An automatic warning system for epileptic seizures recorded on intracerebral EEGs

被引:87
作者
Grewal, S [1 ]
Gotman, J [1 ]
机构
[1] McGill Univ, Montreal Neurol Inst, Montreal, PQ H3A 2B4, Canada
基金
加拿大健康研究院;
关键词
EEG; epilepsy; intracerebral electrodes; seizure detection; warning;
D O I
10.1016/j.clinph.2005.05.020
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: A new clinical seizure waning system for intracerebral EEG is proposed. It is aimed at a better performance than existing systems and at user tuneability. Methods: The system employs data filtering in multiple bands, spectral feature extraction, Bayes' theorem, and spatio-temporal analysis. The a priori information in Bayes' theorem was provided by 407 It of EEG from 19 patients having 152 seizures. Results: The testing data (19 patients, 389 h, 100 seizures, independent of the training data) yielded a sensitivity of 89.4%, a false detection rate of 0.22/h, and median delay time of 17.1 s when tuning was used, and 86%, 0.47/h and 16.2 s without tuning. Missed seizures were of short duration or had subtle seizure activity. False detections were caused by technical artefacts, non-epileptic large amplitude rhythmic bursts or very low amplitude activity. It was established that performance could easily be tuned. Results were also compared to the clinical system of Gotman (1990). Conclusions: The system offers a performance that is acceptable for clinical use. User tuneability allows for reduction in false detection with minimal loss to sensitivity. Significance: Epilepsy monitoring generates large amounts of recordings and requires intense observation. Automatic seizure detection and warning systems reduce review time and facilitate observation. We propose a method with high sensitivity and few false alarms. (c) 2005 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:2460 / 2472
页数:13
相关论文
共 18 条
  • [1] VERY HIGH-FREQUENCY RHYTHMIC ACTIVITY DURING SEEG SUPPRESSION IN FRONTAL-LOBE EPILEPSY
    ALLEN, PJ
    FISH, DR
    SMITH, SJM
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1992, 82 (02): : 155 - 159
  • [2] How well can epileptic seizures be predicted? An evaluation of a nonlinear method
    Aschenbrenner-Scheibe, R
    Maiwald, T
    Winterhalder, M
    Voss, HU
    Timmer, J
    Schulze-Bonhage, A
    [J]. BRAIN, 2003, 126 : 2616 - 2626
  • [3] Hippocampal and entorhinal cortex high-frequency oscillations (100-500 Hz) in human epileptic brain and in kainic acid-treated rats with chronic seizures
    Bragin, A
    Engel, J
    Wilson, CL
    Fried, I
    Mathern, GW
    [J]. EPILEPSIA, 1999, 40 (02) : 127 - 137
  • [4] FISHER RS, 1992, J CLIN NEUROPHYSIOL, V9, P441
  • [5] AUTOMATIC RECOGNITION OF EPILEPTIC SEIZURES IN THE EEG
    GOTMAN, J
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1982, 54 (05): : 530 - 540
  • [6] AUTOMATIC SEIZURE DETECTION - IMPROVEMENTS AND EVALUATION
    GOTMAN, J
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 76 (04): : 317 - 324
  • [7] Automatic seizure detection in the newborn: methods and initial evaluation
    Gotman, J
    Flanagan, D
    Zhang, J
    Rosenblatt, B
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1997, 103 (03): : 356 - 362
  • [8] GOTMAN J, 2004, HAND CLIN N, V3, P155
  • [9] AN AUTOMATED SEIZURE MONITORING-SYSTEM FOR PATIENTS WITH INDWELLING RECORDING ELECTRODES
    HARDING, GW
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1993, 86 (06): : 428 - 437
  • [10] Wavelet based automatic seizure detection in intracerebral electroencephalogram
    Khan, YU
    Gotman, J
    [J]. CLINICAL NEUROPHYSIOLOGY, 2003, 114 (05) : 898 - 908