Effect of phospholipidic boundary lubrication in rigid and compliant hemiarthroplasty models

被引:27
作者
Foy, JR
Williams, PF
Powell, GL
Ishihara, K
Nakabayashi, N
LaBerge, M
机构
[1] Virginia Polytech Inst & State Univ, Dept Engn Sci & Mech, Blacksburg, VA 24061 USA
[2] Howmed Inc, Rutherford, NJ USA
[3] Clemson Univ, Dept Biol Sci, Clemson, SC 29634 USA
[4] Univ Tokyo, Dept Mat Sci, Tokyo 113, Japan
[5] Tokyo Med & Dent Univ, Inst Med & Dent Engn, Div Organ Mat, Tokyo, Japan
[6] Clemson Univ, Dept Bioengn, Clemson, SC USA
关键词
articular cartilage; friction; phospholipid; poly(MPC-co-BMA; polyurethane elastomer;
D O I
10.1243/0954411991534762
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Hemiarthroplasty may benefit from materials which produce lower friction and improved boundary lubrication protection during start-up conditions. The purpose of this study was to evaluate the effect of phospholipidic boundary lubrication in both rigid and compliant hemiarthroplasty. An in vitro model was designed to dissociate the relative contribution of implant material compliance and the presence of phospholipid to the overall friction of a hemiarthroplasty contact using bovine articular cartilage. Normal bovine articular cartilage was articulated against four flat materials using reciprocating motion: (a) borosilicate glass; (b) borosilicate glass coated with dipalmitoylphosphatidylcholine (DPPC); (c) polyurethane (PU) elastomer (Tecoflex(R)SG93A, a medical-grade aliphatic thermoplastic PU, Thermedics Incorporated, Woburn, Massachusetts); and (d) surface-coated PU (Tecoflex(R)SG93A substrate coated with lipid-attracting copolymer poly [methacryloyloxyethyl phosphorylcholine (MPC)-co-butyl methacrylate (BMA)]). Tests were conducted in physiologically simulated tribological conditions for a non-conformal point contact. Friction and lubrication analysis was performed using both static and kinetic coefficients of friction mu measured for each group as a function of time for a sliding distance of up to 60 m. Results showed that the inclusion of supplemental phospholipid, DPPC, on a rigid substrate significantly decreased mu in comparison with the control (cartilage-glass). Additionally, removal of phospholipid components from the articular cartilage surface produced a significantly greater start-up mu in comparison with normal cartilage at the test onset. The use of a material with a lower modulus resulted in lower mu for the entire duration of the test. Polyurethane elastomer coated with the lipid-attracting copolymer, poly(MPC-co-BMA), resulted in the lowest frictional response. As seen in this study, the improvement of low-modulus hemiarthroplasty may involve the optimization of chemical modification and incorporation of lipid-attracting MPC copolymers onto compliant materials. However, further tests are warranted to determine whether lipid-attracting MPC copolymers perform as well during long-time, in vivo wear studies.
引用
收藏
页码:5 / 18
页数:14
相关论文
共 58 条
  • [51] SMITH ST, 1996, T 5 WORLD BIOM C, P387
  • [52] Swann A C, 1989, Proc Inst Mech Eng H, V203, P143, DOI 10.1243/PIME_PROC_1989_203_026_01
  • [53] THE LUBRICATING ACTIVITY OF SYNOVIAL-FLUID GLYCOPROTEINS
    SWANN, DA
    HENDREN, RB
    RADIN, EL
    SOTMAN, SL
    DUDA, EA
    [J]. ARTHRITIS AND RHEUMATISM, 1981, 24 (01): : 22 - 30
  • [54] UEDA T, 1991, J BIOMAT SCI-POLYM E, V3, P185
  • [55] UNSWORTH A, P IMECHE C TRIB FRIC, P715
  • [56] Evaluation of the frictional properties of an elastomer with enhanced lipid-adsorbing ability
    Williams, PF
    Iwasaki, Y
    Ishihara, K
    Powell, GL
    Gilbert, JA
    Nakabayashi, N
    LaBerge, M
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (05) : 359 - 368
  • [57] FABRICATION AND CHARACTERIZATION OF DIPALMITOYLPHOSPHATIDYLCHOLINE-ATTRACTING ELASTOMERIC MATERIAL FOR JOINT REPLACEMENTS
    WILLIAMS, PF
    POWELL, GL
    LOVE, B
    ISHIHARA, K
    NAKABAYASHI, N
    JOHNSON, R
    LABERGE, M
    [J]. BIOMATERIALS, 1995, 16 (15) : 1169 - 1174
  • [58] WILLIAMS PT, 1993, P I MECH ENG A-J POW, V207, P55, DOI 10.1243/PIME_PROC_1993_207_268_02