A Low-Pressure Oxygen Storage System for Oxygen Supply in Low-Resource Settings

被引:15
作者
Rassool, Roger P. [1 ,2 ]
Sobott, Bryn A. [1 ,2 ]
Peake, David J. [1 ]
Mutetire, Bagayana S. [3 ]
Moschovis, Peter P. [4 ]
Black, Jim F. P. [1 ,5 ]
机构
[1] FREO2 Fdn Australia, Melbourne, Vic, Australia
[2] Univ Melbourne, Sch Phys, Melbourne, Vic, Australia
[3] Mbarara Reg Referral Hosp, Mbarara, Uganda
[4] Massachusetts Gen Hosp, Boston, MA 02114 USA
[5] Univ Melbourne, Nossal Inst Global Hlth, Melbourne, Vic, Australia
关键词
pneumonia; hypoxia; LMIC; oxygen concentrator; infant; child; preschool; PAPUA-NEW-GUINEA; SEVERE PNEUMONIA; IMPLEMENTATION; CONCENTRATORS; COUNTRIES; GAMBIA;
D O I
10.4187/respcare.05532
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Widespread access to medical oxygen would reduce global pneumonia mortality. Oxygen concentrators are one proposed solution, but they have limitations, in particular vulnerability to electricity fluctuations and failure during blackouts. The low-pressure oxygen storage system addresses these limitations in low-resource settings. This study reports testing of the system in Melbourne, Australia, and nonclinical field testing in Mbarara, Uganda. METHODS: The system included a power-conditioning unit, a standard oxygen concentrator, and an oxygen store. In Melbourne, pressure and flows were monitored during cycles of filling/emptying, with forced voltage fluctuations. The bladders were tested by increasing pressure until they ruptured. In Mbarara, the system was tested by accelerated cycles of filling/emptying and then run on grid power for 30 d. RESULTS: The low-pressure oxygen storage system performed well, including sustaining a pressure approximately twice the standard working pressure before rupture of the outer bag. Flow of 1.2 L/min was continuously maintained to a simulated patient during 30 d on grid power, despite power failures totaling 2.9% of the total time, with durations of 1-176 min (mean 36.2, median 18.5). CONCLUSIONS: The low-pressure oxygen storage system was robust and durable, with accelerated testing equivalent to at least 2 y of operation revealing no visible signs of imminent failure. Despite power cuts, the system continuously provided oxygen, equivalent to the treatment of one child, for 30 d under typical power conditions for sub-Saharan Africa. The low-pressure oxygen storage system is ready for clinical field trials.
引用
收藏
页码:1582 / 1587
页数:6
相关论文
共 12 条
  • [1] [Anonymous], 2015, MILLENNIUM DEV GOALS
  • [2] Implementation and 8-year follow-up of an uninterrupted oxygen supply system in a hospital in The Gambia
    Bradley, B. D.
    Light, J. D.
    Ebonyi, A. O.
    N'Jai, P. C.
    Ideh, R. C.
    Ebruke, B. E.
    Nyassi, E.
    Peel, D.
    Howie, S. R. C.
    [J]. INTERNATIONAL JOURNAL OF TUBERCULOSIS AND LUNG DISEASE, 2016, 20 (08) : 1130 - 1134
  • [3] Child Trends, 2015, Attitudes Toward Spanking."
  • [4] Duke T, 2001, INT J TUBERC LUNG D, V5, P511
  • [5] Paediatric emergency and acute care in resource poor settings
    Duke, Trevor
    Cheema, Baljit
    [J]. JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2016, 52 (02) : 221 - 226
  • [6] Duke Trevor, 2010, Papua New Guinea Medical Journal, V53, P126
  • [7] Implementation of an oxygen concentrator system in district hospital paediatric wards throughout Malawi
    Enarson, Penny
    La Vincente, Sophie
    Gie, Robert
    Maganga, Ellubey
    Chokani, Codewell
    [J]. BULLETIN OF THE WORLD HEALTH ORGANIZATION, 2008, 86 (05) : 344 - 348
  • [8] Hill SE, 2009, INT J TUBERC LUNG D, V13, P587
  • [9] The functioning of oxygen concentrators in resource-limited settings: a situation assessment in two countries
    La Vincente, S. F.
    Peel, D.
    Carai, S.
    Weber, M. W.
    Enarson, P.
    Maganga, E.
    Soyolgerel, G.
    Duke, T.
    [J]. INTERNATIONAL JOURNAL OF TUBERCULOSIS AND LUNG DISEASE, 2011, 15 (05) : 693 - 699
  • [10] Evaluation of oxygen concentrators for use in countries with limited resources
    Peel, D.
    Neighbour, R.
    Eltringham, R. J.
    [J]. ANAESTHESIA, 2013, 68 (07) : 706 - 712