Accuracy of a 3D fluoroscopic navigation system using a flat-panel detector-equipped C-arm

被引:12
|
作者
Takao, Masaki [1 ]
Yabuta, Kentaro [1 ]
Nishii, Takashi
Sakai, Takashi [1 ]
Sugano, Nobuhiko
机构
[1] Osaka Univ, Grad Sch Med, Dept Orthopaed Sugery, Suita, Osaka 5650871, Japan
基金
日本学术振兴会;
关键词
3D fluoroscopic navigation; flat-panel detector; accuracy; image distortion; SCREW PLACEMENT; CLINICAL ACCURACY; SURGERY; EXCISION; ISO-C-3D;
D O I
10.3109/10929088.2011.602117
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: The aim of this study was to evaluate the accuracy of a novel 3-dimensional (3D) fluoroscopic navigation system using a flat-panel detector-equipped C-arm, focusing on the influence of the distance from the center of fluoroscopic imaging on navigation accuracy. Materials and Methods: A geometric phantom was made using a Styrofoam cube with 25 markers, each consisting of a metal ball 1.5 mm in diameter, fixed in a cross arrangement at 1-cm intervals. Hip joint surgery was simulated using a set of dry pelvic and femoral bones. A total of eight markers were fixed to the acetabulum and proximal femur. Results: In the geometric phantom study, mean target registration error (TRE) was 0.7 mm (range: 0.1-1.5). The TRE of markers located at 5 cm from the imaging center was significantly higher than the TRE of markers located at 1 and 2 cm. However, the TRE was <1 mm in 90% of the overall trials and <1.5 mm in 100%. In the dry bone study, the mean TRE was 0.9 mm (range: 0.7-1.5) over the acetabulum and 1.0 mm (range: 0.5-1.4) over the femur. No significant difference in TRE was seen between the acetabulum and proximal femur. Conclusion: The accuracy of this novel 3D fluoroscopic navigation system was considered acceptable for clinical application. A 3D C-arm equipped with a flat-panel detector could increase the feasibility of 3D fluoroscopic navigation by reducing the effects of image distortion on navigation accuracy.
引用
收藏
页码:234 / 239
页数:6
相关论文
共 50 条
  • [1] Reliability and Usefulness of Intraoperative 3-Dimensional Imaging by Mobile C-Arm With Flat-Panel Detector
    Fujimori, Takahito
    Iwasaki, Motoki
    Nagamoto, Yukitaka
    Kashii, Masafumi
    Takao, Masaki
    Sugiura, Tsuyoshi
    Yoshikawa, Hideki
    CLINICAL SPINE SURGERY, 2017, 30 (01): : E64 - E75
  • [2] Image quality and effective dose of a robotic flat panel 3D C-arm vs computed tomography
    Kraus, Michael
    Fischer, Eric
    Gebhard, Florian
    Richter, Peter H.
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2016, 12 (04) : 743 - 750
  • [3] Clinical Applications of Quantitative Perfusion Imaging with a C-Arm Flat-Panel Detector-A Systematic Review
    Al-Kaylani, Abdallah H. A. Zaid
    Schuurmann, Richte C. L.
    Maathuis, Wouter D.
    Slart, Riemer H. J. A.
    De Vries, Jean-Paul P. M.
    Bokkers, Reinoud P. H.
    DIAGNOSTICS, 2023, 13 (01)
  • [4] Evaluation of image quality and assessability of a new flat-panel 3D C-arm compared to mobile and fixed computed tomography in posterior spinal fixation
    Keil, Holger
    Luxenhofer, Miriam
    Vetter, Sven Y.
    Beisemann, Nils
    Grutzner, Paul A.
    Franke, Jochen
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2021, 17 (02)
  • [5] Volume CT with a flat-panel detector on a mobile, isocentric C-arm: Pre-clinical investigation in guidance of minimally invasive surgery
    Siewerdsen, JH
    Moseley, DJ
    Burch, S
    Bisland, SK
    Bogaards, A
    Wilson, BC
    Jaffray, DA
    MEDICAL PHYSICS, 2005, 32 (01) : 241 - 254
  • [6] Radiation Dose of Interventional Radiology System Using a Flat-Panel Detector
    Chida, Koichi
    Inaba, Yohei
    Saito, Haruo
    Ishibashi, Tadashi
    Takahashi, Shoki
    Kohzuki, Masahiro
    Zuguchi, Masayuki
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (06) : 1680 - 1685
  • [7] C-arm flat-panel CT arthrography of the shoulder: Radiation dose considerations and preliminary data on diagnostic performance
    Guggenberger, Roman
    Ulbrich, Erika J.
    Dietrich, Tobias J.
    Scholz, Rosemarie
    Kaelin, Pascal
    Koehler, Christoph
    Elsaesser, Thilo
    Le Corroller, Thomas
    Pfammatter, Thomas
    Alkadhi, Hatem
    Andreisek, Gustav
    EUROPEAN RADIOLOGY, 2017, 27 (02) : 454 - 463
  • [8] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Masaki Takao
    Hidetoshi Hamada
    Takashi Sakai
    Nobuhiko Sugano
    Archives of Orthopaedic and Trauma Surgery, 2019, 139 : 189 - 195
  • [9] Factors influencing the accuracy of iliosacral screw insertion using 3D fluoroscopic navigation
    Takao, Masaki
    Hamada, Hidetoshi
    Sakai, Takashi
    Sugano, Nobuhiko
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2019, 139 (02) : 189 - 195
  • [10] Detection of Pulmonary Embolism Using a Novel Dynamic Flat-Panel Detector System in Monkeys
    Miyatake, Hidemitsu
    Tabata, Takahisa
    Tsujita, Yasuyuki
    Fujino, Kazunori
    Tanaka, Rie
    Eguchi, Yutaka
    CIRCULATION JOURNAL, 2021, 85 (04) : 361 - 368