Quantifying image quality in chest computed tomography angiography: Evaluation of different contrast-to-noise ratio measurement methods

被引:3
|
作者
Chatzaraki, Vasiliki [1 ,2 ]
Kubik-Huch, Rahel A. [1 ]
Thali, Michael [2 ]
Niemann, Tilo [1 ]
机构
[1] Kantonsspital Baden, Inst Radiol, Ergel 1, CH-5404 Baden, Switzerland
[2] Univ Zurich, Inst Forens Med, Zurich, Switzerland
关键词
Computed tomography angiography; thorax; radiographic image enhancement; contrast to noise ratio; image enhancement; TUBE-VOLTAGE SELECTION; AORTIC CT ANGIOGRAPHY; ITERATIVE RECONSTRUCTION; PULMONARY-ARTERY; REDUCTION; OPACIFICATION; PROTOCOL; DETECTOR;
D O I
10.1177/02841851211041813
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Contrast-to-noise ratio is used to objectively evaluate image quality in chest computed tomography angiography (CTA). Different authors define and measure contrast-to-noise ratio using different methods. Purpose To summarize and evaluate the different contrast-to-noise ratio calculation formulas in the current literature. Material and Methods A systematic review of the recent literature for studies using contrast-to-noise ratio was performed. Contrast-to-noise ratio measurement methods reported by the different authors were recorded and reproduced in three patients who underwent chest CTA in our department for exploring variations among the different measurement methods. Results The search resulted in 109 articles, of which 26 were included. The studies involved 69 different measurements and overall, three different formula patterns. In all three, aorta and pulmonary arteries comprised the objects of interest in the numerator. In the denominator, standard deviation of the attenuation of the object of interest itself or of another background were used to reflect image noise. Some authors averaged the ratio values at different levels to obtain global ratio values. Using the object of interest itself for image noise calculation in the denominator compared to the usage of another background caused the most prominent variances of contrast-to-noise ratio between the two different protocols used for the reproduction of the measurements. Conclusion We recommend using the standard deviation of the attenuation of a background indicator as image noise rather than the object of interest itself for more reliable and comparative values. Global contrast-to-noise ratios based on averaging the values of different measurement levels should be avoided.
引用
收藏
页码:1353 / 1362
页数:10
相关论文
共 50 条
  • [1] Measurement of contrast-to-noise ratio for differential phase contrast computed tomography
    Zambelli, Joseph
    Bevins, Nicholas
    Qi, Zhihua
    Chen, Guang-Hong
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [2] Contrast-to-Noise Ratio Optimization in Coronary Computed Tomography Angiography: Validation in a Swine Model
    Hubbard, Logan
    Malkasian, Shant
    Zhao, Yixiao
    Abbona, Pablo
    Molloi, Sabee
    ACADEMIC RADIOLOGY, 2019, 26 (06) : E115 - E125
  • [3] The Contrast-to-Noise Ratio for Image Quality Evaluation in Scanning Electron Microscopy
    Timischl, F.
    SCANNING, 2015, 37 (01) : 54 - 62
  • [4] Mammographic image quality assessment by a contrast-to-noise ratio for clinical images
    Hill, Melissa L.
    Mackenzie, Alistair
    Warren, Lucy M.
    Khan, Nabeel
    Wang, Kaier
    Highnam, Ralph P.
    15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [5] Quantification of Radiographic Image Quality Based on Patient Anatomical Contrast-to-Noise Ratio - A Preliminary Study with Chest Images
    Lin, Yuan
    Wang, Xiaohui
    Sehnert, William J.
    Foos, David H.
    Barski, Lori
    Samei, Ehsan
    MEDICAL IMAGING 2010: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2010, 7627
  • [6] Contrast-to-noise ratio comparison between X-ray fluorescence emission tomography and computed tomography
    Debrosse, Hadley
    Jadick, Giavanna
    Meng, Ling Jian
    La Riviere, Patrick
    JOURNAL OF MEDICAL IMAGING, 2024, 11
  • [7] Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
    Mousavi-Gazafroudi, Seyedeh Shokoofeh
    Sajjadieh-Khajouei, Amirreza
    Moradi, Maryam
    Mousavi-Gazafroudi, Seyedeh Shabnam
    Yadegarfar, Ghasem
    Tavakoli, Mohammad Bagher
    ARYA ATHEROSCLEROSIS, 2019, 15 (05) : 205 - 210
  • [8] Radiation Dose Reduction With a Low-Tube Voltage Technique for Pediatric Chest Computed Tomographic Angiography Based on the Contrast-to-Noise Ratio Index
    Masuda, Takanori
    Funama, Yoshinori
    Nakaura, Takeshi
    Tahara, Masahiro
    Yamashita, Yukari
    Kiguchi, Masao
    Imada, Naoyuki
    Sato, Tomoyasu
    Awai, Kazuo
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2018, 69 (04): : 390 - 396
  • [9] Optimizing Image Quality and Dose for Digital Radiography of Distal Pediatric Extremities Using the Contrast-to-Noise Ratio
    Hess, R.
    Neitzel, U.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2012, 184 (07): : 643 - 649
  • [10] Image quality of spectral brain computed tomography angiography using halved dose of iodine contrast medium
    Fransson, Veronica
    Mellander, Helena
    Ramgren, Birgitta
    Andersson, Henrik
    Arena, Francesco
    Ydstroem, Kristina
    Ullberg, Teresa
    Wasselius, Johan
    NEURORADIOLOGY, 2023, 65 (09) : 1333 - 1342