Effect of the atrial blanking time on the detection of atrial fibrillation in dual chamber pacing

被引:12
|
作者
Nowak, B
Kracker, S
Rippin, G
Horstick, G
Vincent, A
Geil, S
Himmrich, E
Meyer, J
机构
[1] Univ Mainz, Med Clin 2, D-6500 Mainz, Germany
[2] Vitatron Inc, Cologne, Germany
[3] Univ Mainz, Inst Med Stat, D-6500 Mainz, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 04期
关键词
atrial fibrillation; atrial blanking time; atrial sensing; ventricular far-field sensing;
D O I
10.1046/j.1460-9592.2001.00496.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with paroxysmal atrial fibrillation (PAF) and dual chamber pacemakers frequently have short postventricular atrial blanking times and sensitive atrial sensing thresholds used to provide reliable detection and mode switching during AF. However, short atrial blanking times increase the risk of atrial sensing of ventricular far-field signals. We evaluated if the length of the atrial blanking time influences the detection of AF. The study included ten patients with a VDDR (n = 7) or DDDR system (n = 3), who presented with AF at 18 follow-up visits. Bipolar atrial sensing was programmed to the most sensitive value. Atrial blanking times were programmed from 100 to 200 ms in 25-ms steps in each patient. Using marker annotation, the following parameters were measured at ten consecutive ventricular beats: VAF = the interval between ventricular events; and XAF = the interpolated number of atrial-sensed atrial-sensed events between two ventricular events; and XAF = the interpolated number of atrial-sensed events during atrial blanking time. The intervals between ventricular events and between atrial-sensed event markers showed no significant differences for the five blanking times tested. There was no significant influence of the atrial blanking time onto the measured parameters (least square means +/- standard error) with VAF between 281 +/- 12 and 300 +/- 12 ms (P = NS), AFs between 3.4 +/- 0.2 and 3.6 +/- 0.2 beats (P = NS) and XAF between 1.84 +/- 0.12 and 2.03 +/- 0.12 beats (P = NS). At ventricular rates < 100/min, the atrial sensing of AF in dual chamber pacemakers demonstrated no evidence for deterioration by an increase of the atrial blanking time from 100 to 200 ms. Thus, the risk of ventricular far-field sensing may be reduced without compromising atrial sensing.
引用
收藏
页码:496 / 499
页数:4
相关论文
共 50 条
  • [31] Effect of renal sympathetic denervation on the inducibility of atrial fibrillation during rapid atrial pacing
    Qingyan Zhao
    Shengbo Yu
    Minghui Zou
    Zixuan Dai
    Xule Wang
    Jinping Xiao
    Congxin Huang
    Journal of Interventional Cardiac Electrophysiology, 2012, 35 : 119 - 125
  • [32] Evaluation of bi-atrial pacing and single site right atrial pacing for the prevention of atrial fibrillation
    Enjoji, Y
    Sugi, K
    Noro, M
    Nakae, T
    Tezuka, N
    Takami, M
    Kawase, A
    Kumagai, K
    Sakai, T
    Ikeda, T
    Yamaguchi, T
    CIRCULATION JOURNAL, 2002, 66 (01) : 70 - 74
  • [33] Inflammatory cytokines in cardiac pacing patients with atrial fibrillation and asymptomatic atrial fibrillation
    Pan, Jiaxi
    Wang, Weiwei
    Wu, Xiaoxue
    Kong, Fanqi
    Pan, Jialin
    Lin, Jiafeng
    Zhang, Minxue
    PANMINERVA MEDICA, 2018, 60 (03) : 86 - 91
  • [34] Effect of atrial overdrive pacing on pulmonary vein focal discharge in patients with atrial fibrillation
    Arentz, T
    Ott, P
    von Rosenthal, J
    Blum, T
    Kalusche, D
    EUROPACE, 2003, 5 (01): : 25 - 31
  • [35] Optimizing transesophageal atrial pacing in mice to detect atrial fibrillation
    Murphy, Matthew B.
    Kim, Kyungsoo
    Kannankeril, Prince J.
    Subati, Tuerdi
    Van Amburg, Joseph C.
    Barnett, Joey, V
    Murray, Katherine T.
    AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2022, 322 (01): : H36 - H43
  • [36] Sensing of atrial fibrillation by a dual chamber pacemaker: How should atrial sensing be programmed to ensure adequate mode shifting?
    Walfridsson, H
    Aunes, M
    Capocci, M
    Edvardsson, N
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (07): : 1089 - 1093
  • [37] Atrial septal pacing: A new approach to prevent atrial fibrillation
    Padeletti, L
    Michelucci, A
    Pieragnoli, P
    Colella, A
    Musilli, N
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (06): : 850 - 854
  • [38] Efficacy and tolerability of continuous overdrive atrial pacing in atrial fibrillation
    Lam, CTF
    Lau, CP
    Leung, SK
    Tse, HF
    Lee, KLF
    Tang, MO
    Tsang, V
    EUROPACE, 2000, 2 (04): : 286 - 291
  • [39] Atrial Burst Pacing with Biphasic and Monophasic Waveforms for Atrial Fibrillation
    Barsheshet, Alon
    Wakslak, Menachem
    Mower, Morton M.
    Goldenberg, Ilan
    Hall, Burr
    ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2012, 17 (01) : 22 - 27
  • [40] Pacing therapy for prevention of atrial fibrillation
    Ellenbogen, Kenneth A.
    HEART RHYTHM, 2007, 4 (03) : S84 - S87