Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method

被引:6
|
作者
Okahara, Shoko [1 ]
Handoh, Tetsuya [2 ]
Wakita, Mitsuru [2 ]
Yamamoto, Takamasa [2 ]
Misawa, Shigeki [2 ]
Miida, Takashi [2 ]
Sumikura, Hiroyuki [1 ]
机构
[1] Juntendo Univ Hosp, Dept Anesthesiol, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
[2] Juntendo Univ Hosp, Dept Clin Lab, Bunkyo Ku, 3-1-3 Hongo, Tokyo 1138431, Japan
关键词
Clauss method; FibCare; Fibrinogen; Hemorrhage; Point-of-care testing; TRAUMA; PLASMA;
D O I
10.1007/s00540-021-02987-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Timely fibrinogen replacement is key to treating critical hemorrhage. Measuring fibrinogen concentration by conventional laboratory tests requires centrifugation of blood samples and is often time-consuming. A point-of-care testing device (A&T, Yokohama, Japan), CG02N, has been available in Japan since 2011 to measure fibrinogen concentration without centrifugation. However, it has not been widely used as it requires dilution of blood samples using manual micropipetting. To further speed up and simplify the fibrinogen measurement, an improved device called FibCare (Atom Medical, Tokyo, Japan) was developed to avoid diluting blood samples. The purpose of this study is to verify the reliability of FibCare against laboratory measurement using the Clauss method. Fibrinogen concentrations with 60 sodium citrated whole blood samples were measured by both FibCare and Clauss methods in the laboratory. Measured values with the Clauss method were distributed in the 88-300 mg/dL range. By comparing these results, a significant positive correlation was observed between the FibCare and Clauss method (Y = 12.402 + 0.982 X; R = 0.891; P < 0.01). The study indicated that FibCare allows accurate measurement of fibrinogen concentration and shows a possibility to contribute to optimal fibrinogen replacement therapy during critical hemorrhage.
引用
收藏
页码:757 / 760
页数:4
相关论文
共 50 条
  • [1] Fibrinogen measurement by a novel point-of-care device with whole blood: comparison of values against Clauss method
    Shoko Okahara
    Tetsuya Handoh
    Mitsuru Wakita
    Takamasa Yamamoto
    Shigeki Misawa
    Takashi Miida
    Hiroyuki Sumikura
    Journal of Anesthesia, 2021, 35 : 757 - 760
  • [2] The qLabs® FIB system, a novel point-of-care technology for a rapid and accurate quantification of functional fibrinogen concentration from a single drop of citrated whole blood
    Sanfilippo, Sandra
    Buisson, Leslie
    Rouabehi, Hafsa
    Dujaric, Marie-Eglantine
    Donnet, Thibault
    de Raucourt, Emmanuelle
    Dumont, Benedicte
    Peynaud-Debayle, Edith
    THROMBOSIS RESEARCH, 2023, 226 : 159 - 164
  • [3] A point-of-care microfluidic channel-based device for rapid and direct detection of fibrinogen in whole blood
    Ban, Qinan
    Zhang, Yulong
    Li, Yuxuan
    Cao, Daye
    Ye, Weifeng
    Zhan, Linsheng
    Wang, Daming
    Wang, Xiaohui
    LAB ON A CHIP, 2022, 22 (14) : 2714 - 2725
  • [4] A novel point-of-care device accurately measures thyrotropin in whole blood, capillary blood and serum
    Kahaly, George J.
    Lotz, Johannes
    Walder, Sara
    Hammad, Cara
    Kraemer, Rebecca
    Frommer, Lara
    Koenig, Jochem
    Wolf, Jan
    Gottwald-Hostalek, Ulrike
    Urgatz, Bogumila
    Lackner, Karl J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2022, 60 (10) : 1607 - 1616
  • [5] Point-of-care measurement of fetal blood lactate - Time to trust a new device
    Wang, Mandy
    Chua, Seng C.
    Bouhadir, Lilain
    Treadwell, Erin L.
    Gibbs, Emma
    McGee, Therese M.
    AUSTRALIAN & NEW ZEALAND JOURNAL OF OBSTETRICS & GYNAECOLOGY, 2018, 58 (01) : 72 - 78
  • [6] Standard vs. point-of-care measurement of fibrinogen: potential impact on clinical decisions
    Urwyler, N.
    Theiler, L.
    Hirschberg, M.
    Kleine-Brueggeney, M.
    Colucci, G.
    Greif, R.
    MINERVA ANESTESIOLOGICA, 2012, 78 (05) : 550 - 555
  • [7] Comparison of point-of-care device DiaSpect against the HemoCue and laboratory analyser in an ICU population
    Salmond, B.
    Rodseth, R. N.
    Farina, Z.
    von Randen, R. P.
    SOUTHERN AFRICAN JOURNAL OF ANAESTHESIA AND ANALGESIA, 2021, 27 (02) : 99 - 103
  • [8] Evaluation of a novel point-of-care test kit, ABSOGEN™ PCT, in semi-quantitative measurement of procalcitonin in whole blood
    Kwon, Hi Jeong
    Lee, Jehoon
    Park, Hae-il
    Han, Kyungja
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2017, 31 (06)
  • [9] Assessment of the reliability and validity of a novel point-of-care fibrinogen (F-Point) device against an industry standard at fibrinogen levels &gt;2 g/L in non-haemorrhage scenarios
    Katz, R.
    Efremov, V
    Mooney, C.
    El-Khuffash, A.
    Heaphy, L.
    Cosgrave, D.
    Loughrey, J.
    Thornton, P.
    INTERNATIONAL JOURNAL OF OBSTETRIC ANESTHESIA, 2020, 43 : 91 - 96
  • [10] Exploring the utility of a novel point-of-care whole blood thrombin generation assay following trauma: A pilot study
    Ferrara, Michael J.
    MacArthur, Taleen A.
    Butenas, Saulius
    Mann, Kenneth G.
    Immermann, Joseph M.
    Spears, Grant M.
    Bailey, Kent R.
    Kozar, Rosemary A.
    Heller, Stephanie F.
    Loomis, Erica A.
    Stephens, Daniel
    Park, Myung S.
    RESEARCH AND PRACTICE IN THROMBOSIS AND HAEMOSTASIS, 2021, 5 (03) : 395 - 402