Integration of an RF coil and commercial field camera for ultrahigh-field MRI

被引:14
作者
Gilbert, Kyle M. [1 ,2 ]
Dubovan, Paul, I [1 ,2 ]
Gati, Joseph S. [1 ,2 ]
Menon, Ravi S. [1 ,2 ]
Baron, Corey A. [1 ,2 ]
机构
[1] Univ Western Ontario, Ctr Funct & Metab Mapping, 1151 Richmond St North, London, ON N6A 5B7, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON, Canada
基金
加拿大自然科学与工程研究理事会; 加拿大创新基金会;
关键词
field monitoring; functional imaging; magnetic resonance imaging; radiofrequency coil; ultrahigh field; DIFFUSION MRI; STEADY-STATE; NMR PROBES; BRAIN; RECONSTRUCTION; COMPENSATION; FLUCTUATIONS; ARRAY; ARTIFACTS; SYSTEM;
D O I
10.1002/mrm.29130
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To develop an RF coil with an integrated commercial field camera for ultrahigh field (7T) neuroimaging. The RF coil would operate within a head-only gradient coil and be subject to the corresponding design constraints. The RF coil can thereafter be used for subject-specific correction of k-space trajectories-notably in gradient-sensitive sequences such as single-shot spiral imaging. Methods The transmit and receive performance was evaluated before and after the integration of field probes, whereas field probes were evaluated when in an optimal configuration external to the coil and after their integration. Diffusion-weighted EPI and single-shot spiral acquisitions were employed to evaluate the efficacy of correcting higher order field perturbations and the consequent effect on image quality. Results Field probes had a negligible effect on RF-coil performance, including the transmit efficiency, transmit uniformity, and mean SNR over the brain. Modest reductions in field-probe signal lifetimes were observed, caused primarily by nonidealities in the gradient and shim fields of the head-only gradient coil at the probe positions. The field-monitoring system could correct up to second-order field perturbations in single-shot spiral imaging. Conclusion The integrated RF coil and field camera was capable of concurrent-field monitoring within a 7T head-only scanner and facilitated the subsequent correction of k-space trajectories during spiral imaging.
引用
收藏
页码:2551 / 2565
页数:15
相关论文
共 64 条
[21]   Radiofrequency coil for routine ultra-high-field imaging with an unobstructed visual field [J].
Gilbert, Kyle M. ;
Klassen, L. Martyn ;
Mashkovtsev, Alexander ;
Zeman, Peter ;
Menon, Ravi S. ;
Gati, Joseph S. .
NMR IN BIOMEDICINE, 2021, 34 (03)
[22]   A conformal transceive array for 7 T neuroimaging [J].
Gilbert, Kyle M. ;
Belliveau, Jean-Guy ;
Curtis, Andrew T. ;
Gati, Joseph S. ;
Klassen, L. Martyn ;
Menon, Ravi S. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 67 (05) :1487-1496
[23]   A radiofrequency coil to facilitate B1+ shimming and parallel imaging acceleration in three dimensions at 7 T [J].
Gilbert, Kyle M. ;
Curtis, Andrew T. ;
Gati, Joseph S. ;
Klassen, L. Martyn ;
Menon, Ravi S. .
NMR IN BIOMEDICINE, 2011, 24 (07) :815-823
[24]   AN ACTIVE TRANSMIT/RECEIVE NMR MAGNETOMETER FOR FIELD MONITORING IN ULTRA HIGH FIELD MRI SCANNERS [J].
Handwerker, J. ;
Bonehi, V ;
Eschelbach, M. ;
Scheffler, K. ;
Ortmanns, M. ;
Anders, J. .
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2013, 58
[25]   Safety testing and operational procedures for self-developed radiofrequency coils [J].
Hoffmann, Jens ;
Henning, Anke ;
Giapitzakis, Ioannis A. ;
Scheffler, Klaus ;
Shajan, G. ;
Pohmann, Rolf ;
Avdievich, Nikolai I. .
NMR IN BIOMEDICINE, 2016, 29 (09) :1131-1144
[26]  
International Electrotechnical Commission, 2012, MED EL EQ 1
[27]  
International Electrotechnical Commission, 2013, MED EL EQ 2
[28]   A rapid and robust gradient measurement technique using dynamic single-point imaging [J].
Jang, Hyungseok ;
McMillan, Alan B. .
MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (03) :950-962
[29]   Rapid anatomical brain imaging using spiral acquisition and an expanded signal model [J].
Kasper, Lars ;
Engel, Maria ;
Barmet, Christoph ;
Haeberlin, Maximilian ;
Wilm, Bertram J. ;
Dietrich, Benjamin E. ;
Schmid, Thomas ;
Gross, Simon ;
Brunner, David O. ;
Stephan, Klaas E. ;
Pruessmann, Klaas P. .
NEUROIMAGE, 2018, 168 :88-100
[30]   A 64-channel 3T array coil for accelerated brain MRI [J].
Keil, Boris ;
Blau, James N. ;
Biber, Stephan ;
Hoecht, Philipp ;
Tountcheva, Veneta ;
Setsompop, Kawin ;
Triantafyllou, Christina ;
Wald, Lawrence L. .