Acetylsalicylic acid determination in pharmaceutical samples by FIA-potentiometry using a salicylate-sensitive tubular electrode with an ethylene-vinyl acetate membrane

被引:44
作者
Rover, L [1 ]
Garcia, CAB [1 ]
Neto, CD [1 ]
Kubota, LT [1 ]
Galembeck, F [1 ]
机构
[1] UNICAMP, Inst Chem, BR-13083970 Campinas, SP, Brazil
关键词
acetylsalicylic acid determination; ethylene-vinyl acetate; salicylate-sensitive tubular detector; FIA;
D O I
10.1016/S0003-2670(97)00723-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The construction of a salicylate-sensitive tubular potentiometric electrode for flow injection analysis (FIA) is described, in which an ethylene-vinyl acetate (EVA) copolymeric membrane, 40% in vinyl acetate, is applied directly onto a conducting support consisting of a mixture of an epoxy resin and graphite. The tubular salicylate ion-selective electrode is based on tricaprylyl-trimethyl-ammonium (Aliquat(TM)) salicylate as the ion-exchanger, supported on the polymeric matrix without plasticizer solvent. The tubular electrode showed a response of 58.3 mV per concentration decade within the 5.0x10(-3)-5.0x10(-1) mol l(-1) salicylate range in 0.1 mol l(-1) Tris-HCl buffer (pH=7.5). Iodide is the main analytical interferent; the electrode response time was 5 s at 25 degrees C. Determinations of acetylsalicylic acid in tablet samples by the proposed method exhibit relative errors lower than 2%, when compared with a spectrophotometric method. The useful lifetime of the tubular sensor is longer than 3 months. (C) 1998 Elsevier Science B.V.