Results with a cementless alumina-coated cup with dual mobility - A twelve-year follow-up study

被引:134
作者
Farizon, F [1 ]
de Lavison, R [1 ]
Azoulai, JJ [1 ]
Bousquet, G [1 ]
机构
[1] Exercise GIP Hop Bellevue, Univ Hosp Ctr St Etienne, Traumatol & Orthopaed Ctr, St Etienne, France
关键词
D O I
10.1007/s002640050246
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Between April 1980 and December 1981, 135 alumina-coated hemispherical clips were inserted without cement in patients with degenerative lesions. The clips were anchored by press fit and secured with a tripod attachment. The follow up was for over ten years. A bipolar polyethylene insert was used. Only one patient was lost to follow up in less than 2 years. Forty-four implants worked perfectly lip to the time of death of the patients. Four clips were replaced two after the patients sustained violent trauma, and there was an isolated instance of replacement of the insert only. The cumulative survival rate was 95.4% at 10 and 12 years follow up. The 78 surviving patients (85 cups) were examined, and radiographs were taken in 62 of them. Three patients showed clinical signs of deterioration and radiological signs of loosening, 2 others showed radiological signs of cup displacement without clinical symptoms. There was no evidence of osteolysis at the bone-cup interface in well functioning implants. These results confirm that cementless hemispherical cups survive well and demonstrate the value of a clip without holes for screws which can allow migration of polyethylene particles which might induce osteolysis of the adjacent ilum.
引用
收藏
页码:219 / 224
页数:6
相关论文
共 30 条
  • [11] SOCKET FIXATION USING A METAL-BACKED ACETABULAR COMPONENT FOR TOTAL HIP-REPLACEMENT - A MINIMUM 5-YEAR FOLLOW-UP
    HARRIS, WH
    WHITE, RE
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (05) : 745 - 748
  • [12] FURTHER FOLLOW-UP ON SOCKET FIXATION USING A METAL-BACKED ACETABULAR COMPONENT FOR TOTAL HIP-REPLACEMENT - A MINIMUM 10-YEAR FOLLOW-UP-STUDY
    HARRIS, WH
    PENENBERG, BL
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (08) : 1140 - 1143
  • [13] CLINICAL AND RADIOGRAPHIC EVALUATION OF TOTAL HIP-REPLACEMENT - A STANDARD SYSTEM OF TERMINOLOGY FOR REPORTING RESULTS
    JOHNSTON, RC
    FITZGERALD, RH
    HARRIS, WH
    POSS, R
    MULLER, ME
    SLEDGE, CB
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (02) : 161 - 168
  • [14] INVIVO WEAR OF POLYETHYLENE ACETABULAR COMPONENTS
    KABO, JM
    GEBHARD, JS
    LOREN, G
    AMSTUTZ, HC
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02): : 254 - 258
  • [15] DISLOCATION FOLLOWING TOTAL HIP-REPLACEMENT
    KHAN, MAA
    BRAKENBURY, PH
    REYNOLDS, ISR
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1981, 63 (02): : 214 - 218
  • [16] LECLERCQ S, 1995, REV CHIR ORTHOP, V81, P389
  • [17] Wear of the polyethylene liner-metallic shell interface in modular acetabular components - An in vitro analysis
    Lieberman, JR
    Kay, RM
    Hamlet, WP
    Park, SH
    Kabo, JM
    [J]. JOURNAL OF ARTHROPLASTY, 1996, 11 (05) : 602 - 608
  • [18] Massin P, 1989, J Arthroplasty, V4, P245, DOI 10.1016/S0883-5403(89)80020-8
  • [19] MCCOY TH, 1988, ORTHOP CLIN N AM, V19, P467
  • [20] TOTAL HIP-REPLACEMENT WITH INSERTION OF AN ACETABULAR COMPONENT WITHOUT CEMENT AND A FEMORAL COMPONENT WITH CEMENT - 4-YEAR TO 7-YEAR RESULTS
    MOHLER, CG
    KULL, LR
    MARTELL, JM
    ROSENBERG, AG
    GALANTE, JO
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (01) : 86 - 96