Characteristics of the absorbed dose to water standard at ENEA

被引:20
作者
Guerra, AS
Laitano, RF
Pimpinella, M
机构
[1] Ist. Naz. Metrologia delle R., Dipartimento Ambiente, Centro Ricerche Casaccia, 00100 Roma AD
关键词
D O I
10.1088/0031-9155/41/4/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The primary standard of absorbed dose to water established at ENEA for the Co-60 gamma-ray quality is based on a graphite calorimeter and an ionometric transfer system. This standard was recently improved after a more accurate assessment of some perturbation effects in the calorimeter and a modification of the water phantom shape and size. The conversion procedure requires two corresponding depths, one in graphite and one in water, where the radiation energy spectra must be the same. The energy spectra at the corresponding points were determined by a Monte Carlo simulation in water and graphite scaled phantoms. A thorough study of the calorimeter gap effect corrections was also made with regard to their dependence on depth and field size. A comparison between the ionization chamber calibration procedures based on the standards of absorbed dose to water and of air kerma was also made, confirming the consistency of the two methods.
引用
收藏
页码:657 / 674
页数:18
相关论文
共 29 条
[1]   ABSORBED DOSE BEAM QUALITY FACTORS FOR THE DOSIMETRY OF HIGH-ENERGY PHOTON BEAMS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (12) :2189-2211
[2]   UNCERTAINTIES IN DOSIMETRIC DATA AND BEAM CALIBRATION [J].
ANDREO, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (05) :1233-1247
[3]  
[Anonymous], 1983, MED PHYS, V10, P741
[4]   PARTITION OF KERMA TO ACCOUNT FOR BREMSSTRAHLUNG [J].
ATTIX, FH .
HEALTH PHYSICS, 1979, 36 (03) :347-354
[7]  
BOUTILLON M, 1993, CCEMRI9333 BIPM COM
[8]   ABSORBED-DOSE CALIBRATIONS IN HIGH-ENERGY PHOTON BEAMS AT THE NATIONAL-PHYSICAL-LABORATORY - CONVERSION PROCEDURE [J].
BURNS, JE .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (10) :1555-1575
[9]  
BURNS JE, 1990, 7 NPL RSA EXT NAT PH
[10]   DISPLACEMENT CORRECTIONS USED IN ABSORBED DOSE DETERMINATION [J].
CUNNINGHAM, JR ;
SONTAG, MR .
MEDICAL PHYSICS, 1980, 7 (06) :672-676