Effect of the atrial blanking time on the detection of atrial fibrillation in dual chamber pacing

被引:12
|
作者
Nowak, B
Kracker, S
Rippin, G
Horstick, G
Vincent, A
Geil, S
Himmrich, E
Meyer, J
机构
[1] Univ Mainz, Med Clin 2, D-6500 Mainz, Germany
[2] Vitatron Inc, Cologne, Germany
[3] Univ Mainz, Inst Med Stat, D-6500 Mainz, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2001年 / 24卷 / 04期
关键词
atrial fibrillation; atrial blanking time; atrial sensing; ventricular far-field sensing;
D O I
10.1046/j.1460-9592.2001.00496.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with paroxysmal atrial fibrillation (PAF) and dual chamber pacemakers frequently have short postventricular atrial blanking times and sensitive atrial sensing thresholds used to provide reliable detection and mode switching during AF. However, short atrial blanking times increase the risk of atrial sensing of ventricular far-field signals. We evaluated if the length of the atrial blanking time influences the detection of AF. The study included ten patients with a VDDR (n = 7) or DDDR system (n = 3), who presented with AF at 18 follow-up visits. Bipolar atrial sensing was programmed to the most sensitive value. Atrial blanking times were programmed from 100 to 200 ms in 25-ms steps in each patient. Using marker annotation, the following parameters were measured at ten consecutive ventricular beats: VAF = the interval between ventricular events; and XAF = the interpolated number of atrial-sensed atrial-sensed events between two ventricular events; and XAF = the interpolated number of atrial-sensed events during atrial blanking time. The intervals between ventricular events and between atrial-sensed event markers showed no significant differences for the five blanking times tested. There was no significant influence of the atrial blanking time onto the measured parameters (least square means +/- standard error) with VAF between 281 +/- 12 and 300 +/- 12 ms (P = NS), AFs between 3.4 +/- 0.2 and 3.6 +/- 0.2 beats (P = NS) and XAF between 1.84 +/- 0.12 and 2.03 +/- 0.12 beats (P = NS). At ventricular rates < 100/min, the atrial sensing of AF in dual chamber pacemakers demonstrated no evidence for deterioration by an increase of the atrial blanking time from 100 to 200 ms. Thus, the risk of ventricular far-field sensing may be reduced without compromising atrial sensing.
引用
收藏
页码:496 / 499
页数:4
相关论文
共 50 条
  • [21] Atrial conduction velocity and risk of recurrent atrial fibrillation after ablation: Time to blank the blanking period?
    John, Roy M.
    Mitra, Raman
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (08) : 1950 - 1952
  • [22] Design and Implementation of the Dual Site Atrial Pacing to Prevent Atrial Fibrillation (DAPPAF) Clinical Trial
    Stephanie M Fitts
    Michael R.S. Hill
    Rahul Mehra
    Paul Friedman
    Stephen Hammill
    G. Neal Kay
    Atul Prakash
    Charles Webb
    Sanjeev Saksena
    Journal of Interventional Cardiac Electrophysiology, 1998, 2 : 139 - 144
  • [23] Detection of atrial fibrillation and flutter by a dual-chamber implantable cardioverter-defibrillator
    Swerdlow, CD
    Schsls, W
    Dijkman, B
    Jung, W
    Sheth, NV
    Olson, WH
    Gunderson, BD
    CIRCULATION, 2000, 101 (08) : 878 - 885
  • [24] Design and implementation of the Dual Site Atrial Pacing to Prevent Atrial Fibrillation (DAPPAF) clinical trial
    Fitts, SM
    Hill, MRS
    Mehra, R
    Friedman, P
    Hammill, S
    Kay, GN
    Prakash, A
    Webb, C
    Saksena, S
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (02) : 139 - 144
  • [25] Paradoxical atrial undersensing due to inappropriate atrial noise reversion of atrial fibrillation in dual-chamber pacemakers
    Kolb, C
    Halbfass, P
    Zrenner, B
    Schmitt, C
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (07) : 696 - 700
  • [26] Permanent atrial fibrillation in patients with a dual-chamber pacemaker
    Debski, Maciej
    Ulman, Mateusz
    Zabek, Andrzej
    Boczar, Krzysztof
    Haberka, Kazimierz
    Kuniewicz, Marcin
    Lelakowski, Jacek
    Malecka, Barbara
    KARDIOLOGIA POLSKA, 2019, 77 (12) : 1140 - 1146
  • [27] Dynamic and dual-site atrial pacing in the prevention of atrial fibrillation: The STimolazione atrial DInamica Multisito (STADIM) study
    De Simone, Antonio
    Senatore, Gaetano
    Donnici, Giovanni
    Turco, Pietro
    Romano, Enrico
    Gazzola, Carlo
    Stabile, G.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 : S71 - S74
  • [28] Left atrial vein pacing: A technique of biatrial pacing for the prevention of atrial fibrillation
    Birnie, D
    Connors, SP
    Veinot, JP
    Green, M
    Stinson, WA
    Tang, ASL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (02): : 240 - 245
  • [29] Effect of atrial pacing on ventricular rate during atrial fibrillation. A human study
    Capucci, Alessandro
    Villani, Giovanni Quinto
    Igel, David
    Marotta, Tiziana
    JOURNAL OF CARDIOVASCULAR MEDICINE, 2008, 9 (03) : 256 - 262
  • [30] Effect of renal sympathetic denervation on the inducibility of atrial fibrillation during rapid atrial pacing
    Zhao, Qingyan
    Yu, Shengbo
    Zou, Minghui
    Dai, Zixuan
    Wang, Xule
    Xiao, Jinping
    Huang, Congxin
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2012, 35 (02) : 119 - 125