Multiple headspace single-drop microextraction coupled with gas chromatography for direct determination of residual solvents in solid drug product

被引:31
作者
Yu, Yingjia [1 ]
Chen, Bin [1 ]
Shen, Cidan [1 ]
Cai, Yi [1 ]
Xie, Meifen [1 ]
Zhou, Wei [1 ]
Chen, Yile [1 ]
Li, Yan [1 ]
Duan, Gengli [1 ]
机构
[1] Fudan Univ, Sch Pharm, Dept Pharmaceut Anal, Shanghai 201203, Peoples R China
关键词
Residual solvents; Multiple headspace single-drop microextraction; Gas chromatography; Drug product; LIQUID-PHASE MICROEXTRACTION; MASS-SPECTROMETRY; QUANTITATIVE-DETERMINATION; PHARMACEUTICAL PRODUCTS; BULK PHARMACEUTICALS; SEPARATION; COLUMN; WATER;
D O I
10.1016/j.chroma.2010.06.009
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper proposed a multiple headspace single-drop microextraction (MHS-SDME) method coupled to gas chromatography with Flame-ionization detection (GC-FID) for direct determination of residual solvents in solid drug product. The MHS-SDME technique is based on extrapolation to an exhaustive extraction of consecutive extractions from the same sample which eliminates the matrix effect on the quantitative analysis of solid samples. The total peak area of analyte is calculated with a beta constant which can be obtained from the slope of the linear regression that related to the peak area of each extraction and the number of extraction times. In this work, a model drug powder was chosen and the amounts of residues of two solvents, methanol and ethanol, were investigated. The factors influencing the extraction process including extraction solvent, microdrop volume, extraction time, sample amount, thermostatting temperature and incubation time were studied. 10 mg of drug powder was incubated for 3 h at 140 degrees C prior to the first extraction and thermostatted for 15 min at 140 degrees C between each extraction. Extraction was carried out with 2 mu L of dimethyl sulfoxide (DMSO) as the microdrop for 5 min. The features of the method were established using standard solutions. Validation of the proposed method showed good agreement with the traditional dissolution method for analysis of residual solvents in drug product. The results indicated that MHS-SDME has a great potential for the quantitative determination of residual solvents directly from the solid drug products due to its low cost, ease of operation, sensitivity, reliability and environmental protection. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:5158 / 5164
页数:7
相关论文
共 39 条