Continuous Signal Enhancement for Sensitive Aptamer Affinity Probe Electrophoresis Assay Using Electrokinetic Concentration

被引:66
作者
Cheow, Lih Feng [1 ]
Han, Jongyoon [1 ,2 ]
机构
[1] MIT, Dept Elect Engn & Comp Sci, Cambridge, MA 02139 USA
[2] MIT, Dept Biol Engn, Cambridge, MA 02139 USA
关键词
TYPE-1; REVERSE-TRANSCRIPTASE; CAPILLARY-ELECTROPHORESIS; BINDING; PROTEIN; IGE; LIGANDS; PRECONCENTRATION; ANTIBODIES; MOLECULES; BIOSENSOR;
D O I
10.1021/ac201307d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe an electrokinetic concentration-enhanced aptamer affinity probe electrophoresis assay to achieve highly sensitive and quantitative detection of protein targets in a microfluidic device. The key weaknesses of aptamer as a binding agent (weak binding strength/fast target dissociation) were counteracted by continuous injection of fresh sample while band-broadening phenomena were minimized due to self-focusing effects. With 30 min of continuous signal enhancement, we can detect 4.4 pM human immunoglobulin E (IgE) and 9 pM human immunodeficiency virus 1 reverse transcriptase (HIV-1 RT), which are among the lowest limits of detection (LOD) reported. IgE was detected in serum sample with a LOD of 39 pM due to nonspecific interactions between aptamers and serum proteins. The method presented in this paper also has broad applicability to improve sensitivities of various other mobility shift assays.
引用
收藏
页码:7086 / 7093
页数:8
相关论文
共 36 条
  • [1] Effect of buffer, electric field, and separation time on detection of aptamer-ligand complexes for affinity probe capillary electrophoresis
    Buchanan, DD
    Jameson, EE
    Perlette, J
    Malik, A
    Kennedy, RT
    [J]. ELECTROPHORESIS, 2003, 24 (09) : 1375 - 1382
  • [2] ASSOCIATION OF ASTHMA WITH SERUM IGE LEVELS AND SKIN-TEST REACTIVITY TO ALLERGENS
    BURROWS, B
    MARTINEZ, FD
    HALONEN, M
    BARBEE, RA
    CLINE, MG
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1989, 320 (05) : 271 - 277
  • [3] Increasing the Sensitivity of Enzyme-Linked Immunosorbent Assay Using Multiplexed Electrokinetic Concentrator
    Cheow, Lih Feng
    Ko, Sung Hee
    Kim, Sung Jae
    Kang, Kwan Hyoung
    Han, Jongyoon
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (08) : 3383 - 3388
  • [4] Optimization of aptamer microarray technology for multiple protein targets
    Cho, EJ
    Collett, JR
    Szafranska, AE
    Ellington, AD
    [J]. ANALYTICA CHIMICA ACTA, 2006, 564 (01) : 82 - 90
  • [5] Selection of smart aptamers by equilibrium capillary electrophoresis of equilibrium mixtures (ECEEM)
    Drabovich, A
    Berezovski, M
    Krylov, SN
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2005, 127 (32) : 11224 - 11225
  • [6] INVITRO SELECTION OF RNA MOLECULES THAT BIND SPECIFIC LIGANDS
    ELLINGTON, AD
    SZOSTAK, JW
    [J]. NATURE, 1990, 346 (6287) : 818 - 822
  • [7] HIV-1 viral load assays for resource-limited settings
    Fiscus, Susan A.
    Cheng, Ben
    Crowe, Suzanne M.
    Demeter, Lisa
    Jennings, Cheryl
    Miller, Veronica
    Respess, Richard
    Stevens, Wendy
    [J]. PLOS MEDICINE, 2006, 3 (10) : 1743 - 1750
  • [8] Study of binding stoichiometries of the human immunodeficiency virus type 1 reverse transcriptase by capillary electrophoresis and laser-induced fluorescence polarization using aptamers as probes
    Fu, H
    Guthrie, JW
    Le, XC
    [J]. ELECTROPHORESIS, 2006, 27 (02) : 433 - 441
  • [9] Aptamers as ligands in affinity probe capillary electrophoresis
    German, I
    Buchanan, DD
    Kennedy, RT
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (21) : 4540 - 4545
  • [10] DNA aptamer-based bioanalysis of IgE by fluorescence anisotropy
    Gokulrangan, G
    Unruh, JR
    Holub, DF
    Ingram, B
    Johnson, CK
    Wilson, GS
    [J]. ANALYTICAL CHEMISTRY, 2005, 77 (07) : 1963 - 1970