2015, 101 (06) :667-673
[3]   Factors influencing initial cup stability in total hip arthroplasty [J].
Amirouche, Farid ;
Solitro, Giovanni ;
Broviak, Stefanie ;
Gonzalez, Mark ;
Goldstein, Wayne ;
Barmada, Riad .
CLINICAL BIOMECHANICS, 2014, 29 (10) :1177-1185
[4]   Primary stability of two uncemented acetabular components of different geometry: hemispherical or peripherally enhanced? [J].
Antoniades, G. ;
Smith, E. J. ;
Deakin, A. H. ;
Wearing, S. C. ;
Sarungi, M. .
BONE & JOINT RESEARCH, 2013, 2 (12) :264-269
[5]   A novel method to assess primary stability of press-fit acetabular cups [J].
Crosnier, Emilie A. ;
Keogh, Patrick S. ;
Miles, Anthony W. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (11) :1126-1134
[6]   THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[7]   Cementless total hip arthroplasty: a review [J].
Diehl, Peter ;
Haenle, Maximilian ;
Bergsehmidt, Philipp ;
Gollwitzer, Hans ;
Schauwecker, Johannes ;
Bader, Rainer ;
Mittelmeier, Wolfram .
BIOMEDIZINISCHE TECHNIK, 2010, 55 (05) :251-264
[8]   Screw-in performance of threaded cups [J].
Effenberger, H. ;
Boesl, O. ;
Witzel, U. ;
Linke, B. .
ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2008, 146 (02) :185-193
[9]  
Ellenrieder M, 2016, ACTA ORTHOP BELG, V82, P60
[10]   Experimental investigations of the insertion and deformation behavior of press-fit and threaded acetabular cups for total hip replacement [J].
Fritsche, Andreas ;
Bialek, Kristin ;
Mittelmeier, Wolfram ;
Simnacher, Martin ;
Fethke, Klaus ;
Wree, Andreas ;
Bader, Rainer .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2008, 13 (03) :240-247