ISFET-based differential pCO(2) sensors employing a low-resistance gas-permeable membrane

被引:37
作者
Shin, JH [1 ]
Lee, HJ [1 ]
Kim, CY [1 ]
Oh, BK [1 ]
Rho, KL [1 ]
Nam, H [1 ]
Cha, GS [1 ]
机构
[1] KWANGWOON UNIV,DEPT CHEM,NOWAN KU,SEOUL 139701,SOUTH KOREA
关键词
D O I
10.1021/ac960473h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new type of ISFET-based differential pCO(2) sensor system is devised by employing a low-resistance gas-permeable membrane, Both the pCO(2)-FET and the reference FET (REFET) are made by coating their gate surfaces with an RTV silicone rubber film doped with valinomycin. The pCO(2) FET is covered with a hydrogel-based recipient layer inside the gas-permeable RTV membrane, but the REFET does not. include the hydrogel layer, Because of low resistance of the gas-permeable membrane, the reference electrode can be placed outside as a common reference for both FETs. Output signals from the pCO(2)-FET and the REFET are measured against the common reference, and the difference of these two signals is measured through a differential amplifier; therefore, calibration drifts of the two identical base ISFETs, caused by their thermal sensitivity, for example, cancel out It is also shown that the ion response or the potential drifts caused by both parts of the RTV membranes cancel out exactly the same way as in the case of the base ISFETs.
引用
收藏
页码:3166 / 3172
页数:7
相关论文
共 24 条
[21]   MULTIION SENSING SYSTEM BASED ON GLASS-ENCAPSULATED PH-ISFETS AND A PSEUDO-REFET [J].
VANDENVLEKKERT, HH ;
DEROOIJ, NF ;
VANDENBERG, A ;
GRISEL, A .
SENSORS AND ACTUATORS B-CHEMICAL, 1990, 1 (1-6) :395-400
[22]  
VANDERSCHOOT BH, 1987, BIOSENSORS, V3, P161
[23]   NEW MEMBRANE MATERIALS FOR POTASSIUM-SELECTIVE ION-SENSITIVE FIELD-EFFECT TRANSISTORS [J].
VANDERWAL, PD ;
SKOWRONSKAPTASINSKA, M ;
VANDENBERG, A ;
BERGVELD, P ;
SUDHOLTER, EJR ;
REINHOUDT, DN .
ANALYTICA CHIMICA ACTA, 1990, 231 (01) :41-52
[24]   POLYMER MEMBRANE-BASED ION-SELECTIVE, GAS-SELECTIVE AND BIO-SELECTIVE POTENTIOMETRIC SENSORS [J].
YIM, HS ;
KIBBEY, CE ;
MA, SC ;
KLIZA, DM ;
LIU, D ;
PARK, SB ;
TORRE, CE ;
MEYERHOFF, ME .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (01) :1-38