A readout circuit for an intra-ocular pressure sensor

被引:50
作者
Coosemans, J [1 ]
Catrysse, M [1 ]
Puers, R [1 ]
机构
[1] Katholieke Univ Leuven, Dept ESAT MICAS, B-3001 Heverlee, Belgium
关键词
intra-ocular pressure; passive telemetry; micro coils; readout circuit;
D O I
10.1016/j.sna.2003.09.015
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A readout circuit for a passive telemetric intra-ocular pressure (IOP) sensor is being developed. The intra-ocular sensor consists of capacitive pressure sensor in parallel with a planar coil. This inductor-capacitor (LC) resonant circuit transduces the pressure into a of resonance frequency. A voltage controlled oscillator (VCO) is used to excite the sensor over a large frequency range hereby detecting resonance of the internal sensor, and thus enabling the measurement of the intra-ocular pressure. This low power is extremely compact, making it suitable for long-term ambulant patient monitoring. The circuit allows wireless readout of the pressure transducers. Tests show promising results at mutual coil distances up to 7.5 mm. (C) 2003 Published by Elsevier B.V.
引用
收藏
页码:432 / 438
页数:7
相关论文
共 16 条
[1]   A wireless batch sealed absolute capacitive pressure sensor [J].
Akar, O ;
Akin, T ;
Najafi, K .
SENSORS AND ACTUATORS A-PHYSICAL, 2001, 95 (01) :29-38
[2]   PASSIVE SILICON TRANSENSOR INTENDED FOR BIOMEDICAL, REMOTE PRESSURE MONITORING [J].
BACKLUND, Y ;
ROSENGREN, L ;
HOK, B ;
SVEDBERGH, B .
SENSORS AND ACTUATORS A-PHYSICAL, 1990, 21 (1-3) :58-61
[4]   FASTHENRY - A MULTIPOLE-ACCELERATED 3-D INDUCTANCE EXTRACTION PROGRAM [J].
KAMON, M ;
TSUK, MJ ;
WHITE, JK .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1994, 42 (09) :1750-1758
[5]  
Kinget P, 1999, ANALOG CIRCUIT DESIGN, P353
[6]   Hermetically sealed inductor-capacitor (LC) resonator for remote pressure monitoring [J].
Park, EC ;
Yoon, JB ;
Yoon, E .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS BRIEF COMMUNICATIONS & REVIEW PAPERS, 1998, 37 (12B) :7124-7128
[7]   A CAPACITIVE PRESSURE SENSOR WITH LOW IMPEDANCE OUTPUT AND ACTIVE SUPPRESSION OF PARASITIC EFFECTS [J].
PUERS, B ;
PEETERS, E ;
VANDENBOSSCHE, A ;
SANSEN, W .
SENSORS AND ACTUATORS A-PHYSICAL, 1990, 21 (1-3) :108-114
[8]   Electrodeposited copper inductors for intraocular pressure telemetry [J].
Puers, R ;
Vandevoorde, G ;
De Bruyker, D .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2000, 10 (02) :124-129
[9]  
RAZAVI B, 2000, DESIGN ANALOG CMOS I, P505
[10]   RADIOFREQUENCY COILS IN IMPLANTABLE DEVICES - MISALIGNMENT ANALYSIS AND DESIGN PROCEDURE [J].
SOMA, M ;
GALBRAITH, DC ;
WHITE, RL .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (04) :276-282