Accurate ultrasound of carotid intima-media layer by pixel compounding

被引:0
作者
Yang, Z [1 ]
Raunig, D [1 ]
Tuthill, T [1 ]
Fox, MD [1 ]
Analoui, M [1 ]
机构
[1] Univ Connecticut, Storrs, CT 06269 USA
来源
Medical Imaging 2005: Ultrasonic Imaging and Signal Processing | 2005年 / 5750卷
关键词
ultrasound image; pixel compounding; anisotropic diffusion; super-resolution reconstruction; homomorphic transformation; carotid artery; IMT;
D O I
10.1117/12.597680
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The intima-media thickness (IMT) of the carotid artery is an important biomarker for the clinical prognosis and diagnosis of atherosclerosis and stroke. This paper presents a new approach, pixel compounding, to enhance the resolution of the intima-media vascular layers in ultrasound B-scan images and provide increased image resolution for a more precise measurement. First, homomorphic transformation is used to estimate the lumped point spread function (PSF) of the images, then, the images are deblurred with the estimated PSF, and finally, a non-homogeneous anisotropic diffusion algorithm is used to further enhance the resolution of the image. The homogeneous part of the algorithm is used to suppress speckle while enhancing the coherent structures, specifically the edges; the non-homogeneous part (likelihood estimator) progressively adds the details from succeeding frames in the image sequence for an optimal and sub-pixel resolved image. Phantom studies have shown 300% improvement on Peak Distance Standard Deviation and nearly 100% improvement on Average Half Peak Width, indicating significant resolution enhancement.
引用
收藏
页码:281 / 288
页数:8
相关论文
共 15 条
[1]  
ABDELMONIEM KZ, 2002, IEEE T BIOMEDICAL EN, V49
[2]  
BIGGS DSC, 1997, APPL OPTICS, V36
[3]  
Gustavsson T, 1997, IEEE COMPUTERS CARDI, V24
[4]  
HANISCH RJ, 1997, DECONVULTION IMAGES
[5]  
LIANG Q, 1998, IEEE INT C IM P, V2, P867
[6]  
Oppenheim A. V., 1989, DISCRETE TIME SIGNAL
[7]   Reproducibility of ultrasonographically determined intima-media thickness is dependent on arterial wall thickness - The Tromso study [J].
StenslandBugge, E ;
Bonaa, KH ;
Joakimsen, O .
STROKE, 1997, 28 (10) :1972-1980
[8]   Two-dimensional noise-robust blind deconvolution of ultrasound images [J].
Taxt, T ;
Strand, J .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2001, 48 (04) :861-866
[9]   Comparison of cepstrum-based methods for radial blind deconvolution of ultrasound images [J].
Taxt, T .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (03) :666-674
[10]   Noise robust one-dimensional blind deconvolution of medical ultrasound images [J].
Taxt, T ;
Frolova, GV .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (02) :291-299