Nonenzymatic Sensor for Lactate Detection in Human Sweat

被引:108
作者
Zaryanov, Nikolay V. [1 ]
Nikitina, Vita N. [1 ]
Karpova, Elena V. [1 ]
Karyakina, Elena E. [1 ]
Karyakin, Arkady A. [1 ]
机构
[1] Moscow MV Lomonosov State Univ, Chem Fac, Moscow 119991, Russia
基金
俄罗斯科学基金会;
关键词
ANAEROBIC THRESHOLD; BLOOD LACTATE; BORONIC ACIDS; OXIDASE; BIOSENSOR; ELECTRODE; SACCHARIDES; STABILITY; GLUCOSE; IMMOBILIZATION;
D O I
10.1021/acs.analchem.7b03662
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
For noninvasive diagnostics of hypoxia, we propose the nonenzymatic sensor based on screen-printed structures with the working surface modified in course of electropolymerization of 3-aminophenylboronic acid (3-APBA) with imprinting of lactate. Impedimetric sensor allows lactate detection in the range from 3 mM to 100 mM with the detection limit of 1.5 mM; response time is 2-3 min. Sensor sensitivity remains unchanged within 6 months of storage unpacked in dry state at a room temperature, which is unachievable for enzyme based devices. Analysis of human sweat with poly(3-APBA) based sensor is possible due to (i) much higher lactate content compared to other polyols and (ii) high sensor selectivity (K-lactate(glucose) < 3 10(-2)). Successful detection of lactate in human sweat by means of the poly(3-APBA) based sensor has been confirmed using the highly specific reference method based on lactate oxidase enzyme (correlation coefficient r > 0.9). The attractive performance characteristics of poly(3-APBA) based enzyme-free sensors justify their future use for noninvasive clinical analysis and sports medicine.
引用
收藏
页码:11198 / 11202
页数:5
相关论文
共 50 条
[21]   Novel flexible enzyme laminate-based sensor for analysis of lactate in sweat [J].
Tur-Garcia, Eva L. ;
Davis, Frank ;
Collyer, Stuart D. ;
Holmes, Joanne L. ;
Barr, Hugh ;
Higson, Seamus P. J. .
SENSORS AND ACTUATORS B-CHEMICAL, 2017, 242 :502-510
[22]   Transcutaneous Flexible Sensor for In Vivo Photonic Detection of pH and Lactate [J].
Dat Nguyen ;
Lawrence, Micah M. ;
Berg, Haley ;
Lyons, Monika Aya ;
Shreim, Samir ;
Keating, Mark T. ;
Weidling, John ;
Botvinick, Elliot L. .
ACS SENSORS, 2022, 7 (02) :441-452
[23]   A contact lens with an integrated lactate sensor [J].
Thomas, N. ;
Laehdesmaeki, I. ;
Parviz, B. A. .
SENSORS AND ACTUATORS B-CHEMICAL, 2012, 162 (01) :128-134
[24]   Ultrastable Lactate Biosensor Linearly Responding in Whole Sweat for Noninvasive Monitoring of Hypoxia [J].
Daboss, Elena, V ;
Tikhonov, Dmitrii, V ;
Shcherbacheva, Elizaveta, V ;
Karyakin, Arkady A. .
ANALYTICAL CHEMISTRY, 2022, 94 (25) :9201-9207
[25]   A high-performance nonenzymatic piezoelectric sensor based on molecularly imprinted transparent TiO2 film for detection of urea [J].
Yang, Zheng-peng ;
Liu, Xuan ;
Zhang, Chun-jing ;
Liu, Bao-zhong .
BIOSENSORS & BIOELECTRONICS, 2015, 74 :85-90
[26]   In-ear integrated sensor array for the continuous monitoring of brain activity and of lactate in sweat [J].
Xu, Yuchen ;
de la Paz, Ernesto ;
Paul, Akshay ;
Mahato, Kuldeep ;
Sempionatto, Juliane R. ;
Tostado, Nicholas ;
Lee, Min ;
Hota, Gopabandhu ;
Lin, Muyang ;
Uppal, Abhinav ;
Chen, William ;
Dua, Srishty ;
Yin, Lu ;
Wuerstle, Brian L. ;
Deiss, Stephen ;
Mercier, Patrick ;
Xu, Sheng ;
Wang, Joseph ;
Cauwenberghs, Gert .
NATURE BIOMEDICAL ENGINEERING, 2023, 7 (10) :1307-+
[27]   Electrocodeposition of silver and silicomolybdate hybrid nanocomposite for nonenzymatic hydrogen peroxide sensor [J].
Lin, Kuo Chiang ;
Wu, Tsung Han ;
Chen, Shen Ming .
RSC ADVANCES, 2015, 5 (51) :41224-41229
[28]   A nonenzymatic cholesterol sensor constructed by using porous tubular silver nanoparticles [J].
Li, Yanfen ;
Bai, Hongyan ;
Liu, Qian ;
Bao, Jianchun ;
Han, Min ;
Dai, Zhihui .
BIOSENSORS & BIOELECTRONICS, 2010, 25 (10) :2356-2360
[29]   Research on nonenzymatic electrochemical sensor using HO-BiONO3 nanocomposites for glucose detection [J].
Liu, Gen-qing ;
Zhong, Hui ;
Li, Xiao-rong ;
Yang, Kai ;
Jia, Fei-fei ;
Cheng, Zhi-peng ;
Zhang, Li-li ;
Yin, Jing-zhou ;
Guo, Li-ping ;
Qian, Hai-yan .
SENSORS AND ACTUATORS B-CHEMICAL, 2017, 242 :484-491
[30]   Selective nonenzymatic bilirubin detection in blood samples using a Nafion/Mn-Cu sensor [J].
Noh, Hui-Bog ;
Won, Mi-Sook ;
Shim, Yoon-Bo .
BIOSENSORS & BIOELECTRONICS, 2014, 61 :554-561