Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method

被引:6
|
作者
Okahara, Shoko [1 ]
Handoh, Tetsuya [2 ]
Wakita, Mitsuru [2 ]
Yamamoto, Takamasa [2 ]
Misawa, Shigeki [2 ]
Miida, Takashi [2 ]
Sumikura, Hiroyuki [1 ]
机构
[1] Juntendo Univ Hosp, Dept Anesthesiol, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
[2] Juntendo Univ Hosp, Dept Clin Lab, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
关键词
Clauss method; FibCare; Fibrinogen; Hemorrhage; Point-of-care testing; TRAUMA; PLASMA;
D O I
10.1007/s00540-021-02987-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Timely fibrinogen replacement is key to treating critical hemorrhage. Measuring fibrinogen concentration by conventional laboratory tests requires centrifugation of blood samples and is often time-consuming. A point-of-care testing device (A&T, Yokohama, Japan), CG02N, has been available in Japan since 2011 to measure fibrinogen concentration without centrifugation. However, it has not been widely used as it requires dilution of blood samples using manual micropipetting. To further speed up and simplify the fibrinogen measurement, an improved device called FibCare (Atom Medical, Tokyo, Japan) was developed to avoid diluting blood samples. The purpose of this study is to verify the reliability of FibCare against laboratory measurement using the Clauss method. Fibrinogen concentrations with 60 sodium citrated whole blood samples were measured by both FibCare and Clauss methods in the laboratory. Measured values with the Clauss method were distributed in the 88-300 mg/dL range. By comparing these results, a significant positive correlation was observed between the FibCare and Clauss method (Y = 12.402 + 0.982 X; R = 0.891; P < 0.01). The study indicated that FibCare allows accurate measurement of fibrinogen concentration and shows a possibility to contribute to optimal fibrinogen replacement therapy during critical hemorrhage.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [41] Development of a hybridization chain reaction-powered lab-on-fiber device for ultrafast point-of-care testing of circulating tuor DNA in whole blood
    Xu, Jiaxin
    Han, Xiangzhi
    Xu, Wenjuan
    Liu, Jiayuan
    Zhu, Anna
    Song, Dan
    Long, Feng
    TALANTA, 2023, 259
  • [42] Measurement accuracy of two professional-use systems for point-of-care testing of blood glucose
    Baumstark, Annette
    Jendrike, Nina
    Kamecke, Ulrike
    Liebing, Christina
    Pleus, Stefan
    Freckmann, Guido
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (03) : 445 - 455
  • [43] A Comparison Study Between Point-of-Care Testing Systems and Central Laboratory for Determining Blood Glucose in Venous Blood
    Wei, Huiping
    Lan, Fang
    He, Qitian
    Li, Haiwei
    Zhang, Fuyong
    Qin, Xue
    Li, Shan
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (03)
  • [44] Measures of Blood Hemoglobin and Hematocrit During Cardiac Surgery: Comparison of Three Point-of-Care Devices
    Kolotiniuk, Nikolai V.
    Manecke, Gerald R.
    Pinsky, Michael R.
    Banks, Dalia
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2018, 32 (04) : 1638 - 1641
  • [45] Point-of-care detection of glycated hemoglobin using a novel dry chemistry-based electrochemiluminescence device
    Zhou, Xinya
    Lai, Wei
    Zhong, Jinbiao
    Yang, Yang
    Chen, Zhenyu
    Zhang, Chunsun
    ANALYTICA CHIMICA ACTA, 2023, 1279
  • [46] Evaluating the effect of sample type on American alligator (Alligator mississippiensis) analyte values in a point-of-care blood analyser
    Hamilton, Matthew T.
    Finger, John W., Jr.
    Winzeler, Megan E.
    Tuberville, Tracey D.
    CONSERVATION PHYSIOLOGY, 2016, 4
  • [47] Photoacoustic lateral flow assay for point-of-care testing of thrombin in human whole blood without background interference
    Guo, Lan
    Zhao, Dong-Mei
    Chen, Shuai
    Yu, Yong-Liang
    MICROCHEMICAL JOURNAL, 2025, 208
  • [48] Point-of-care blood clotting test and its correlation with fibrinogen level: Potential in goal-directed transfusion in postpartum hemorrhage
    Mishra, Nalini
    Ekka, Suma, V
    Pal, Sanjita
    Mishra, Ishan
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2021, 154 (02) : 343 - 351
  • [49] Point-of-Care Testing Blood Coagulation Detectors Using a Bio-Microfluidic Device Accompanied by Raman Spectroscopy
    Huang, Song-Jeng
    Chiang, Chao-Ching
    Immanuel, Philip Nathaniel
    Subramania, Murugan
    COATINGS, 2022, 12 (07)
  • [50] Evaluation of the Freestyle Optium Neo H point-of-care device for measuring blood glucose concentrations in sick calves
    Karapinar, Tolga
    Tumer, Kenan Cagri
    Buczinski, Sebastien
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2020, 34 (04) : 1650 - 1656