Does an infra pectineal plate alone provide adequate fixation in anterior column posterior hemitransverse acetabular fractures? A comparative biomechanical study

被引:1
作者
Simsek, Ekin Kaya [1 ]
Haberal, Bahtiyar [2 ]
Mahmuti, Ates [3 ]
Balcik, Bedi Cenk [4 ]
Demirors, Huseyin [2 ]
机构
[1] Beypazari State Hosp, Dept Orthopaed & Traumatol, Ankara, Turkey
[2] Baskent Univ Hosp, Dept Orthopaed & Traumatol, Ankara, Turkey
[3] Silivri Ceza Infaz Kurumu State Hosp, Dept Orthopaed & Traumatol, Istanbul, Turkey
[4] Baskent Univ, Fac Engn, Mech Engn Program, Ankara, Turkey
来源
ULUSAL TRAVMA VE ACIL CERRAHI DERGISI-TURKISH JOURNAL OF TRAUMA & EMERGENCY SURGERY | 2022年 / 28卷 / 08期
关键词
Acetabulum; anterior column posterior hemi transverse; biomechanics; fracture; SCREW PLACEMENT; STRENGTH;
D O I
10.14744/tjtes.2021.99544
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: The purpose of this study is to compare biomechanical properties of suprapectineal (SP) plate fixation, infrapectineal (IP) plate fixation, and both SP and IP plate fixation in anterior column posterior hemitransverse (ACPHT) fractures of the acetabulum using posterior and anterior column screws. METHODS: In 21 hard plastic left hemipelvis models, ACPHT fractures of the acetabulum were created, and in three different fixation groups, the methods were compared: Group 1: SP plating using a 3.5 mm reconstruction plate and cortical screw fixation, Group 2: Infrapectineal plating using 3.5 mm reconstruction plate and cortical screws combined with posterior and anterior column screws, and Group 3: Combined fixation with SP and IP plating using 3.5 mm reconstruction plates and cortical screws. Maximum load to failure (strength) of these three groups was compared between groups. RESULTS: The mean maximum load of failure for three groups was 2921 N, 2018 N, and 3658 N, respectively. When strength was compared considering the force that causing implant failure, it was determined that the strongest fixation was achieved when SP and IP fixation method were applied together, followed by SP only fixation and IP fixation supported by anterior and posterior column screws, respectively. CONCLUSIONS: The combined application of SP and IP fixation provides the most stable fixation of the ACPHT acetabular fractures, and IP fixation does not provide comparable biomechanical stability despite reinforcement with three-column screws placed away from the plate. Although IP fixation supported by anterior and posterior column screws with the limited combined approach is less invasive approach for patients, SP fixation should be included in the surgical treatment method to ensure adequate stability.
引用
收藏
页码:1052 / 1058
页数:7
相关论文
共 16 条
[1]   Management of Acetabular Fractures in the Elderly [J].
Antell, Nicholas B. ;
Switzer, Julie A. ;
Schmidt, Andrew H. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2017, 25 (08) :577-585
[2]   Screw fixation of ACPHT acetabular fractures offers sufficient biomechanical stability when compared to standard buttress plate fixation [J].
Busuttil, Tatjana ;
Teuben, Michel ;
Pfeifer, Roman ;
Cinelli, Paolo ;
Pape, Hans-Christoph ;
Osterhoff, Georg .
BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1)
[3]   Biomechanical Comparison of Different Fixation Techniques for Typical Acetabular Fractures in the Elderly The Role of Special Quadrilateral Surface Buttress Plates [J].
Chen, Kaifang ;
Yang, Fan ;
Yao, Sheng ;
Xiong, Zekang ;
Sun, Tingfang ;
Guo, Xiaodong .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2020, 102 (14) :E81
[4]   Different stabilisation techniques for typical acetabular fractures in the elderly-A biomechanical assessment [J].
Culemann, U. ;
Holstein, J. H. ;
Koehler, D. ;
Tzioupis, C. C. ;
Pizanis, A. ;
Tosounidis, G. ;
Burkhardt, M. ;
Pohlemann, T. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2010, 41 (04) :405-410
[5]   Infra-Acetabular Corridor-Technical Tip for an Additional Screw Placement to Increase the Fixation Strength of Acetabular Fractures [J].
Culemann, Ulf ;
Marintschev, Ivan ;
Gras, Florian ;
Pohlemann, Tim .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2011, 70 (01) :244-246
[6]   Biomechanical Comparison of Intrapelvic and Extrapelvic Fixation for Acetabular Fractures Involving the Quadrilateral Plate [J].
Gillispie, Gregory J. ;
Babcock, Sharon N. ;
McNamara, Kyle P. ;
Dimoff, Michaela E. ;
Aneja, Arun ;
Brown, Philip J. ;
Carroll, Eben A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (11) :570-576
[7]   Screw- versus plate-fixation strength of acetabular anterior column fractures: A biomechanical study [J].
Gras, Florian ;
Marintschev, Ivan ;
Schwarz, Christoph E. ;
Hofmann, Gunther O. ;
Pohlemann, Tim ;
Culemann, Ulf .
JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2012, 72 (06) :1664-1670
[8]  
Hobbie R., 2007, Intermediate physics for medicine and biology
[9]   BIOMECHANICAL FIXATION STRENGTH COMPARISON OF ILIOINGUNIAL AND MEDIAL STOPPA APPROACHES ON ANTERIOR COLUMN FRACTURES [J].
Kacira, Burkay Kutluhan ;
Ozkaya, Mustafa ;
Kiran, Uygar ;
Turkmen, Faik ;
Arazi, Mehmet ;
Demir, Teyfik .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2016, 16 (06)
[10]   Are Quadrilateral Surface Buttress Plates Comparable to Traditional Forms of Transverse Acetabular Fracture Fixation? [J].
Kistler, Brian J. ;
Smithson, Ian R. ;
Cooper, Seth A. ;
Cox, Jacob L. ;
Nayak, Aniruddh N. ;
Santoni, Brandon G. ;
Sagi, H. Claude .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (11) :3353-3361