Determination of cyclosporine, tacrolimus, sirolimus and everolimus by liquid chromatography coupled to electrospray ionization and tandem mass spectrometry: Assessment of matrix effects and assay performance

被引:20
|
作者
Vethe, Nils T. [1 ]
Gjerdalen, Laila C. [1 ]
Bergan, Stein [2 ,3 ]
机构
[1] Natl Hosp Norway, Oslo Univ Hosp, Dept Med Biochem, NO-0027 Oslo, Norway
[2] Oslo Univ Hosp, Dept Pharmacol, Oslo, Norway
[3] Univ Oslo, Sch Pharm, N-0316 Oslo, Norway
关键词
LC-MS/MS; mass spectrometry; matrix effects; immunosuppressants; cyclosporine; tacrolimus; sirolimus; everolimus; WHOLE-BLOOD; SIMULTANEOUS QUANTIFICATION; ROUTINE METHOD; MS/MS METHOD; HPLC-MS/MS; IMMUNOASSAY; PLASMA; DRUGS;
D O I
10.3109/00365513.2010.531141
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective. The immunosuppressants cyclosporine, tacrolimus, sirolimus and everolimus are used in rejection prophylaxis after transplantation. Liquid chromatography tandem mass spectrometry (LC-MS/MS) has become a widely used methodology for monitoring of the drug levels to ensure therapeutic exposure. The main objective of the study was to evaluate the existence and potential influence of matrix effects on LC-MS/MS measurements of the immunosuppressants in clinical blood samples. Methods. The samples were prepared by protein precipitation and thereafter analysed by reversed-phase chromatography coupled to MS/MS via an electrospray interface. Assay performance including within-and between-series imprecision and deviations from external controls were examined. Elution of overall matrix components and glycerophosphocholines were investigated. The MS/MS signals were monitored in post-column infusion experiments, and post-precipitation addition of compounds provided a basis for quantification of the matrix effects. The influence of matrix effects on assay performance was investigated after dilution of quality controls with blood from multiple individuals. Results. Between-series coefficients of variation were <= 5.1, <= 6.6, <= 11.0 and <= 7.4 %, and the mean deviations from external controls were -10.3, -6.7, 15.6 and 4.3% for cyclosporine, tacrolimus, sirolimus and everolimus, respectively. The elution of matrix components including glycerophosphocholines overlapped to some extent with the target compounds, and the average ion suppression ranged from 8.5-21%. However, the drugs and internal standards were correspondingly influenced. Conclusion. The internal standards consistently corrected the between-individual variability of matrix effects. These findings consolidate the reliability of the assay.
引用
收藏
页码:583 / 591
页数:9
相关论文
共 50 条
  • [21] Determination of porphyrins in oral bacteria by liquid chromatography electrospray ionization tandem mass spectrometry
    Jonas Fyrestam
    Nadja Bjurshammar
    Elin Paulsson
    Annsofi Johannsen
    Conny Östman
    Analytical and Bioanalytical Chemistry, 2015, 407 : 7013 - 7023
  • [22] Simultaneous determination of six alkaloids in blood and urine using a hydrophilic interaction liquid chromatography method coupled with electrospray ionization tandem mass spectrometry
    Wu, Ximei
    Huang, Weixiong
    Lu, Lin
    Lin, Li
    Yang, Xingfen
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 398 (03) : 1319 - 1327
  • [23] Collaborative investigation of matrix effects in mycotoxin determination by high performance liquid chromatography coupled to mass spectrometry
    Malachova, A.
    Sulyok, M.
    Schuhmacher, R.
    Berthiller, F.
    Hajslova, J.
    Veprikova, Z.
    Zachariasova, M.
    Lattanzio, V. M. T.
    De Saeger, S.
    Di Mavungu, J. Diana
    Malysheva, S. V.
    Biselli, S.
    Winkelmann, O.
    Breidbach, A.
    Hird, S.
    Krska, R.
    QUALITY ASSURANCE AND SAFETY OF CROPS & FOODS, 2013, 5 (02) : 91 - 103
  • [24] Electrospray ionization mass spectrometry of tetrodotoxin and its analogs: Liquid chromatography/mass spectrometry, tandem mass spectrometry, and liquid chromatography/tandem mass spectrometry
    Shoji, Y
    Yotsu-Yamashita, M
    Miyazawa, T
    Yasumoto, T
    ANALYTICAL BIOCHEMISTRY, 2001, 290 (01) : 10 - 17
  • [25] Identification and determination of coumateralyl and coumafuryl in animal tissues by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
    Jin, Mi-cong
    Xu, Guo-zhang
    Ren, Yi-ping
    Chen, Xiao-hong
    Xu, Xiao-ming
    JOURNAL OF APPLIED TOXICOLOGY, 2008, 28 (05) : 621 - 627
  • [26] Determination of soyasaponins Ba and Bb in human serum by high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry
    Jin, Micong
    Yang, Yiwen
    Su, Baogen
    Ren, Qilong
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 846 (1-2): : 169 - 175
  • [27] Simultaneous Determination of Eight Tranquilizers in Pork by Ultra-Performance Liquid Chromatography Coupled with Electrospray Tandem Mass Spectrometry
    Zhang, Liqun
    Wu, Pinggu
    Jin, Quan
    Ye, Haipeng
    Huang, Xihui
    Liu, Shaoying
    FOOD ANALYTICAL METHODS, 2017, 10 (02) : 354 - 362
  • [28] A validated assay for the quantitative analysis of tranexamic acid in human serum by liquid chromatography coupled with electrospray ionization mass spectrometry
    Grassin Delyle, Stanislas
    Abe, Emuri
    Batisse, Anne
    Tremey, Benjamin
    Fischler, Marc
    Devillier, Philippe
    Alvarez, Jean Claude
    CLINICA CHIMICA ACTA, 2010, 411 (5-6) : 438 - 443
  • [29] Quantitative determination of the diastereoisomers of hexabromocyclododecane in human plasma using liquid chromatography coupled with electrospray ionization tandem mass spectrometry
    Tang, Caiming
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (31): : 3317 - 3322
  • [30] Effective use of liquid chromatography-mass spectrometry (LC/MS) in the routine clinical laboratory for monitoring sirolimus, tacrolimus, and cyclosporine
    Poquette, MA
    Lensmeyer, GL
    Doran, TC
    THERAPEUTIC DRUG MONITORING, 2005, 27 (02) : 144 - 150