An improved application for the enzyme multipled immunoassay technique for caffeine, amikacin, and methotrexate assays on the dade-behring dimension RxL max clinical chemistry system

被引:20
作者
Mendu, Damodara Rao
Chou, Peter P.
Soldin, Steven J.
机构
[1] Childrens Natl Med Ctr, Dept Lab Med, Washington, DC 20010 USA
[2] Georgetown Univ, Dept Med, Washington, DC USA
[3] Georgetown Univ, Dept Oncol, Washington, DC USA
[4] Georgetown Univ, Dept Pharmacol, Washington, DC USA
[5] Georgetown Clin Res Ctr, Washington, DC USA
[6] Quest Diagnost, Dept Clin Lab Management, Chantilly, VA USA
[7] George Washington Univ, Sch Med, Dept Pediat & Pathol, Washington, DC USA
关键词
caffeine; amikacin and methotrexate; Dade-Behring Dimension RxL Max clinical chemistry system; quantitative measurement; comparative analysis;
D O I
10.1097/FTD.0b013e3181570791
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Caffeine is widely used in children's hospitals to treat neonatal apnea. Amikacin is used for treating hospital-acquired infections caused by Gram-negative bacteria resistant to other aminoglycosides. The blood levels, however, have to be monitored carefully because of its ototoxicity and nephrotoxicity. Methotrexate (MTX) is used as a chemotherapeutic agent in the treatment of leukemia and lymphoma as well as of certain solid tumors. Current enzyme multiplied immunoassay technique (EMIT) assays for caffeine, amikacin, and MTX lack low-end precision. In addition, the EMIT assays for MTX lack the sensitivity of reliable quantification to 0.05 mu mol/L needed because of today's more rigorous requirements. The goal of the present study was to optimize the EMIT method parameters on the Dimension RxL Max, thereby providing applications with improved precision for all the three analytes and enhancing the sensitivity of the EMIT methotrexate assay. Serum samples were measured for caffeine, amikacin, and MTX by EMIT on the Dimension RxL Max and by EMIT (on the Olympus AU 600 for caffeine) and fluorescence polarization immunoassay [for MTX and amikacin (FPIA; TDx FLx)] at Quest Diagnostics. The new instrument method parameters that use larger sample volumes and longer observation of optical density changes (caffeine, MTX) provide improved sensitivity for MTX permitting reliable measurement at 0.05 mu mol/L and improved precision for all three analytes. Within- and between-day imprecision were less than 6% for low to high concentrations of caffeine and amikacin controls and are less than 7.5% for MTX concentrations greater than 0.3 mu mol/L and 12.3% at 0.06 mu mol/L. The correlation coefficients for caffeine, amikacin, and MTX plotted for the Dimension RxL Max versus the methods used at Quest Diagnostics were 0.973, 0.986, and 0.992, respectively. These EMIT method applications now compare well with other established assays. The new Dimension RxL Max method parameters provide greatly improved precision and also meet today's clinical sensitivity guidelines (0.05 mu mol/L) for MTX.
引用
收藏
页码:632 / 637
页数:6
相关论文
共 25 条
[1]   INTERCONVERSION OF THEOPHYLLINE AND CAFFEINE IN NEWBORN-INFANTS [J].
BADA, HS ;
KHANNA, NN ;
SOMANI, SM ;
TIN, AA .
JOURNAL OF PEDIATRICS, 1979, 94 (06) :993-995
[2]   DETERMINATION OF AMIKACIN IN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ULTRAVIOLET DETECTION [J].
BARENDS, DM ;
BLAUW, JS ;
SMITS, MH ;
HULSHOFF, A .
JOURNAL OF CHROMATOGRAPHY, 1983, 276 (02) :385-394
[3]   Liquid chromatographic method for the simultaneous determination of caffeine and fourteen caffeine metabolites in urine [J].
Bendriss, EK ;
Markoglou, N ;
Wainer, IW .
JOURNAL OF CHROMATOGRAPHY B, 2000, 746 (02) :331-338
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
BLEYER WA, 1978, CANCER-AM CANCER SOC, V41, P36, DOI 10.1002/1097-0142(197801)41:1<36::AID-CNCR2820410108>3.0.CO
[6]  
2-I
[7]  
CHABNER BA, 1986, METHOTREXATE CANC CH, P1
[8]   ANALYSIS OF METHOTREXATE AND 7-HYDROXYMETHOTREXATE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND PRELIMINARY CLINICAL-STUDIES [J].
COLLIER, CP ;
MACLEOD, SM ;
SOLDIN, SJ .
THERAPEUTIC DRUG MONITORING, 1982, 4 (04) :371-380
[9]  
COOK CE, 1976, J PHARMACOL EXP THER, V199, P679
[10]   Simple high-performance liquid chromatography method for the simultaneous determination of serum caffeine and paraxanthine following rapid sample preparation [J].
Holland, DT ;
Godfredsen, KA ;
Page, T ;
Conner, JD .
JOURNAL OF CHROMATOGRAPHY B, 1998, 707 (1-2) :105-110